RS PRO RSDM-9060 Скачать руководство пользователя страница 342

 

 

 

342 

rspro.com

 

[SENSe:]FREQuency:NULL[:STATe]? 

Returns the relative function state of Frequency measurement. 
Return parameter: 0 | 1, 0=OFF, 1=ON 

[SENSe:]FREQuency:NULL:VALue 

Sets the relative value for Frequency measurement. 
Parameter: <NRf> (-1.2e6~1.2e6 Hz) | MIN | MAX | DEF 
Example: SENS:FREQ:NULL:VAL 10 
Sets the relative value to 10Hz for Frequency measurement. 

[SENSe:]FREQuency:NULL:VALue? 

Returns the current relative value of Frequency measurement. 

[SENSe:]FREQuency:NULL:VALue:AUTO 

Sets the relative value auto on/off for Frequency measurement. 
Parameter: 0 | 1 | ON | OFF 
Example: SENS:FREQ:NULL:STAT ON 
      SENS:FREQ:NULL:VAL:AUTO ON 
The unit automatically sets the 1st count of measurement as 
null value. 

[SENSe:]FREQuency:NULL:VALue:AUTO? 

Returns the null value auto state of Frequency measurement. 

[SENSe:]FREQuency:TIMeout:AUTO 

Assigns timeout time at the frequency measurement. 
Parameter: 0 | 1 | ON | OFF 
Example: SENS:FREQ:TIM:AUTO OFF 
Sets the timeout time at 1 seconds. 

[SENSe:]FREQuency:TIMeout:AUTO? 

Returns the assigned timeout time used for the frequency 
function. 
Return parameter: 0 | 1, 0:timeout time = 1 second, 1:timeout 
time is different in according with ac filter bandwith (gate 
time). 

[SENSe:]FREQuency:TRIGger:DELay 

Sets the trigger delay time that minimum step is microseconds 
of Frequency measurement. 
Parameter: <NRf> (0 ~ 3600 s) | MIN | MAX | DEF 
Example: SENS:FREQ:TRIG:DEL 0.5 
Sets the trigger delay time to 0.5s of Frequency measurement. 

[SENSe:]FREQuency:TRIGger:DELay? 

Returns the trigger delay time in seconds of Frequency 
measurement. 
Return parameter: <NRf> 

Содержание RSDM-9060

Страница 1: ...User Manual RSDM 906X series Stock number 1919720 RSDM 9061 1919721 RSDM 9060 EN rspro com...

Страница 2: ......

Страница 3: ...age Measurement 34 AC DC Current Measurement 42 2W 4W Resistance Measurement 47 Continuity Test 51 Diode Measurement 54 Frequency Period Measurement 57 Capacitance Measurement 63 Temperature Measureme...

Страница 4: ...e Update 162 MENU SETTING 165 Configure System 166 Configure Display 189 SCREENHOT LOG 211 Capture 212 Save Reading 215 DISPLAY SETTING 219 Digit 220 Display 222 REMOTE CONTROL 245 Configure Interface...

Страница 5: ...5 Dimensions 409 Declaration of Conformity 410...

Страница 6: ...r in this manual or on the instrument WARNING Warning Identifies conditions or practices that could result in injury or loss of life CAUTION Caution Identifies conditions or practices that could resul...

Страница 7: ...or at building installations Note below Do not disassemble the instrument unless you are qualified as service personnel Make sure that the Sense LO terminal to Input LO is limited to 2Vpk the Sense HI...

Страница 8: ...risk of electric shock may occur if switching the input switch when high voltage or current is present Fuse WARNING Fuse type T0 25A 100 120 VAC T0 125A 220 240 VAC Make sure the correct type of fuse...

Страница 9: ...Normally only non conductive pollution occurs Occasionally however a temporary conductivity caused by condensation must be expected Pollution degree 3 Conductive pollution occurs or dry non conductive...

Страница 10: ...ected to the terminal which is marked with the letter N or coloured Blue or Black The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or...

Страница 11: ...ase note the information in this manual was correct at the time of printing However as RS PRO continues to improve its products changes can occur at any time without notice Please see the RS PRO websi...

Страница 12: ...onse 300 kHz The fastest sampling rate 1k Readings sec RSDM 9060 10k Readings sec RSDM 9061 Internal memory 10k read memory RSDM 9060 100k read memory RSDM 9061 Data Logging to USB Features 6 2 1 digi...

Страница 13: ...mber Description CD ROM UM Software Driver 82DM 90610M01 Safety Instruction Sheet GTL 217 Test leads GTL 246 USB Cable USB 2 0 A B type 1200mm Optional Accessories Part number Description 11DM 9061010...

Страница 14: ...pro com Front Panel Overview 5 6 7 9 10 11 12 13 14 4 2 1 3 RSDM 9060 5 6 7 8 9 10 11 12 13 14 4 2 1 3 RSDM 9061 Item Description 1 ESC Escape Key 2 Print screen Data log Key 3 USB Host Port 4 Power S...

Страница 15: ...Key Single press to escape from current page Presses and holds the ESC key for 2 seconds to toggle between full display and simple display which conceals the status bar math display as well as additio...

Страница 16: ...rm setting Arrow Keys Presses the left or right arrow keys to move parameter cursor rightward or leftward per requirement Range Selection Keys Presses the Auto key to activate auto range mode whilst c...

Страница 17: ...upper 2 rows of measurement keys are used for basic RSDM 9060 9061 measurements such as voltage current resistance continuity diode frequency period capacitance and temperature Each key has a primary...

Страница 18: ...Frequency Measures Frequency Shift FREQ Capacitance Measures Capacitance TEMP Temperature Measures Temperature Measurement Keys advanced Background The lower 2 rows of measurement keys are used for m...

Страница 19: ...tes the Trigger function Shift TRIG TRIG Manually sets the parameters for the Trigger function Menu Enters the setting pages in various Menus Shift Menu Filter Manually sets the parameters for the Fil...

Страница 20: ...Overview 240 1 3 4 5 7 8 9 10 RSDM 9060 240 1 2 3 4 5 6 7 8 9 10 11 12 RSDM 9061 Item Description 1 3A Current Terminal Fuse 2 3A Current Terminal RSDM 9061 only 3 RS 232 Interface Connector 4 USB Int...

Страница 21: ...ket Accepts the power cord AC 100 120 220 240V 10 50Hz 60Hz 400Hz 10 For power on sequence Fuse Socket Holds the main fuse 100 120 VAC T0 25A 220 240 VAC T0 125A For fuse replacement details RS 232C p...

Страница 22: ...e line in 4W resistance measurement Sense HI Terminal RSDM 9061 only Accepts HI sense line in 4W resistance measurement For details Input LO Terminal RSDM 9061 only Accepts ground COM line in all meas...

Страница 23: ...23 DC AC 3 15A Input Current Fuse Holds the current fuse T3 15A 500V 5 20mm For fuse replacement details...

Страница 24: ...ror icon for commands from remote control 4 Rear panel switch icon 5 Shift key identification icon 6 The first and second function menu switch icon 7 Digital I O mode icon User 4094 8 USB flash drive...

Страница 25: ...front panel for measurement Shift It indicates the shift key is being pressed ready for in conjunction with other keys for additional functions First function menu It indicates the active bottom menu...

Страница 26: ...ound Beep It indicates sound of beep is enabled Sound Key It indicates sound of key is enabled Sound All It indicates sounds of beep and key are both enabled Sound Off It indicates sounds of beep and...

Страница 27: ...rtical Applications Pull out the handle sideways and rotate it clockwise for the applications below Horizontal Tilt Place the unit horizontally Rotate the handle for tilt stand Vertical Place the hand...

Страница 28: ...er cord to the AC Voltage input Note Make sure the ground connector on the power cord is connected to a safety ground This will affect the measurement accuracy 3 Push the power button until click to t...

