Ex3:
Set the line trigger is line 16
Command:
TRIGger:LINE 17
NOTE:
The line trigger must be done before setting trigger, otherwise this statement doesn't work. Similar with the
words (such as TRIGer) content, if not set, the setting (LINE) may not work. The same way works as below.
5. Mark
Command
Means
CALCulate:MARKer:STATe <Boolean>
CALCulate:MARKer:STATe?
Set /Get the marker state
CALCulate:MARKer:FCOunt[:STATe]<Boo
lean>
CALCulate:MARKer:FCOunt[:STATe]?
Set /Get the frequency count state
CALCulate:MARKer:AOFF
Set the marker to all off
CALCulate:MARKer:X <freq>
CALCulate:MARKer:X?
Set /Get the frequency of marker
CALCulate:MARKer:Y?
Get the level of marker
CALCulate:MARKer:NORMal:XPOS?
Get the x-coordinate of marker
CALCulate:MARKer:TT:XPOS?
Get the x-coordinate of delta deactivate marker
CALCulate:MARKer:TA:XPOS?
Get the x-coordinate of delta activate marker
CALCulate:MARKer:MAXimum
Seek maximum marker
CALCulate:MARKer:MAXimum:NEXT
Seek next maximum marker
CALCulate:MARKer:MAXimum:LEFT
Seek maximum marker on left
CALCulate:MARKer:MAXimum:RIGHt
Seek maximum marker on right
CALCulate:MARKer[:SET]:CENTer
Set marker to center
CALCulate:MARKer[:SET]:RLEVel
Set marker to reference level
Ex1:
Set the marker to ON
Command:
CALCulate:MARKer:STATe ON
Ex2:
Seek maximum marker
Command:
CALCulate:MARKer:MAXimum
Ex3:
Get the power of current marker
Command:
CALCulate:MARKer:Y?
Return:
80.83
Means the level of current marker is 80.83 dBuv ( As the unit of reference level is dBuv)