![RS PRO RSDM-9060 Скачать руководство пользователя страница 311](http://html1.mh-extra.com/html/rs-pro/rsdm-9060/rsdm-9060_user-manual_1496059311.webp)
311
CONFigure:CURRent:AC
Sets measurement to AC Current on the 1
st
display and
specifies range.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX | DEF)]
Example: CONF:CURR:AC 10e-2
Sets the measurement mode to AC Current with a 100mA
range.
CONFigure:RESistance
Sets measurement to 2-wire Resistance on the 1
st
display and
specifies range/resolution.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example:
CONF:RES 10e3,MIN
Sets the measurement mode to 2-wire Resistance with a 10k
Ω
range at the lowest resolution.
●
Autoranging (AUTO or DEFault), will generate an error if you
specify a <resolution> because the instrument cannot
accurately resolve the integration time (especially if the input
continuously changes). If your application requires
autoranging, specify DEFault for the <resolution> or omit the
<resolution> altogether.
CONFigure:FRESistance
Sets measurement to 4-wire Resistance on the 1
st
display and
specifies the range/resolution.
Parameter: [None] | [Range(<NRf> | AUTO | MIN | MAX |
DEF)[,Resolution(<NRf> | MIN | MAX | DEF)]]
Example: CONF:FRES 1e3,MAX
Sets the measurement mode to 4-wire Resistance with a range
of 1k
Ω
at the maximum resolution.
●
Autoranging (AUTO or DEFault), will generate an error if you
specify a <resolution> because the instrument cannot
accurately resolve the integration time (especially if the input
continuously changes). If your application requires
autoranging, specify DEFault for the <resolution> or omit the
<resolution> altogether.
Содержание RSDM-9060
Страница 1: ...User Manual RSDM 906X series Stock number 1919720 RSDM 9061 1919721 RSDM 9060 EN rspro com...
Страница 2: ......
Страница 5: ...5 Dimensions 409 Declaration of Conformity 410...
Страница 23: ...23 DC AC 3 15A Input Current Fuse Holds the current fuse T3 15A 500V 5 20mm For fuse replacement details...
Страница 141: ...141 EOM pulse width timing...
Страница 160: ...160 rspro com SYSTEM FIRMWARE View System Info 161 Firmware Update 162...
Страница 211: ...211 SCREENHOT LOG Capture 212 Save Reading 215...
Страница 219: ...219 DISPLAY SETTING Digit 220 Display 222 Number 222 Bar Meter 223 Trend Chart 227 Histogram 238...
Страница 409: ...409 Dimensions All dimensions are shown in millimeters...