Страница 29: ...Select Resistance Range 48 General Resistance Setting 49 Continuity Test 51 Set Continuity Threshold 53 Diode Measurement 54 Frequency Period Measurement 57 Frequency Period In Depth Setting 60 Capac...

Страница 30: ...ced measurement mainly refers to the operation using the result obtained from one or more of the basic measurements Refresh Rate Background Refresh rate defines how frequently the RSDM 9060 9061 captu...

Страница 31: ...change the refresh rate You can also press the F2 Speed key to select a desired rate for measurement Press corresponding function key in accord with the desired option on screen display Also the F6 M...

Страница 32: ...ering Overview By default the RSDM 9060 9061 automatically triggers according to the refresh rate See the previous page for refresh rate setting details The TRIG key on the other hand is used to manua...

Страница 33: ...rigger Press and hold the TRIG key for 2 seconds to return to the Automatic Internal Trigger Press hold for 2 seconds Indicator Auto Internal Trigger Mode Note Single triggering is not supported for c...

Страница 34: ...spectively or ACV DCV mode display appears The mode will switch to ACV DCV mode immediately See the figure below for example DC or AC Voltage Indicates DC or AC Voltage mode 5 s Indicates the active r...

Страница 35: ...to select the range The Auto indicator turns to indicating Manual range selection If the appropriate range is unknown select the highest range You can also press the F1 Range key to select a range fo...

Страница 36: ...below ACV Press the F1 F3 key to select the desired rate F3 Auto Zero key to enable Auto Zero DCV mode only Background Autozero provides the most accurate measurements but requires additional time to...

Страница 37: ...e Background Specify the input impedance to the test leads Input R This specifies the measurement terminal input impedance which is either Auto or 10 M The Auto mode selects high impedance Hi Z for th...

Страница 38: ...ted the display shows an icon indicating the Auto mode is currently being activated F5 DCV Ratio key to enable DCV Ratio Background The RSDM 9060 9061 is able to calculate DCV ration by measuring inpu...

Страница 39: ...indicates the measured voltage from the Sense terminals Display DCV Ratio Reading Input Voltage Reading Reference Voltage Reading From the screenshot above for example the INP 00 86308V input voltage...

Страница 40: ...form Peak to Peak AC True RMS DC Sine PK PK 2 828 1 000 0 000 Rectified Sine full wave PK PK 1 414 0 435 0 900 Rectified Sine half wave PK PK 2 000 0 771 0 636 Square PK PK 2 000 1 000 0 000 Rectified...

Страница 41: ...0 voltage measurement will not result in error due to dynamic range limitations at full scale If the crest factor is more than 3 0 it usually indicates an abnormal waveform as seen from the below tabl...

Страница 42: ...terminals nor 10A terminal it offers merely a 3A terminal for current measurement less than 3A Current type RSDM 9060 AC DC 3A RSDM 9061 AC DC 3A 10A Activate ACI DCI Measure Press the Shift ACV or Sh...

Страница 43: ...use the 10A terminal Select Current Range Auto range To turn the automatic range selection On Off press the AUTO key The most appropriate range for the currently used input jack will be automatically...

Страница 44: ...measurement Press the F1 F5 key to select a desired range for the measurement Press the F6 More 1 2 key for next page with more options as the figure shown below Selectable Current Ranges Range Resolu...

Страница 45: ...enable Auto Zero DCI mode only Background Autozero provides the most accurate measurements but requires additional time to perform the zero measurement With autozero enabled On the RSDM 9060 9061 inte...

Страница 46: ...060 9061 measures the offset once and subtracts the offset from all subsequent measurements Display When turning On the Auto Zero the display shows an icon indicating the Auto Zero mode is currently b...

Страница 47: ...smaller than 1k Activate 2W or 4W Measurement Press the 2W key to activate 2W resistance measurement Press the Shift 2W key to activate 4W resistance measurement 2W 4W resistance mode display appears...

Страница 48: ...another sense leads between the SENSE LO and HI terminals Select Resistance Range Auto range To turn the automatic range selection On Off press the Auto key Manual range Press the or the key to select...

Страница 49: ...99M Note For more details see the specifications on page 353 General Resistance Setting F2 Speed key to select the rate Press the F1 F5 key to select the desired rate Press the F6 More 1 2 key for nex...

Страница 50: ...oltages present on the RSDM 9060 9061 s input circuitry from affecting measurement accuracy With autozero disabled Off the RSDM 9060 9061 measures the offset once and subtracts the offset from all sub...

Страница 51: ...e continuity testing Continuity mode display appears The mode will switch to continuity testing immediately Press on the front panel as figure shown below Continuity Indicates Continuity measurement 6...

Страница 52: ...he most accurate measurements but requires additional time to perform the zero measurement With autozero enabled On the RSDM 9060 9061 internally measures the offset following each measurement It then...

Страница 53: ...hreshold defines the maximum resistance allowed in the DUT when testing the continuity Threshold Range Threshold 1 to 1000 Default Threshold 10 Resolution 1 Procedure Press the F5 key to enter the Thr...

Страница 54: ...de test Press the Shift key to activate diode measurement Diode mode display appears The screen will switch to Diode mode immediately as the figure shown below Diode Indicates the Diode measurement 60...

Страница 55: ...des the most accurate measurements but requires additional time to perform the zero measurement With autozero enabled On the RSDM 9060 9061 internally measures the offset following each measurement It...

Страница 56: ...56 rspro com Display When turning On the Auto Zero the display shows an icon indicating the Auto Zero mode is currently being activated...

Страница 57: ...re menu Click the F1 Frequency key and the measured frequency will be displayed on the primary screen with the period value displayed on the sub section beneath To measure Period press the FREQ key fo...

Страница 58: ...mmediately Press on the front panel followed by clicking F3 key to choose Frequency as shown below Frequency Indicates Frequency measurement 100ms Indicates the active refresh rate Indicates Manual ra...

Страница 59: ...rms of voltage connect test leads between the Input HI terminal and the LO terminal The display updates the reading In terms of current connect test leads between the 3A terminal and the LO terminal o...

Страница 60: ...nter gate time menu Click the F1 F3 key for the desired gate time See the figure below with available options F4 InJack key to select voltage or current Background In accordance with the target inputs...

Страница 61: ...ilable options Note When selecting Auto the timeout setting will fully sync with the gate time value F1 AC Range key to manually select range setting Press the or the key to promptly select the range...

Страница 62: ...below When InJack is Voltage Press the F1 F6 key to select a desired range for the measurement When InJack is 3A Press the F1 F5 key to select a desired range for the measurement Press the F6 More 1 2...

Страница 63: ...itance mode display appears The screen will switch to capacitance mode immediately Press on the front panel as shown below Capacitance Indicates the Capacitance measurement 2 s Indicates the active re...

Страница 64: ...itance range is between 1nF and 10nF It is required to proceed to Cable Open function when capacitance is between 1nF and 10nF in which test leads connected will result in measuring capacity in small...

Страница 65: ...automatic range selection On Off press the Auto key Manual range Press the or the key to select the range The Auto indicator turns to indicating Manual range selection If the appropriate range is unk...

Страница 66: ...le 1nF 1pF 1 199nF 10nF 10pF 11 99nF 100nF 100pF 119 9nF 1 F 1nF 1 199 F 10 F 10nF 11 99 F 100 F 100nF 119 9 F Note For further details please see the specifications on page 355 Note The refresh rate...

Страница 67: ...put and calculates the temperature from the voltage fluctuation Temperature Range Thermocouple 200 C 1820 C vary by sensor types RTD 200 C 630 C Thermistor 80 C 150 C Activate temperature measurement...

Страница 68: ...kground Autozero provides the most accurate measurements but requires additional time to perform the zero measurement With autozero enabled On the RSDM 9060 9061 internally measures the offset followi...

Страница 69: ...it See the figure shown below Thermocouple Sensor Type Background The RSDM 9060 9061 accepts thermocouple inputs and calculates the temperature from the voltage difference of two dissimilar metals The...

