65
Save files
Menu path:
[File] > [Save] > [Save]/[Save as...]
.
Fig. 6-27 Save files
Recall of file
1) Call status and calibration data:
a) sta file
The status data of the instrument are saved in the sta file, including the Setups, Trace data, limit line and
cursor.
b) cal file
Calibration data are saved in the cal file, excluding status data.
The correction accuracy of calibration data is
associated with the status Setup. To realize the highest accuracy of measurement, the instrument Setups should be
identical in file calling and calibration; otherwise, the accuracy of calibration will not be guaranteed.
c) cst file
All measurement status and calibration data of the instrument are saved in the cst file, therefore, the cst file
can be called to save time.
2) File calling:
Menu path:
[File] > [Recall]
. The dialog box will appear.
Select the load file type in the
[File type]
box.
Set the call file directory through the
[Search range]
box and the
[File list]
below.