Fig.9.3 System menu (page 3)
9.1 System Calibration
For the analyzer to meet its specifications and characteristics the self-calibration routines should be performed
periodically or whenever the ambient temperature changes.
System calibration includes frequency calibration and amplitude calibration. Frequency calibration is to rectify the
frequency error caused by IF filter. Amplitude calibration is to rectify the amplitude error, which can be caused by
window functions of different RBW or attenuators, so Amplitude calibration should be taken with different RBW or
attenuators.
Allow the analyzer to warm up for 10 minutes before performing the following calibration steps:
Press "CAl" soft-key, and the menu will pop up.
Fig.9.4 Calibration menu
NOTE:
During the process of calibration, the instrument is unable to response to the keyboard commands except the