Страница 70: ...ould be determined by the user Type Range Resolution SIM simulated 20 C 80 C 0 01 C The terminal temperature is manually defined by user Default value Auto Thermocouple Setting Procedure 1 Press the F...

Страница 71: ...ng the simulated baseline is 23 C If choosing Auto by F2 Auto key the subset menu appears with additional option Press the F3 ADJ 00 00 key followed by inputting a desired parameter as the following f...

Страница 72: ...Temperature Probe menu followed by clicking either the F2 RTD 2W or F3 RTD 4W key to activate RTD 2W 4W mode See the figure shown below 2 Press the F5 Type key to enter the Sensor Type menu as the fig...

Страница 73: ...lly as defined by the Callendar Van Dusen equation Type Coefficient Alpha Beta Delta PT100 0 00385 0 10863 1 49990 D100 0 00392 0 10630 1 49710 F100 0 00390 0 11000 1 49589 PT385 0 00385 0 11100 1 507...

Страница 74: ...o enter the desired value default 0 00385 range 0 9 999999 4 Press the F6 Enter key or the Knob key to confirm the input value and repeat the previous steps 2 4 to set up the Beta Delta and R0 coeffic...

Страница 75: ...ure Probe menu followed by clicking either the F4 Therm2W or F5 Therm4W key to activate Therm 2W 4W mode See the figure shown below 2 Press the F5 Type key to enter the Sensor Type menu as the figure...

Страница 76: ...or user to configure the A B and C coefficients individually as defined by the Steinhart Hart equation Type Coefficient A B C 2 2k 0 0014733 0 0002372 1 07E 07 5k 0 0012880 0 0002356 9 56E 08 10k 0 00...

Страница 77: ...s Number keys to enter the desired value A range 0 9 9999 default 1 2880E 03 7 Press the F6 Enter key or the Knob key to confirm the input value and repeat the previous steps 2 4 to set up the B and C...

Страница 78: ...78 rspro com DUAL MEASUREMENT Dual Measurement 79 Refresh Rate 82 Connect the Test Leads 83 The error influence on Dual Measurement V I 85 The error of current shunt 87...

Страница 79: ...displays such as ACV and frequency period measurements If the primary and secondary displays use different measurement functions ranges or rates then separate measurements will be taken for each disp...

Страница 80: ...ay to DCV measurement 2nd Measurement item setting To set a measurement mode for the second display press the F6 2ND key and the 2ND Function options appear subsequently For example press the F2 ACV k...

Страница 81: ...e covering either 1ST or 2ND icon indicates the active display 1 Select active display Toggle the active display between the 1ST and 2ND display by pressing the Knob key Primary display 1ST highlighte...

Страница 82: ...20 s 60 s 100 s 400 s 1k s 1 1 2k s 2 2 4k s 2 4 8k s 2 7 2k s 2 ACV ACI 1 s 5 s 20 s Frequency Period 1s 100ms 10ms Note 1 is applicable to RSDM 9060 whilst 2 is specifically for RSDM 9061 Selection...

Страница 83: ...ing indicator flashes according to the defined refresh rate setting of the active display Reading Indicator Connect the Test Leads Connect the test leads and measure When using the dual measurement fu...

Страница 84: ...eads and internal resistance of the current connection as it is in series with the test circuit The above measuring configuration is used to measure the voltage present on the resistance under test an...

Страница 85: ...voltage and current is being executed the route from DMM internal circuit to the LO terminal circuit for measuring voltage is totally identical with that for measuring current and thus the resistor w...

Страница 86: ...shunt will vary accordingly For example Vs 10V Rload 10 Vs 10V Rload 10 If the total impedance passing through current terminal is Rint 0 5 the ideal measured voltage will be 10V regardless of impact...

Страница 87: ...ing of voltage drop by shunt There will be obvious error occurred while measuring low current due to the measurable voltage generated by a larger shunt An ideal ammeter never changes flowing route of...

Страница 88: ...0 5 The theoretical value for current reading should be 1A in that the DMM internal resistor Rint which contains Shunt Rline Rline Rline Rline and Fuse will cause impact on the measuring reading The...

Страница 89: ...89 1 A 0 1 0 7 V 3 A 0 1 2 V 10 A 10m 0 5 V The above table indicates the maximum burden voltage caused by the maximum current within the applicable range...

Страница 90: ...rement 94 Trigger Setting 98 Automatic Single Triggering 98 Use External Trigger 99 Set Trigger Delay 101 Filter Setting 104 Digital Filter Overview 104 Digital Filter Setting 106 Math Measurement 109...

Страница 91: ...type of measurement which uses the result obtained by one of the basic measurements ACV DCV ACI DCI 2 4W Diode Continuity Frequency Period and Temperature Advanced Measurement Basic Measurement AC DC...

Страница 92: ...es of REL is to eliminate impedance of test lead from measurement Before operating impedance measurement short circuit the test lead followed by pressing the REL button For other measurements press th...

Страница 93: ...ress the Shift key followed by the REL key The setting appears First use function keys to decide unit value Then use the Left Right arrow keys to move cursor and scroll the Knob key or press Number ke...

Страница 94: ...t Press the Hold key to activate Hold measurement Hold measurement display appears Indicator Hold Measurement The Latest Hold Value Hold Indicates Hold measurement 000 6801 mVDC Shows the latest hold...

Страница 95: ...F4 key to select volume level Once the latest hold value is updated the beep sounds based on the defined volume Press the F1 key to set Beep volume off F2 MathDisp key to show STAT Math Press the F2...

Страница 96: ...dicates the maximum data value Average Indicates the mean average value Peak Peak Indicates the peak to peak data STDEV Indicates the standard deviation of the data Count Indicates the latest counts o...

Страница 97: ...he latest hold value Measure 000 7326mV Indicates the originally measured mV value 5 hold values in blue Indicates the latest 5 counts of hold values F6 HoldValue key to restart Press the F6 HoldValue...

Страница 98: ...SDM 9060 9061 triggers according to the refresh rate automatically See the previous page for refresh rate setting details The figure below shows the screen of Automatic Trigger measurement Auto Trigge...

Страница 99: ...er by default for example to count the frequency and the period Using an external trigger allows customized triggering conditions Signal connection Connect the external trigger signal to the Digital I...

Страница 100: ...cursor and scroll the Knob key or press Number keys to enter the desired counts 5 Push the Knob key Enter or press the F6 Enter key to confirm the input value Range 1 1 000 000 or Set Trigger Signal B...

Страница 101: ...ding to the external signal trigger timing Exit external trigger Press the F1 TrigSource key to reenter the TrigSource menu followed by pressing the F1 Auto or the F2 Single key to switch to other tri...

Страница 102: ...h to the manual delay time setting 4 Use F1 F3 keys to decide unit value Then use the Left Right arrow keys to move cursor and scroll the Knob key or press Number keys to enter the desired value 5 Pus...

Страница 103: ...103 2 Press the ESC key to return to the previous page and have the auto trigger delay setting take effect The 1ST display will be shown like the following figure...

Страница 104: ...The filter type defines the averaging method The following diagrams highlight the differences between the Moving and Repeating filter using 4 samples per reading Moving default The Moving filter take...

Страница 105: ...Range 2 100 Filter window Filter window defines the threshold for when the digital filter data is updated again When the AD data falls in the range between TH and TL the filter keeps processing When t...

Страница 106: ...r Setting Filter setting Press the Shift key Menu Filter key The Filter setting menu shows as the figure below Select display Press the F1 Setup key to toggle between the 1ST and the 2ND measurement t...

Страница 107: ...nu Use the Left Right arrow keys to move cursor and scroll Knob key or press Number keys to enter the desired value Press the F6 Enter key or the Knob key until click to confirm the filter count setti...

Страница 108: ...the subsequent menu Press the F1 F5 keys to choose desired Filter Window percentage Range 0 01 0 1 1 10 None Turn off Filter Press the Shift key the Menu Filter key Press the F2 Filter key to turn Of...

