Tel: 886. 909 602 109 Email: [email protected]
47
Fig. 6-9 Setup of Power Level
2) Cutoff of port power
Menu path:
[Stimulus] > [Power] > [Power state ON/off]
. Or, switch the power by changing the Setup of the
[Power ON/off]
check box in the [Power Setup] dialog box.
3) Manual Setup of attenuation
Menu path:
[Stimulus] > [Power] > [Power and attenuation...]
. The
[Power and attenuation]
will appear.
Click to clear the
[Auto attenuation]
zone check box. Click the
[Attenuation]
input box and enter the attenuation
value.
4) Setup of power slope
Menu path:
[Stimulus] > [Power] > [Power...]
. The [Power] dialog box will appear. Click
[Slope]
check box.
Click the
[Slope]
input box and enter the power slope.
5) [Power] dialog box
a) [Power ON/off] check box
If this check box is ticked, the port will normally output the power. If this check box is cleared, the port output
power will be cut off.
b) [Power-port 1]
Set the power level of the port 1.
c) [Power slope] check box
Tick the
[Power slope]
check box and enter the slope value in the
[Slope]
input box to enable the power slope
function.
The power slope is used to compensate the power loss of the cable and test fixture as a result of frequency increase.
If the power slope function is enabled, the output power of the test port will increase (or decrease) with the
increase of Sweep frequency.