6.5
Setup of Sweep
Sweep refers to the process of continuously measuring data points according to the Stimulus values in the
specified sequence.
6.5.1 Overview of Sweep Types
The network analyzer supports six Sweep types shown below.
1) Linear frequency
This is the default Sweep type of the instrument. The whole frequency range is covered continuously by the
frequency linearity.
2) Logarithmic frequency
In the logarithmic frequency mode, the source frequency increases in logarithmic steps, and the frequency ratio of
every two adjacent frequency points is identical.
3) Power Sweep
The power is Sweepned at the point frequency. The maximum Sweep range is 50dB. The default frequency
Sweep range is -30dBm to +20dBm.
4) Point frequency
In the point frequency Sweep mode, a single Sweep frequency is set in the analyzer, measurement data are
accurately and continuously samples at the intervals determined according to the Sweep time and the number of
measurement points, and changes of measurement data with time are displayed.
5) Segment Sweep
Segment Sweep aims to start Sweep of several segments. The power level, intermediate frequency bandwidth and
Sweep time of each segment can be set independently. After calibration of all segments, one or several
segment(s) can be measured according to the calibrations. Segments are defined according to the frequency
increase sequence, and the frequency range must be prevented from overlapping. The attenuator Setups of the
power levels of all segments must be identical to prevent damage caused by frequent switching of the attenuator. If
the attenuator Setups of the current segment and defined segment are different, the analyzer will automatically
change the power level and attenuator Setups of the defined segment.
6) Phase Sweep
Sweep the phase of one or several source(s) relative to another source. The measured value is -360° to +360°.
6.5.2 Setup of Sweep Type
1)
Setup of Sweep type
Menu path:
[Stimulus] > [Sweep] > [Sweep type]
.
Select the corresponding Sweep type in the auxiliary menu bar or in the [Setup of Sweep type] dialog box.