Страница 109: ...he specified upper high and lower low limit MX B Multiplies the reading X by the factor M and adds subtracts offset B 1 X Divides 1 by the reading X Percentage Runs the following equation ReadingX Ref...

Страница 110: ...Watt Vreading2 Rref Parameters Vreading Input Voltage ACV or DCV Rref REF Reference resistance simulating an output load Activate dBm Press the Shift key Math key to activate Math setting menu as the...

Страница 111: ...Knob key Enter or press the F6 Enter key to confirm the input reference resistance or Resistance List 2 4 8 16 50 75 93 110 124 125 135 150 250 300 500 600 800 900 1000 1200 8000 View result in WattW...

Страница 112: ...ollowing chapters Show STAT result Background The STAT page in MathDisp allows you to make statistical calculations for several measurements including Minimum Maximum Average Peak Peak Standard Deviat...

Страница 113: ...lculations for several parameters Operation Press the F3 Math key to show the mathematical analysis instantly as below View Data 00 00000 dBW Indicates the latest dBW value Measure 000 0006V Indicates...

Страница 114: ...0 9061 calculates the dBm using the reading at the first moment and stores it as dBmref Activate dB Press the Shift Math key to activate Math setting menu as the following figure shown Further press t...

Страница 115: ...300 500 600 800 900 1000 1200 8000 F4 Ref Method to select dB reference method Reference method involves the ways to calculate dB value When dBm option is selected user can specify a definite dBm val...

Страница 116: ...esired Ref value Press the F6 Enter key or Knob key to confirm the input value Note When setting voltage Ref value press the function keys to promptly define the unit F6 Ref Value key to set the dBm r...

Страница 117: ...mediately as the figure below View Data 10 28281 dB Indicates the calculated dB value Minimum Indicates the minimum data value Maximum Indicates the maximum data value Average Indicates the mean avera...

Страница 118: ...originally measured m Voltage value Ref 0002 Indicates the defined reference resistance value Ref Voltage 003 0330mV Indicates the measured reference voltage value Ref dBm 023 3727 Indicates the measu...

Страница 119: ...the Shift Math key to activate Math setting menu as the following figure shown Further press the F1 Function key to enter the Math Function menu as the figure shown below Press F4 Compare key to enabl...

Страница 120: ...low limit Press the F5 Low Limit key to enter the setting menu First use the functions keys to determine the unit which varies by different measure modes Then use the Left Right arrow keys to move cur...

Страница 121: ...t the intensity of beep volume via pressing F1 F3 key for desired level as the figure shown below or or Compare mode result When the measured result is within the range of high and low limit the displ...

Страница 122: ...relative pins of digital I O port in action are as the follows Digital I O FAIL Out Pin 6 and LOW Limit FAIL Out Pin 8 are activated Pass If the compare result is Pass the relative pin of digital I O...

Страница 123: ...e figure below View Data 000 4835 m VDC Indicates the currently measured mVDC value Minimum Indicates the minimum data value Maximum Indicates the maximum data value Average Indicates the mean average...

Страница 124: ...alue Low Limit Indicates the defined low limit Low Fail Indicates the counts of below the defined low limit High Limit Indicates the defined high limit High Fail Indicates the counts of above the defi...

Страница 125: ...entical to the contents of STAT display Refer to the previous chapter for details Red Section It is identical to the contents of Math display Refer to the previous chapter for details Compare live res...

Страница 126: ...rement To cancel the Compare measurement press the F1 Function key followed by clicking F1 OFF to deactivate or simply activate another measurement MX B Measurement Applicable to Activate MX B Press t...

Страница 127: ...he Knob key or press Number keys to enter the desired value See the figure below Press the F6 Enter key or the Knob key until click to confirm the input M value or F4 B Value key to set the offset B P...

Страница 128: ...he following chapters Show STAT result Background The STAT page in MathDisp allows you to make statistical calculations for several measurements including Minimum Maximum Average Peak Peak Standard De...

Страница 129: ...round The Math page in MathDisp allows you to view mathematical calculations for several parameters Operation Press the F3 Math key to show the mathematical analysis instantly as below View Data 0 999...

Страница 130: ...measurement 1 X Measurement Applicable to Activate 1 X Press the Shift Math key to activate Math setting menu as the following figure shown Further press the F1 Function key to enter the Math Function...

Страница 131: ...r F3 Math display in accord with the following chapters Show STAT result Background The STAT page in MathDisp allows you to make statistical calculations for several measurements including Minimum Max...

Страница 132: ...ult Background The Math page in MathDisp allows you to view mathematical calculations for several parameters Operation Press the F3 Math key to show the mathematical analysis as the figure below View...

Страница 133: ...following figure shown Further press the F1 Function key to enter the Math Function menu as the figure shown below Press F6 More 1 2 key to enter the next page followed by pressing the F2 Percent key...

Страница 134: ...ey Enter or press the F6 Enter key to confirm the input value or F2 MathDisp key to show STAT Math Press the F2 MathDisp key to show the option menu as the figure below shown Proceed to the F2 STAT or...

Страница 135: ...ean average value Peak Peak Indicates the peak to peak data STDEV Indicates the standard deviation of the data Count Indicates the latest counts of Percent Show Math result Background The Math page in...

Страница 136: ...886mV Indicates the originally measured m Voltage value Ref 000 5908m Indicates the defined reference value Deactivate percent measurement To cancel the percent measurement press the F1 Function key f...

Страница 137: ...O Digital I O Overview 138 Application Compare Mode 142 Application 4094 User Mode 150 User Mode IO Output Mode 150 User Mode Switch Mode LED 152 User Mode Switch Mode Relay 154 4094 Mode 156 Applicat...

Страница 138: ...a secondary function 4094 Mode and third function User Mode the Digital I O port can have the output state of pins 5 8 controlled via remote control By providing separate VCC power for the terminal th...

Страница 139: ...lock OUT1 6 Fail Out Output Enable OUT2 7 High Limit Fail Out Strobe OUT3 8 Low Limit Fail Out Serial Input OUT4 9 EOM Out EOM Out EOM Out Pin1 VCC output 5V Serves as the unregulated max power source...

Страница 140: ...ser for diversified functions as follows Compare 4094 User Mdoe Refer to the page 14 for details of Compare Mode and the page 20 for details of 4094 User Mode Pins 5 8 output wiring diagram Pin9 EOM E...

Страница 141: ...141 EOM pulse width timing...

Страница 142: ...shold or the low threshold the High Fail or Low Fail pin is pulled low When the signal stays within the threshold levels the Pass pin is pulled low Pin Assignment Pin No Compare Mode Description 1 VCC...

Страница 143: ...e setting menu First use the functions keys to determine the unit which varies by different measure modes Then use the Left Right arrow keys to move cursor and scroll the Knob key or press Number Keys...

Страница 144: ...ling beep mode user can be aware of the latest state promptly by beep voice The display shows as the figure below Press the F2 Pass or F3 Fail key to determine the condition of beep alarm or Press the...

Страница 145: ...low for more details of each state in compare mode High If the compare result is High the relative pins of digital I O port in action are as the follows Digital I O FAIL Out Pin 6 and HIGH Limit FAIL...

Страница 146: ...STAT display in accord with the following chapters Show STAT result Background The STAT page in MathDisp allows you to make statistical calculations for several measurements including Minimum Maximum...

Страница 147: ...ath key to show the mathematical analysis as the figure below View Data 000 5625m VDC Indicates the currently measured mVDC value Low Limit Indicates the defined low limit Low Fail Indicates the count...

Страница 148: ...contents of STAT display Refer to the previous chapter for details Red Section It is identical to the contents of Math display Refer to the previous chapter for details Compare live result in MathDisp...

Страница 149: ...mpare function is activated Low limit High limit P P P F F H L Pin 5 PASS Out Pin 6 FAIL Out Pin 7 HIGH Limit FAIL Out Pin 8 LOW Limit FAIL Out Pin 5 8 Output Compare On Deactivate Compare measurement...

