[SENSe:]VOLTage[:DC]:IMPedance:AUTO?
Returns the Automatic input impedance mode.
Return parameter: 0 | 1, 0=OFF, 1=ON
[SENSe:]VOLTage[:DC]:NPLCycles
Sets the integration time for DC Voltage measurement in PLCs
(power line cycles). Where one PLC is equal to 16.6
milliseconds. For any <NRf> parameter, the DMM will
automatically set the PLC to the closest acceptable PLC value
(0.006 | 0.0083 | 0.0125 | 0.025 | 0.05 | 0.15 | 0.6 | 1 | 3 | 12).
Parameter: NPLCycles(<NRf> | MIN | MAX | DEF)
Example: SENS:VOLT:DC:NPLC 12
Sets the integration time to 12 PLCs for DC Voltage
measurements.
[SENSe:]VOLTage[:DC]:NPLCycles?
Returns the integration time for DC Voltage measurement in
PLCs (power line cycles). Where one PLC is equal to 16.6
milliseconds.
Return parameter: 0.006 | 0.0083 | 0.0125 | 0.025 | 0.05 | 0.15 |
0.6 | 1 | 3 | 12
[SENSe:]VOLTage[:DC]:NULL[:STATe]
Sets the relative function on/off for DC Voltage measurement.
Parameter: 0 | 1 | ON | OFF
Example: SENS:VOLT:DC:NULL:STAT OFF
Turns the relative function off for DC Voltage measurement.
[SENSe:]VOLTage[:DC]:NULL[:STATe]?
Returns the relative function state of DC Voltage measurement.
Return parameter: 0 | 1, 0=OFF, 1=ON
[SENSe:]VOLTage[:DC]:NULL:VALue
Sets the relative value for DC Voltage measurement.
Parameter: <NRf> (-1200.0~1200.0 V) | MIN | MAX | DEF
Example: SENS:VOLT:DC:NULL:STAT ON
SENS:VOLT:DC:NULL:VAL 1.2
Sets the relative value to 1.2V for DC Voltage measurement.
[SENSe:]VOLTage[:DC]:NULL:VALue?
Returns the current relative value of DC Voltage measurement.
Содержание RSDM-9060
Страница 1: ...User Manual RSDM 906X series Stock number 1919720 RSDM 9061 1919721 RSDM 9060 EN rspro com...
Страница 2: ......
Страница 5: ...5 Dimensions 409 Declaration of Conformity 410...
Страница 23: ...23 DC AC 3 15A Input Current Fuse Holds the current fuse T3 15A 500V 5 20mm For fuse replacement details...
Страница 141: ...141 EOM pulse width timing...
Страница 160: ...160 rspro com SYSTEM FIRMWARE View System Info 161 Firmware Update 162...
Страница 211: ...211 SCREENHOT LOG Capture 212 Save Reading 215...
Страница 219: ...219 DISPLAY SETTING Digit 220 Display 222 Number 222 Bar Meter 223 Trend Chart 227 Histogram 238...
Страница 409: ...409 Dimensions All dimensions are shown in millimeters...
Страница 412: ...412 rspro com...