background image

Tel: 886. 909 602 109 Email: [email protected]

www.salukitec.com

46

3) Setup of center frequency and frequency span

Menu path:

[Stimulus] > [Freq]

Fig. 6-8 Setup of Center Frequency and Frequency Span

Prompt

[Freq] shortcut key

The frequency can be rapidly set with the

[Freq]

shortcut key in the functional key zone of the front panel.

6.4

Setup of Signal Power Level

The power level refers to the power level of the output signal of the test port. The power level indicators of the port

of S3602 series vector network analyzer are as follows.

Table 6-1 Power Level Indicators of Source Output Signal

Frequency range

Source frequency

range (dBm)

Configuration

10MHz to 13.5GHz/26.5GHz/43.5GHz/50GHz/67GHz

-30 to +20

Standard configuration

10MHz to 13.5GHz/26.5GHz

-100 to +20

Option: program-controlled

step attenuator

10MHz to 43.5GHz/50GHz

-90 to +20

Option: program-controlled

step attenuator

1) Setup of power level

Menu path:

[Stimulus] > [Power]

. Also, the

[Power]

shortcut key is set in the

functional

key zone of the front

panel and the shortcut menu bar.

Содержание S3602 Series

Страница 1: ...S3602 Series Vector Network Analyzer Quick Start Guide Saluki Technology Inc...

Страница 2: ...ter modulation Measurement For inter modulation signal measurement of amplifier Option 400 S80 is needed S3602A 003 Vector Noise Figure Measurement The built in 10MHz 13 5GHz high sensitivity noise re...

Страница 3: ...or the measurement of embedded LO frequency converters Option 400 S82 or S83 S80 are needed S3602A S86 Gain Compression Two Dimension Sweep Measurement For the gain compression two dimension sweep tes...

Страница 4: ...e system Built in crosstalk evaluation modules such as NEXT FEXT PSXT ILD ICR and ICN Built in standards such as IEEE 802 3 PCIe SAS and SATA help to quickly evaluate whether S parameters Meet the des...

Страница 5: ...to build the system and use on the cabinet S3602B EWT1 Extend 1 year warranty S3602C Part No Name Description S3602C 201 2 Port Programmable Step Attenuator Set two 60dB programmable step attenuators...

Страница 6: ...rement For time domain test can locate and analyze the discontinuous positions in devices fixtures or cables S3602C S11 Advanced Time Domain Analysis Used for TDR time domain impedance test eye diagra...

Страница 7: ...MHz Noise figure gain measurement range 0 to 55dB Noise source driving voltage 28V 1V Noise jitter 0 15dB S3602D 008 Pulse Measurement For pulse S parameter measurement S3602D 480 Four port Spread Spe...

Страница 8: ...02D S86 Gain Compression Two Dimension Sweep Measurement For the gain compression two dimension sweep test of amplifier S3602D S88 Phase Scan Option Used for phase scan measurement need option S3602D...

Страница 9: ...r automatic testing and removal of single ended and balanced device measurement fixtures S3602E S10 Time Domain Measurement For time domain test can locate and analyze the discontinuous positions in d...

Страница 10: ...com 10 1 85 1 85 to 1 85mm female Top Rack Easy to build the system and use on the cabinet S3602E 061 Extension Workbench S3602E EWT1 Extend 1 year warranty For S3602E main machine S3602E 3648 EWT1 E...

Страница 11: ...or reproduced or transmitted for the purpose of making profit without its prior permission otherwise Saluki will reserve the right to investigate and affix legal liability of infringement Product Qual...

Страница 12: ...ector 28 3 3 1 Connector Inspection 28 3 3 2 Connection Method 28 3 3 3 Disconnection Method 29 3 3 4 Use of Torque Wrench 29 4 Instrument Power on off 31 4 1 Connection of power supply 31 4 2 Start u...

Страница 13: ...5 1 Overview of Sweep Types 49 6 5 2 Setup of Sweep Type 49 6 5 3 Sweep Time 51 6 5 4 Setup of Sweep 52 6 6 Trigger Mode 52 6 6 1 Simple Trigger Setup 53 6 6 2 Detailed Trigger Setup 53 6 6 3 Trigger...

Страница 14: ...chapter introduces the inspection before operation operating system configuration initial configuration appearance etc of S3602 series vector network analyzer to help the user make preparation for pr...

Страница 15: ...ive function library etc It aims to guide the user to rapidly and comprehensively know the program control commands and methods of the instrument It mainly includes the following chapters Remote contr...

Страница 16: ...ention to the following requirements before operation The improper operation location or measurement Setup may result in damage to the instrument or the connected instrument Pay attention to the follo...