Страница 150: ...ing output as general IO Output usage with up to 4 pins available for use simultaneously Refer to the following introductions along with diagrams for more details Please see the digital I O commands o...

Страница 151: ...151 Pin Diagram Use the built in power supply Note Pin1 and Pin2 Not in use Use in conjunction with the logic gate Note Pin2 Not in use...

Страница 152: ...agrams for more details Please see the digital I O commands on page 259 for full usage details Related Commands DIG INT MODE IO switch to IO mode DIG INT DATA SET 1 0 0 1 OUT1 Pin5 LED OFF OUT2 Pin6 L...

Страница 153: ...153 Pin Diagram Use the built in power supply Note Pin2 and Pin3 Not in use Use the external power Note Pin1 and Pin2 Not in use...

Страница 154: ...more details Please see the digital I O commands on page 259 for full usage details Related Commands DIG INT MODE IO switch to IO mode DIG INT DATA SET 1 0 1 0 OUT1 Pin5 RELAY OFF OUT2 Pin6 RELAY ON...

Страница 155: ...155 Pin Diagram Use the built in power supply which provides the power of maximum 100mA Note Pin3 Not in use Use the external power 5 24V Maximum Ids of each channel 400mA Note Connect Pin2 to Ext Vcc...

Страница 156: ...Please see the digital I O commands on page 259 for full usage details Related Commands DIG INT MODE 4094 switch to 4094 mode 4094 x 1 8 Pin DIG INT DATA OUTP 10 1 4094 Output Out1 Out8 01010000 4094...

Страница 157: ...157 8 Serial Input Use Pin Diagram Use the built in power supply Note Pin2 Not in use Use the external power Note Pin1 and Pin2 Not in use Method of series Note Pin1 and Pin2 Not in use...

Страница 158: ...he RSDM 9060 9061 when the RSDM 9060 9061 is in the external trigger mode See page xxx for details Signal connection Connect the external trigger signal to the Digital I O port located on the rear pan...

Страница 159: ...ading indicator does not flash before triggering can be on or off After triggering the indicator flashes according to the external signal trigger timing Exit external trigger Press the Shift key follo...

Страница 160: ...160 rspro com SYSTEM FIRMWARE View System Info 161 Firmware Update 162...

Страница 161: ...re and Slave Firmware Step 1 Press the Menu key the System configuration menu appears And press the NEXT key repeatedly or scroll the Knob key to move to the Security Info SystemInfo field 2 Press the...

Страница 162: ...updating the latest firmware Step 1 Press the Menu key the System configuration menu appears And press the NEXT key repeatedly or scroll the Knob key to move to the Cali Update Firmware field Press th...

Страница 163: ...o user can check the current Master and Slave firmware version respectively in this menu Note Prior to update please rename the downloaded firmware files as below Master file M_IMAGE bin Slave file S_...

Страница 164: ...164 rspro com 2 Press the NEXT key or scroll Knob key to move to the Update followed by pressing the F5 Enter key or Knob key to Start update...

Страница 165: ...ng 172 Calibration Setting 179 Firmware Update 183 Security Setting 185 View System Info 187 Configure Display 189 Brightness Setting 189 Auto Off Setting 190 Auto Off Time Setting 192 1ST Color Setti...

Страница 166: ...m Beep Setting Background Enable or Disable Beep Sound Step 1 Press the Menu key the System configuration menu appears 2 Press the F5 Enter key or Knob key followed by scrolling Knob key or pressing k...

Страница 167: ...Enable or Disable Key Sound Step 1 Press the Menu key the System configuration menu appears And then press the NEXT key repeatedly or scroll the Knob key to move to the Key Sound field 2 Press the F5...

Страница 168: ...for Key Sound Date Setting Background Manually adjust date for system or automatically set date via TimeSync setting Step 1 Press the Menu key the System configuration menu appears And press the NEXT...

Страница 169: ...it 3 Press the F5 Enter key or Knob key to confirm the input digit for year of Date 4 Repeat steps 2 to 3 for month and day Time Setting Background Manually adjust time for system or automatically set...

Страница 170: ...ollowed by scrolling Knob key or pressing keys to define hour of Time Also you can press Number keys to directly input a specific digit 3 Press the F5 Enter key or Knob key to confirm the input digit...

Страница 171: ...internet with appropriate network setting Step 1 Press the Menu key the System configuration menu appears And press the NEXT key repeatedly or scroll the Knob key to move to the Date Time TimeSync fi...

Страница 172: ...ime from the remote sever Synchronize Time Define an interval to retrieve the currently standard time from the remote sever 7 Days 14 Days 30 Days Set the time zone Set UTC Coordinated Universal Time...

Страница 173: ...Menu key the System configuration menu appears And press the NEXT key repeatedly or scroll the Knob key to move to the Parameter Save Load field 2 Press the F5 Enter key or Knob key to enter the Para...

Страница 174: ...p 1 Press the F5 Enter key or Knob key to open the dropdown menu 2 Scroll the Knob key or pressing keys followed by pressing the F5 Enter key or Knob key to confirm the group selection Note 1 Press th...

Страница 175: ...Press the F2 Backspace key to clear default words 3 Use the Left Right and keys or scroll the Knob key to move the cursor to desired word followed by pressing the F5 Input key or Knob key to input the...

Страница 176: ...4 Press the F4 OK or the Knob key to confirm the input words Enter 5 Press the F5 Enter key or Knob key to saved the input words Load Select a Group 1 Press the F5 Enter key or Knob key to open the d...

Страница 177: ...s keys followed by pressing the F5 Enter key or Knob key to confirm the group selection Note 1 The currently selected group name appears in the Note field Select a Action 1 Press the F5 Enter key or K...

Страница 178: ...s keys followed by pressing the F5 Enter key or Knob key to confirm the action selection 3 Press the F5 Enter key or Knob key to confirm the action selection Parameter None no recall action Power On r...

Страница 179: ...echnician can operate the calibration procedure Refer to the qualified personnel for more details when necessary Step 1 Press the Menu key the System configuration menu appears And press the NEXT key...

Страница 180: ...coefficient default Factory calibration value Check the box to enable Frequency Original Frequency x Compensate Coefficient Uncheck the box to disable Frequency Original Frequency FREQ Cali Mathod Se...

Страница 181: ...ng the F5 Enter key to save the frequency compensation coefficient The value changes as the figure shown below Please Input 1kHz Source Start the frequency compensation coefficient calculation and con...

Страница 182: ...on which is an internally self calibration function that does Not require external signal source It corrects the gain of internal amplifier though it is not necessary for general conditions unless the...

Страница 183: ...ng the latest firmware Step 1 Press the Menu key the System configuration menu appears And press the NEXT key repeatedly or scroll the Knob key to move to the Cali Update Firmware field 2 Press the F5...

Страница 184: ...n the flash drive plugged into the USB port on the front panel Also user can check the current Master and Slave firmware version respectively in this menu 1 Press the F5 Enter key or Knob key first th...

Страница 185: ...key or Knob key to Start update Security Setting Background This section is to change the password and enable or disable Lan password Step 1 Press the Menu key the System configuration menu appears A...

Страница 186: ...to enter the Please Input Password page 3 Use the Left Right and keys or scroll the Knob key to move the cursor followed by pressing the F5 Input key or the Knob key to input the password 4 Press the...

Страница 187: ...onfirm Password Enter the new password again Modify Password Change password by clicking Start View System Info Background View system information including Vendor Model Name Serial Number Master Firm...

Страница 188: ...188 rspro com 2 Press the F5 Enter key or Knob key to enter the System Information where all the critical contents are exposed for check...

Страница 189: ...Press the Menu key followed by pressing the Page Down key repeatedly until the Display configuration menu appears 2 Use the Left Right keys to move the cursor followed by scrolling Knob key or pressi...

