Fig.3.15 Amplitude Unit
NOTE:
The change of unit will not cause the change of “Log/Line” state.
5) Scale/Div
The scale/div is displayed on the upside of the screen.
In logarithm state: the default value is 10dB/grid. After pressing “Scale/Div” soft-key user can enter the integer number
from 1 to 40 or decimal number from 0.1 to 0.9 to change the scale to the state wanted. The scale/div can be changed
by “Up/Down” cursor or by knob too.
In line State: The “scale/div” state is shown as × n. After pressing “Scale/Div” soft-key, the submenu of it will appear.
User can choose ×1, ×2, ×5 or ×10 in this submenu. 1/n is the ratio of voltage difference between the lowest level
displayed on the screen and reference level. For example when reference level is 1mV, if the Scale/Div is ×1, the lowest
voltage display on the screen is 0mV, while if the Scale/Div is ×2, the lowest voltage display on the screen will be
0.5mV.
Fig.3.16 Scale is 10 dB/ in logarithmic state (default value)