Страница 17: ...x are heavy and should be handled by two people with care 2 1 2Instrument placement and measurement In order to ensure normal functions of S3602 series vector network analyzer pay attention to the fol...

Страница 18: ...eration interface of S3602 series vector network analyzer 3 1 Front Panel This section introduces the structure and functions of the front panel of S3602 series vector network analyzer The front panel...

Страница 19: ...he cursor back and delete the original input after entering e Number keys Include 0 9 Enter the numbers in measurement Setup and press the corresponding unit keys to complete input f Unit keys End the...

Страница 20: ...ortcut key to set the measurement frequency The starting frequency ending frequency center frequency frequency span and frequency offset b Power key It is a shortcut key to set the measurement power T...

Страница 21: ...s will be displayed on screen Measurement Format Scale Display and Average f Calibration key Open the main calibration menu Following options will be displayed on screen Calibration Correction ON OFF...

Страница 22: ...ive for Record Run 1 and cannot be used to control Record Run 2 c Macro local key If the analyzer is programmed press this key to switch to the macro state If the analyzer is not programmed press this...

Страница 23: ...nalyzer will automatically exit the application program When the power supply will be turned off the standby state will be enabled This is only a standby switch It cannot be directly connected with th...

Страница 24: ...ent instead of the internal reference frequency The 10MHz reference input port has the following features Input frequency 10MHz 1ppm Input level 15dBm to 20dBm Input impedance 200 b 10MHZ reference ou...

Страница 25: ...onnector can be connected to the USB mouse keyboard or other USB interface devices The rear panel is equipped with four interfaces in total Type B connector is mainly used for control The network anal...

Страница 26: ...RF output signal power 3dBm 2dB Fig 3 16 LO and RF Output Connector 7 Pulse input output connector As shown in Fig 3 17 this is a 15 pin D type female connector The working status of the internal puls...

Страница 27: ...nnector 12 Trigger input output interface connector As shown in Fig 3 22 this interface is an external and auxiliary trigger input output interface Specific functions are as follows External trigger i...

Страница 28: ...Inspection Wear the anti state wristband in connector inspection It is recommended to use the magnifying lens to check the following items 1 Check whether the plating surface is worn or deeply scratch...

Страница 29: ...is set properly before use Ensure that the angle between the torque wrench and the other wrench used to support the connector or cable is less than 90 before applying force Carefully grip the handle...

Страница 30: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 30 Fig 3 24 Use of Torque Wrench...

Страница 31: ...tor above the power switch of the front panel as shown in Fig 4 3 becomes yellow Step 3 Turn on the power switch of the front panel as shown in Fig 4 3 Do not connect any device to the network analyze...

Страница 32: ...the power supply of the instrument otherwise the instrument cannot be shut down normally and may be damaged or the current status measurement data may be lost Shut down the instrument properly In cas...

Страница 33: ...Tel 886 909 602 109 Email sales salukitec com www salukitec com 33 Fig 4 4 System Recovery Screen...

Страница 34: ...use on the screen to display the right click menu Click the Trace bar to select the current activation Trace Right click the mouse on the Trace bar to set the current activation Trace in the right cli...

Страница 35: ...close the status bar display Or press the status bar opening closing item of the auxiliary menu bar to open or close the status bar 5 1 2 Tool bar 1 Six different toolbars can be displayed at the same...

Страница 36: ...imulus including Starting Stimulus value CW frequency in the CW mode Ending Stimulus value time in the CW mode Number of sweep points g Time domain toolbar Time domain ON OFF check box Gate ON OFF che...

Страница 37: ...rsor for relative measurement Stimulus value Response value b Limit table Fig 6 29 Limit Table The limit table is used to display set and modify the limit test parameters including Type minimum maximu...

Страница 38: ...hat the user can observe the current measurement state including Title Trace state Frequency Stimulus state Cursor display Display time state a Title One title can be set for each window The title wil...

Страница 39: ...des the cursor reading ON OFF Trace cursor ON OFF large font ON OFF leftward movement of display position and downward movement of display position The cursor reading information is displayed in the u...

Страница 40: ...le bar to hide all title bars on the screen In this case the display of measurement results on the screen can be maximized Triggering of title bar display Menu path Response Display Click Title bar in...

Страница 41: ...et on the front panel to return to the known default state i e reset state For Details of reset state please refer to User Manual Section 4 1 1 6 1 2 User Reset State The analyzer can reset into the u...

Страница 42: ...ar 6 2 Selection of Measurement Parameter The following parameters can be set in an S3602 series vector network analyzer to measure the electrical characteristics of devices S parameter fixed ratio An...

Страница 43: ...a S Parameter selection zone Select the S parameter measurement Trace to be created b Receiver button Open the dialog box to create any ratio or non ratio power measurement Trace c Balance parameter...