Страница 190: ...to confirm the input digit for backlight brightness Auto Off Setting Background Enable or disable automatic brightness adjustment Step 1 Press the Menu key followed by pressing the Page Down key repea...

Страница 191: ...croll the Knob key to move to the BackLight AutoOff field 3 Press the F5 Enter key or Knob key followed by scrolling Knob key or pressing keys to select the ON option 4 Press the F5 Enter key or Knob...

Страница 192: ...the screen will change to automatic brightness adjustment Note Auto Off Time will be activated only when Auto Off option is turned ON Step 1 Press the Menu key followed by pressing the Page Down key...

Страница 193: ...move the cursor followed by scrolling Knob key or pressing keys to define minutes Also you can press Number keys to directly input a specific minutes 4 Press the F5 Enter key or Knob key to confirm th...

Страница 194: ...he theme color of 1ST display Step 1 Press the Menu key followed by pressing the Page Down key repeatedly until the Display configuration menu appears 2 Press the NEXT key repeatedly or scroll the Kno...

Страница 195: ...llowed by scrolling Knob key or pressing keys to select desired color for 1ST display 4 Press the F5 Enter key or Knob key to confirm the selected color Display result The following figure demonstrate...

Страница 196: ...he theme color of 2ND display Step 1 Press the Menu key followed by pressing the Page Down key repeatedly until the Display configuration menu appears 2 Press the NEXT key repeatedly or scroll the Kno...

Страница 197: ...ollowed by scrolling Knob key or pressing keys to select desired color for 2ND display 4 Press the F5 Enter key or Knob key to confirm the selected color Display result The following figure demonstrat...

Страница 198: ...e theme color of Math functions Step 1 Press the Menu key followed by pressing the Page Down key repeatedly until the Display configuration menu appears 2 Press the NEXT key repeatedly or scroll the K...

Страница 199: ...llowed by scrolling Knob key or pressing keys to select desired color for Math display 4 Press the F5 Enter key or Knob key to confirm the selected color Display result The following figure demonstrat...

Страница 200: ...ned text is shown in the 1ST display only when MathDisp is off Step 1 Press the Menu key followed by pressing the Page Down key repeatedly until the Display configuration menu appears 2 Press the NEXT...

Страница 201: ...y or Knob key followed by scrolling Knob key or pressing keys to select the Time option 2 Press the F5 Enter key or Knob key to confirm the Time option Display result The following figure demonstrates...

Страница 202: ...02 rspro com Text display 1 Press the F5 Enter key or Knob key followed by scrolling Knob key or pressing keys to select the Text option 2 Press the F5 Enter key or Knob key to confirm the Text option...

Страница 203: ...203 3 Press the NEXT key or scroll the Knob key to move to the Math Off Display Edit field 4 Press the F5 Enter key or Knob key to enter the KeyBoard page...

Страница 204: ...e Knob key to move the cursor followed by pressing the F5 Input key or Knob key to input desired words Note F1 Caps Lock key is for high and low case shift 6 Press the F4 OK key to confirm the input w...

Страница 205: ...ailable for up to 1 2k s refresh rate The 2 4k s above refresh rates are Not supported by anti aliasing Note When Auto Zero or dual measure mode both of which lower down computing speed is activated a...

Страница 206: ...or scroll the Knob key to move to the BigFont Option Antialiasing field 3 Press the F5 Enter key or Knob key followed by scrolling Knob key or pressing keys to select the ON option 4 Press the F5 Ente...

Страница 207: ...he additional information display Step 1 Press the Menu key followed by pressing the Page Down key repeatedly until the Display configuration menu appears 2 Press the NEXT key repeatedly or scroll the...

Страница 208: ...the Return option followed by pressing the F5 Enter key or Knob key to have the setting take effect Display result Take the Temperature mode for example as the figure below we can clearly recognize t...

Страница 209: ...for user interface display Step 1 Press the Menu key followed by pressing the Page Down key repeatedly until the Display configuration menu appears 2 Press the NEXT key repeatedly or scroll the Knob...

Страница 210: ...key to select one of the language options Move to the Return option followed by pressing the F5 Enter key or Knob key to have the setting take effect Options English Traditional Chinese Simplified Chi...

Страница 211: ...211 SCREENHOT LOG Capture 212 Save Reading 215...

Страница 212: ...he LOG LOG key and the following menu appears 2 Press the F1 Log Mode key followed by clicking the F1 Capture key to enable the Capture mode for screenshot 3 Press the F2 FileName key to enter the Log...

Страница 213: ...the cursor followed by pressing the F5 Input key or Knob key to input desired words The F1 Caps Lock key is for high and low case shift 5 Press the F4 OK key to confirm the input words Note It is onl...

Страница 214: ...ed file click F1 Yes to overwrite whilst click F2 No to save in a non occupied serial number of file name Click ESC key to simply discard the overwrite action For File Name Manual Under Overwrite Alwa...

Страница 215: ...Press the F2 FileName key to enter the Log FileName Mode menu Further press the F1 Default key to let system saves screenshot by auto name in serial number or press the F2 Manual key to determine fil...

Страница 216: ...move the cursor followed by pressing the F5 Input key or Knob key to input desired words The F1 Caps Lock key is for high and low case shift 5 Press the F4 OK key to confirm the input words Note It i...

Страница 217: ...le click F1 Yes to overwrite whilst click F2 No to save in a non occupied serial number of file name Click ESC key to simply discard the overwrite action For File Name Manual Under Overwrite Always mo...

Страница 218: ...urce to save and read Determine the source mode by further pressing the F1 Count key or the F2 Recent key Count indicates the saved data log contains the total counts of measurement whilst Recent repr...

Страница 219: ...219 DISPLAY SETTING Digit 220 Display 222 Number 222 Bar Meter 223 Trend Chart 227 Histogram 238...

Страница 220: ...by clicking the F1 Digit key the Digit menu appears where several digit options are available to select 2 Further press F1 6 F2 5 F3 4 keys for desired maximum digit numbers on display or press the F1...

Страница 221: ...5 1 2 4 1 2 4 1 2 4 1 2 ACV 6 1 2 5 1 2 4 1 2 DCI 6 1 2 6 1 2 6 1 2 6 1 2 5 1 2 5 1 2 5 1 2 5 1 2 4 1 2 4 1 2 4 1 2 ACI 6 1 2 5 1 2 4 1 2 2W 4W 6 1 2 6 1 2 6 1 2 6 1 2 5 1 2 5 1 2 5 1 2 5 1 2 4 1 2 4...

Страница 222: ...y key the Display menu appears where several display options are available for selection 2 Press the F1 Number key the screen shows the Number mode for measurement display The measured value is presen...

Страница 223: ...een shows the Bar Meter mode for measurement display The measured value is presented in the bar meter way for viewing along with the maximum digits display depending on the Digit selection Display Red...

Страница 224: ...ints of 2 sides are 0 and 100m respectively which perfectly correspond to the specified range of measurement Green Sect The currently measured value Manual Selecting Manual allows user to customize th...

Страница 225: ...user specified range for measurement Green Sect The currently measured value Center When Center is selected it is available to further determine the exact Center value and the Span Scale for the mete...

Страница 226: ...000k in F5 key is identical with the upper value in red frame on the low scale of meter bar Green Sect The specified High Scale 10 000 in F6 key is identical with the upper value in green frame on hi...

Страница 227: ...igure shown Trend Chart Background Shift to the Trend Chart display for each measurement Step 1 Press the DISP key followed by clicking the F2 Display key the Display menu appears where several displa...

Страница 228: ...F3 VScale key to define vertical scale Background Press the F3 Scale key to enter the VScale Setup menu where Normal and Manual options are available for selection Normal Selecting Normal allows the v...

Страница 229: ...the lowest scale 9 0000 correspond to the user specified values in the red section Green Sect The currently measured value Purple Sect Press the F4 Auto Once key to obtain the highest and lowest scal...

