[SENSe:]CURRent:AC:NULL:VALue?
Returns the current relative value of AC Current measurement.
[SENSe:]CURRent:AC:NULL:VALue:AUTO
Sets the relative value auto on/off for AC Current
measurement.
Parameter: 0 | 1 | ON | OFF
Example: SENS:CURR:AC:NULL:STAT ON
SENS:CURR:AC:NULL:VAL:AUTO ON
The unit automatically sets the 1st count of measurement as
null value.
[SENSe:]CURRent:AC:NULL:VALue:AUTO?
Returns the null value auto state of AC Current measurement.
[SENSe:]CURRent:AC:RANGe
Sets the AC current measurement range.
Parameter: Range(<NRf> | MIN | MAX | DEF)
Example: SENS:CURR:AC:RANG 10e-3
Sets the AC current range to 10mA.
[SENSe:]CURRent:AC:RANGe?
Returns the AC current measurement range.
[SENSe:]CURRent:AC:RANGe:AUTO
Sets the AC current Auto-range settings on, off or once.
Parameter: 0 | 1 | ON | OFF | ONCE
Example: SENS:CURR:AC:RANG:AUTO OFF
Turns Auto-range off for AC current measurements.
[SENSe:]CURRent:AC:RANGe:AUTO?
Returns the AC current Auto-range settings.
Return parameter: 0 | 1, 0=OFF, 1=ON
[SENSe:]CURRent:AC:TERMinals
Assigns an input port for the current function.
Parameter: <NR1> RSDM-9060 : 3 / RSDM-9061 : 3 | 10
Example: SENS:CURR:AC:TERM 10
Sets the input jack to the 10A current input port.
[SENSe:]CURRent:AC:TERMinals?
Returns the assigned input port used for the current function.
Return Parameter: +3 | +10
[SENSe:]CURRent:AC:TRIGger:DELay
Sets the trigger delay time that minimum step is microseconds
of AC Current measurement.
Parameter: <NRf>(0 ~ 3600 s) | MIN | MAX | DEF
Example: SENS:CURR:AC:TRIG:DEL 1
Sets the trigger delay time to 1s of AC Current measurement.
Содержание RSDM-9060
Страница 1: ...User Manual RSDM 906X series Stock number 1919720 RSDM 9061 1919721 RSDM 9060 EN rspro com...
Страница 2: ......
Страница 5: ...5 Dimensions 409 Declaration of Conformity 410...
Страница 23: ...23 DC AC 3 15A Input Current Fuse Holds the current fuse T3 15A 500V 5 20mm For fuse replacement details...
Страница 141: ...141 EOM pulse width timing...
Страница 160: ...160 rspro com SYSTEM FIRMWARE View System Info 161 Firmware Update 162...
Страница 211: ...211 SCREENHOT LOG Capture 212 Save Reading 215...
Страница 219: ...219 DISPLAY SETTING Digit 220 Display 222 Number 222 Bar Meter 223 Trend Chart 227 Histogram 238...
Страница 409: ...409 Dimensions All dimensions are shown in millimeters...
Страница 412: ...412 rspro com...