![RS PRO RSDM-9060 Скачать руководство пользователя страница 406](http://html1.mh-extra.com/html/rs-pro/rsdm-9060/rsdm-9060_user-manual_1496059406.webp)
406
rspro.com
[2]
This specification is available when both sine wave and square wave input
≥
100 mV. For the input of 10 mV to 100 mV, the % of reading error needs to be
multiplied by 10 times.
[3]
The amplitude range is from 10% to 120% and is lower than 750 VAC.
[4]
The input
≥
60 mV, for 300 k ~ 1 MHz, within 100mV range.
Temperature Characteristics
[1]
(Exclusive of probe errors)
RTD (Accuracy based on PT100):
(100
Ω
platinum [PT100], D100, F100,PT385, PT3916, or user type)
Range
Resolution
1 Year (23°C
±5°C)
Temperature Coefficient
0°-18°C & 28°-55°C
-200 °C~ -100 °C
0.001 °C
0.09 °C
0.004 °C / °C
-100 °C~-20 °C
0.001 °C
0.08 °C
0.005 °C / °C
-20 °C~20 °C
0.001 °C
0.06 °C
0.005 °C / °C
20 °C~100 °C
0.001 °C
0.08 °C
0.005 °C / °C
100 °C~300 °C
0.001 °C
0.12 °C
0.007 °C / °C
300 °C~600 °C
0.001 °C
0.22 °C
0.009 °C / °C
Thermocouples (Accuracy based on ITS-90):
Type Range
Resolution
90 Day/1 Year
(23 °C±5 °C)*
Temperature Coefficient
0°-18 °C & 28°-55 °C
E
-200 to +1000 °C 0.002 °C
0.2 °C
0.03 °C / °C
J
-210 to +1200 ˚C 0.002 °C
0.2 °C
0.03 °C / °C
T
-200 to +400 °C
0.002 °C
0.3 °C
0.04 °C / °C
K
-200 to +1372 ˚C 0.002 °C
0.3 °C
0.04 °C / °C
N
-200 to +1300 ˚C 0.003 °C
0.4 °C
0.05 °C / °C
R
-50 to +1768 ˚C
0.01 °C
1 °C
0.14 °C / °C
S
-50 to +1768 ˚C
0.01 °C
1 °C
0.14 °C / °C
B
+350 to +1820 ˚C 0.01 °C
1 °C
0.14 °C / °C
*Relative to simulated junction
Thermistor: (2.2 k
Ω
, 5 k
Ω
, 10 k
Ω
or User Type)
Range
Resolution
90 Day/1 Year
(23 °C±5 °C)*
Temperature
Coefficient / °C
–80 ° to 150 °C
0.001 °C
0.1 °C
0.003 °C/ °C
Содержание RSDM-9060
Страница 1: ...User Manual RSDM 906X series Stock number 1919720 RSDM 9061 1919721 RSDM 9060 EN rspro com...
Страница 2: ......
Страница 5: ...5 Dimensions 409 Declaration of Conformity 410...
Страница 23: ...23 DC AC 3 15A Input Current Fuse Holds the current fuse T3 15A 500V 5 20mm For fuse replacement details...
Страница 141: ...141 EOM pulse width timing...
Страница 160: ...160 rspro com SYSTEM FIRMWARE View System Info 161 Firmware Update 162...
Страница 211: ...211 SCREENHOT LOG Capture 212 Save Reading 215...
Страница 219: ...219 DISPLAY SETTING Digit 220 Display 222 Number 222 Bar Meter 223 Trend Chart 227 Histogram 238...
Страница 409: ...409 Dimensions All dimensions are shown in millimeters...
Страница 412: ...412 rspro com...