Страница 230: ...scale of trend chart to be symmetric with the defined refresh rate of measurement Green Sect The user defined refresh rate Red Sect The refreshing frequency of the total counts of measurements is con...

Страница 231: ...e right side 0 to the left hand 400 0s that corresponds to the user specified range of horizontal scale Yellow Sect The currently total counts of measurement Orange Sect The user specified auto stop f...

Страница 232: ...e To put it simply due to the maximum 400 counts at once when setting 400Sec the interval is equal to 400Sec divided by 400 counts 1 second If setting 800Sec it turns out 800 400 2 seconds Blue Sect P...

Страница 233: ...ed Sect Press the F1 Range for range check Green Sect The total counts of measurements before entering the Stop View Yellow Sect Press the Knob key to change the maximum counts for display 1 pixel 4 p...

Страница 234: ...or decrease 40 counts per time F2 key Cursor1 Press the F2 Cursor1 key to check the lowest value of each count Scroll the Knob key rightward or leftward to move cursor on different sections Green Sect...

Страница 235: ...ghest value of the selected count with its affiliated serial number Yellow Sect Press the Knob key to change the maximum counts for display 1 pixel 10 pixels 20 pixels Purple Sect The delta between th...

Страница 236: ...F1 Range F2 Cursor1 and F3 Cursor2 keys Refer to the each section for further details TIM Press the F4 COU TIM key to toggle between the 2 modes COU TIM In association with the previous F1 Range F2 Cu...

Страница 237: ...llowed by entering this mode where the statistics are almost identical to those of the previous single measurement Display User can view each data for dual measurements or toggle on off for either 1ST...

Страница 238: ...ress the F4 Histogram key the screen shows the Histogram mode for measurement display The measured value is presented in the way of histogram for viewing along with the maximum digits display dependin...

Страница 239: ...ay F3 Bins key to define bin numbers Background Press the F3 Bins key to enter the Bins Setup menu where user can customize the maximum numbers of stripe like bins for display Note The available optio...

Страница 240: ...5 s 2 4k s 4 8k s 7 2k s 10k s Max Bin Number 400 200 100 20 F4 HScale key to define horizontal scale Background Press the F4 HScale key to enter the HScale Setup menu where Auto and Manual options a...

Страница 241: ...Sect The range of horizontal scale of histogram display varies according to the currently measured value Manual Selecting Manual allows the horizontal scale of histogram display to be customized by us...

Страница 242: ...e Take the figure below for instance the right and left ends in horizontal scale are irregular values 1 1022 and 1 0740 which come from the latest measurement of bins F5 Stop View key for data Press t...

Страница 243: ...ghtward or leftward to change bin number for checking Yellow Sect It indicates the total accumulated counts of measurement categorized within the selected bin number Orange Sect It indicates the exact...

Страница 244: ...ue section Up F6 Start key to restart measurement After entering the View Setup His mode system will halt the measurement right away Exit the View Setup His mode and press the F6 Start key to restart...

Страница 245: ...andshake 259 Set the EOL Character 259 Set the Separation Character 259 Insert GPIB Card Optional 259 Configure GPIB Interface 260 Activate Ethernet Interface 264 Reboot LAN Setup 265 Configure Ethern...

Страница 246: ...Local control mode front panel operation press the Shift key Configure SCPI ID Setting Background The IDN query returns the manufacturer model number serial number and system firmware version number...

Страница 247: ...b key to move to the SCPI ID field 3 Press the F5 Enter key or Knob key followed by scrolling Knob key or pressing keys to land on the desired SCPI ID Identity option 4 Press the F5 Enter key or Knob...

Страница 248: ...st Unit side connector Real panel Type B device USB Speed 2 0 Full speed Steps 1 Press the Menu key and then the Page Down key repeatedly until the Interface configuration menu appears 2 Press the F5...

Страница 249: ...select the USB option 4 Press the F4 NEXT key repeatedly or scroll the Knob key to move to the USB Protocol field 5 Press the F5 Enter key or Knob key followed by scrolling Knob key or pressing keys...

Страница 250: ...tilizing the CDC or TMC USB class install the appropriate CDC or TMC USB driver included on the User Manual CD USBCDC The USB port on the RSDM 9060 9061 will appear as a virtual COM port to a connecte...

Страница 251: ...19200 9600 Data bits 8 Parity none Stop bits 1 Flow control none RTS CTS DTR DSR Step 1 Press the Menu key and then the Page Down key repeatedly until the Interface configuration menu appears 2 Press...

Страница 252: ...o select the RS232 option 4 Press the F4 NEXT key repeatedly or scroll the Knob key to move to the RS232 Baud Rate field 5 Press the F5 Enter key or Knob key followed by scrolling Knob key or pressing...

Страница 253: ...the desired RS232 Baud Rate option 7 Press the F4 NEXT key repeatedly or scroll the Knob key to move to the RS232 FlowCtrl field 8 Press the F5 Enter key or Knob key followed by scrolling Knob key or...

Страница 254: ...esired RS232 FlowCtrl option 10 Press the F4 NEXT key repeatedly or scroll the Knob key to move to the RS232 TX Term EndOfLine field 11 Press the F5 Enter key or Knob key followed by scrolling Knob ke...

Страница 255: ...r Knob key again to confirm the desired RS232 TX Term EndOfLine option Note GPIB USBTMC and LAN are fixed with only LF option 13 Press the F4 NEXT key repeatedly or scroll the Knob key to move to the...

Страница 256: ...er key or Knob key again to confirm the desired RS232 TX Term Separation option Note GPIB USBTMC and LAN are fixed with only COMMA option 16 Connect the RS232 cable to the rear panel terminal RS232 Pi...

Страница 257: ...ified factories RS 232 Connection RSDM 906X provides the complete RS 232 signals control Select the corresponding null modem cable which has the DB 9 female connectors on both ends when the port of DB...

Страница 258: ...sly link the pin 7 RTS to pin 8 CTS for advanced function of GTL 234 More Connections If more other cables are applied the diagram of full connections is illustrated as the following figure where the...

Страница 259: ...d from the PC include CR LF LF CR CR or LF The most common EOL character is CR LF Note The USBTMC GPIB and LAN s EOL character is fixed with LF EOL CR LF LF CR CR LF default CR LF Set the Separation C...

Страница 260: ...ater reuse Insert the GPIB card Insert the GPIB card into the slot Close the cover by tightening the screws Power On Connect the power cord and turn On the power Configure GPIB Interface GPIB Configur...

Страница 261: ...ss the F5 Enter key or Knob key followed by scrolling Knob key or pressing keys to land on the GPIB option Note As the figure shown below GPIB option won t appear when optional GPIB card is not well i...

Страница 262: ...or Knob key to select the GPIB option 4 Press the F4 NEXT key repeatedly or scroll the Knob key to move to the GPIB Address field Note As the figure shown below the GPIB Address field won t appear wh...

Страница 263: ...c digit 6 Press the F5 Enter key or Knob key again to confirm the input digit for GPIB Address 7 Connect the GPIB cable to the rear panel optional communication port after the GPIB card has been insta...

Страница 264: ...12 SHIELD Ground 24 Single GND Activate Ethernet Interface Overview Speed 10BaseT 100BaseTx Ethernet LAN port activation 1 Press the Menu key and then the Page Down key repeatedly until the Interface...

Страница 265: ...LAN Setup Background To reboot is used to reset the Ethernet configuration when new settings have been made When the Lan Setup settings have been edited reboot to validate the changes and reset the Et...

Страница 266: ...and other configuration parameters automatically assigned by a DHCP server DHCP Configuration 1 Press the Menu key and then the Page Down key repeatedly until the Lan Setup configuration menu appears...

Страница 267: ...ddresses including the subnet mask gateway DNS1 and DNS2 Note The IP Address Setup can only be edited if DHCP is off IP Address Configuration 1 Press the Menu key and then the Page Down key repeatedly...

Страница 268: ...IP Address Also you can press Number keys to directly input a specific digit 4 Press the F5 Enter key or Knob key to confirm the input digit for IP1 Address And the cursor will automatically jump to...

