Fig. 6-4 Definition of Parameter
6.2.3 Non-ratio Power Measurement
The absolute power of the receiver A, B, C, D, R1, R2, R3 and R4 can be measured in the non-ratio power
measurement mode, however, the phase, group delay and other items under the ―average
‖
function cannot
be measured.
1) New non-ratio power measurement Trace
Menu path:
[Trace] > [New Trace]
. Click “
Receiver
” in the dialog box.
2) [New Trace] dialog box (any ratio and non-ratio power measurement)
a)
[Activation]
check box
Tick it to create a new Trace.
b) [Numerator] zone
Select the numerator of any ratio measurement or the receiver for non-ratio power measurement.
c) [Denominator] zone
Select the denominator of any ratio measurement. The value is 1 in non-ratio power measurement.
d) [Source port] box
Select the source signal output port of the analyzer.
e) [Channel selection] box
Select the channel for the new Trace.
f) [Create in new window] check box
If this check box is selected, a Trace will be created in the new window. If this check box is cleared, a Trace will
be created in the current active window.