Страница 44: ...dialog box any ratio and non ratio power measurement a Activation check box Tick it to create a new Trace b Numerator zone Select the numerator of any ratio measurement or the receiver for non ratio...

Страница 45: ...in the window The corresponding Trace will be activated Fig 6 5 Change of Trace Activation Status 2 Change the measurement parameter of the activated Trace Menu path Response Measure The measurement...

Страница 46: ...gnal of the test port The power level indicators of the port of S3602 series vector network analyzer are as follows Table 6 1 Power Level Indicators of Source Output Signal Frequency range Source freq...

Страница 47: ...ower Power The Power dialog box will appear Click Slope check box Click the Slope input box and enter the power slope 5 Power dialog box a Power ON off check box If this check box is ticked the port w...

Страница 48: ...uation can be set manually for certain measurement such as the reflection amplifier oscillator or unsteady state amplifier measurement In this case the matching source impedance better than the return...

Страница 49: ...ed according to the Sweep time and the number of measurement points and changes of measurement data with time are displayed 5 Segment Sweep Segment Sweep aims to start Sweep of several segments The po...

Страница 50: ...the independent intermediate frequency bandwidth of each segment Independent Sweep time check box tick it to set the independent Sweep time of each segment Display table button click it to create and...

Страница 51: ...dth of the segment the intermediate frequency bandwidth option is opened Double click the Sweep time box of the segment and set the Sweep time the independent Sweep time option is opened 6 5 3 Sweep T...

Страница 52: ...k in the analog Sweep or step Sweep mode depending on the Setup of the Sweep time and intermediate frequency bandwidth c Residence time input box Set the residence time at each point before obtaining...

Страница 53: ...e current active channel can be set by simple trigger Setup Menu path Stimulus Trigger The trigger sub menu will appear Click Continuous Single or Hold in the sub menu and select the trigger mode Fig...

Страница 54: ...one measurement Manual The manual trigger signal is generated in the following methods 1 Click the Manual trigger button in the Trigger dialog box 2 Click the Manual trigger item in the Trigger sub m...

Страница 55: ...the channel will enter the holding mode Hold The channel will not receive the trigger signal and Sweep will be stopped Spot Trigger mode check box The spot Sweep mode can be selected only when the tr...

Страница 56: ...er output signal in the radio box Positive pulse or Negative pulse Position Set the time to send the trigger output signal in the radio box Before data capturing or After data capturing Pulse width In...

Страница 57: ...or can be set respectively in the pulse measurement interface The external pulse synchronization input signal will be input through the pulse input output interface of the rear panel of the analyzer T...

Страница 58: ...he characteristics of the tested device This section introduces nine data formats and the method of scale Setup so as to better display the measurement information 6 7 1 Data Format 1 Rectangular coor...

Страница 59: ...smission time through the tested device Y axis unit time second Applicable typical measurement group delay d Linear amplitude format Only display the positive value Y axis no unit U in ratio measureme...

Страница 60: ...d by the displacement of this point relative to the central point or zero point In the default mode the amplitude is a linear scale and the excircle scale is set as 1 b The phase of any point is deter...

Страница 61: ...nce of the upper part of Smith chart is positive Therefore the upper part is the sensitive zone e The reactance of the lower part of Smith chart is negative Therefore the lower part is the capacitive...

Страница 62: ...vertical grids of the window Reference position button The reference position refers to the position of the reference line in the rectangular coordinate figure The position of the bottom line in the...

Страница 63: ...3 Setup D Four Traces will be displayed in two windows in the logarithmic format in two channels S11 and S21 will be displayed in Window 1 of Channel 1 and S12 and S22 in Window 2 of Channel 2 1 Preco...

Страница 64: ...rtically overlapped Three window arrangement All Traces are displayed in three windows Two of the three windows are arranged in the upper part and the other in the lower part Four window arrangement A...

Страница 65: ...tion accuracy of calibration data is associated with the status Setup To realize the highest accuracy of measurement the instrument Setups should be identical in file calling and calibration otherwise...

Страница 66: ...t contents through the printer or printing into the specified file The local or network printer can be used The printer can be of parallel serial or USB interface type Add the printer in the WINDOWS o...

Страница 67: ...Measurement Information Printing into File The analyzer supports the output of print contents into the bitmap BMP file To print several pages various BMP files will be created automatically respectiv...

Страница 68: ...t up process refer to the System recovery of analyzer in 2 2 Configuration of Operating System Check whether the measurement application program is running In case of failure in running click the prog...

Страница 69: ...h Handle with care on the box body 6 Transport the network analyzer as a precision instrument 7 Keep the copies of all transport Documents Pay attention to the packing of the network analyzer The inst...

Отзывы: