![RS PRO RSDM-9060 Скачать руководство пользователя страница 88](http://html1.mh-extra.com/html/rs-pro/rsdm-9060/html/rs-pro/rsdm-9060/rsdm-9060_user-manual_1496059088.webp)
88
rspro.com
Example
Vs = Voltage source
RLoad = Load under test
Rint = Current terminal total impedance
containing Fuse + Rline
Rline
+ Rline
Rline
When different current range for measurement is
selected, Rshunt will vary accordingly.
For example,
Vs = 10V, Rload = 10 Ω, Rint = total impedance
flowing through current terminal 0.5Ω
The theoretical value for current reading should be
= 1A in that the DMM internal
resistor Rint, which contains Shunt, Rline
Rline
,
Rline
Rline
and Fuse, will cause impact on the
measuring reading.
The measured value is I
=
= 0.952381 A.
Error (%) =
* 100
This error is applicable to not only DC but AC
measurement, and the burden voltage, per varied
current measuring range, is generally within the
range of several hundreds mV.
DC Current
Range
Shunt
Burden Voltage
100 μA 100 Ω
<0.011 V
1 mA
100 Ω
<0.11 V
10 mA 1 Ω
<0.04 V
100 mA 1 Ω
<0.4 V
Содержание RSDM-9060
Страница 1: ...User Manual RSDM 906X series Stock number 1919720 RSDM 9061 1919721 RSDM 9060 EN rspro com...
Страница 2: ......
Страница 5: ...5 Dimensions 409 Declaration of Conformity 410...
Страница 23: ...23 DC AC 3 15A Input Current Fuse Holds the current fuse T3 15A 500V 5 20mm For fuse replacement details...
Страница 141: ...141 EOM pulse width timing...
Страница 160: ...160 rspro com SYSTEM FIRMWARE View System Info 161 Firmware Update 162...
Страница 211: ...211 SCREENHOT LOG Capture 212 Save Reading 215...
Страница 219: ...219 DISPLAY SETTING Digit 220 Display 222 Number 222 Bar Meter 223 Trend Chart 227 Histogram 238...
Страница 409: ...409 Dimensions All dimensions are shown in millimeters...
Страница 412: ...412 rspro com...