Страница 269: ...eatedly or scroll the Knob key to move to the IP Address Setup Subnet field 7 Use the Left Right keys to move the cursor followed by scrolling Knob key or pressing keys to define Subnet Also you can p...

Страница 270: ...S2 S3 and S4 Note The Subnet is divided in 4 groups S1 S2 S3 S4 Gateway Configuration 10 Press the F4 NEXT key repeatedly or scroll the Knob key to move to the IP Address Setup Gateway field 11 Use t...

Страница 271: ...G1 And the cursor will automatically jump to next groups 13 Repeat steps 11 to 12 for G2 G3 and G4 Note The Gateway is divided in 4 groups G1 G2 G3 G4 DNS1 Configuration 14 Press the F4 NEXT key repe...

Страница 272: ...a specific digit 16 Press the F5 Enter key or Knob key again to confirm the input digit for D11 And the cursor will automatically jump to next groups 17 Repeat steps 15 to 16 for D12 D13 and D14 Note...

Страница 273: ...e DNS2 Also you can press Number keys to directly input a specific digit 20 Press the F5 Enter key or Knob key again to confirm the input digit for D21 And the cursor will automatically jump to next g...

Страница 274: ...TCP Web Configuration 1 Press the Menu key and then the Page Down key repeatedly until the Lan Setup configuration menu appears 2 Press the F4 NEXT key repeatedly or scroll the Knob key to move to th...

Страница 275: ...irm the Web ON option Telnet Configuration 5 Press the F4 NEXT key repeatedly or scroll the Knob key to move to the Protocol Telnet field 6 Press the F5 Enter key or Knob key followed by scrolling Kno...

Страница 276: ...rspro com 7 Press the F5 Enter key or Knob key to confirm the Telnet ON option Telnet Port Configuration 8 Press the F4 NEXT key repeatedly or scroll the Knob key to move to the Protocol Telnet Port f...

Страница 277: ...et Port Also you can press Number keys to directly input a specific digit 10 Press the F5 Enter key or Knob key to confirm the input digit for Telnet Port Range 1024 65535 Default 3000 Telnet ECHO Con...

Страница 278: ...ed by scrolling Knob key or pressing keys to land on the ON option 13 Press the F5 Enter key or Knob key again to confirm the Telnet ECHO ON option TCP Configuration 14 Press the F4 NEXT key repeatedl...

Страница 279: ...scrolling Knob key or pressing keys to land on the ON option 16 Press the F5 Enter key or Knob key again to confirm the TCP ON option TCP Port Configuration 17 Press the F4 NEXT key repeatedly or scro...

Страница 280: ...5 Default 3001 19 Press the F5 Enter key or Knob key again to confirm the input digit for TCP Port Remote Terminal Session Telnet TCP Background A terminal application can be used to remotely control...

Страница 281: ...n this query via the terminal application idn The command will return the instrument manufacturer model number serial number and firmware version in the following format RSPRO RSDM9061 000000000 M0 69...

Страница 282: ...a SCPI command is separated by a colon For example the diagram below shows an SCPI sub structure and a command example DCAC DC AC CONFigure VOLTage DC VOLTage CONFigure Command Types There are a numbe...

Страница 283: ...written commands Long form CONFigure DIODe CONFIGURE DIODE Configure diode Short form CONF DIOD conf diod Square Brackets Commands that contain square brackets indicate that the contents are optional...

Страница 284: ...d in place of any numerical parameter where indicated For queries it will return the highest possible value allowed for the particular setting DEF For commands this will set the setting to the default...

Страница 285: ...OL Remote Command Marks the end of a command line The following messages are in accordance with IEEE488 2 standard LF CR CR LF LF CR The most common EOL character is CR LF Message Separator EOL or sem...

Страница 286: ...CALCulate AVERage AVERage 304 CALCulate AVERage CLEar IMMediate 304 CALCulate AVERage COUNt 304 CALCulate AVERage MAXimum 304 CALCulate AVERage MINimum 304 CALCulate AVERage PTPeak 304 CALCulate AVERa...

Страница 287: ...ANsform HISTogram CLEar IMMediate 308 CALCulate TRANsform HISTogram COUNt 308 CALCulate TRANsform HISTogram DATA 308 CALCulate TRANsform HISTogram POINts 308 CALCulate TRANsform HISTogram POINts 308 C...

Страница 288: ...umber WAIT 315 DIGital INTerface MODE 316 DIGital INTerface MODE 316 DIGital INTerface DATA OUTPut 316 DIGital INTerface DATA SETup 316 DISPlay STATe 316 DISPlay STATe 316 DISPlay TEXT CLEar 317 DISPl...

Страница 289: ...ENSe CAPacitance RANGe 325 SENSe CAPacitance RANGe 325 SENSe CAPacitance RANGe AUTO 325 SENSe CAPacitance RANGe AUTO 325 SENSe CONTinuity NPLCycles 325 SENSe CONTinuity NPLCycles 325 SENSe CONTinuity...

Страница 290: ...VOLTage AC BANDwidth 330 SENSe VOLTage AC NULL STATe 330 SENSe VOLTage AC NULL STATe 330 SENSe VOLTage AC NULL VALue 330 SENSe VOLTage AC NULL VALue 330 SENSe VOLTage AC NULL VALue AUTO 331 SENSe VOL...

Страница 291: ...AC TERMinals 335 SENSe CURRent AC TERMinals 335 SENSe CURRent AC TRIGger DELay 335 SENSe CURRent AC TRIGger DELay 336 SENSe RESistance NPLCycles 336 SENSe RESistance NPLCycles 336 SENSe RESistance NUL...

Страница 292: ...SENSe FREQuency CURRent RANGe AUTO 341 SENSe FREQuency INPutjack 341 SENSe FREQuency INPutjack 341 SENSe FREQuency NULL STATe 341 SENSe FREQuency NULL STATe 342 SENSe FREQuency NULL VALue 342 SENSe F...

Страница 293: ...ture NULL VALue AUTO 346 SENSe TEMPerature NULL VALue AUTO 346 SENSe TEMPerature RESolution 346 SENSe TEMPerature RESolution 347 SENSe TEMPerature TRANsducer TYPE 347 SENSe TEMPerature TRANsducer TYPE...

Страница 294: ...ure FRTD TYPE 350 SENSe TEMPerature THERmistor APARameter 350 SENSe TEMPerature THERmistor APARameter 350 SENSe TEMPerature THERmistor BPARameter 350 SENSe TEMPerature THERmistor BPARameter 350 SENSe...

Страница 295: ...STem BEEPer HOLD VOLume 356 SYSTem BEEPer HOLD VOLume 356 SYSTem CLICk STATe 356 SYSTem CLICk STATe 356 SYSTem DATE 356 SYSTem DATE 356 SYSTem DISPlay 356 SYSTem DISPlay 356 SYSTem ERRor NEXT 356 SYST...

Страница 296: ...Tem COMMunicate LAN TELNet ENABle 361 SYSTem COMMunicate LAN TELNet ENABle 362 SYSTem COMMunicate LAN TELNet PORT 362 SYSTem COMMunicate LAN TELNet PORT 362 SYSTem COMMunicate LAN TELNet PROMpt 362 SY...

Страница 297: ...ition 365 STATus QUEStionable ENABle 365 STATus QUEStionable ENABle 365 STATus QUEStionable EVENt 366 CLS 366 ESE 366 ESE 366 ESR 366 IDN 367 OPC 367 OPC 367 OPT 367 PSC 368 PSC 368 RCL 368 RST 368 SA...

Страница 298: ...7 5 0E 07 1 0E 06 2 0E 06 4 0E 06 5 0E 06 100m 1 0E 07 2 0E 07 3 0E 07 1 0E 06 2 0E 06 5 0E 06 1 0E 05 2 0E 05 4 0E 05 5 0E 05 1 1 0E 06 2 0E 06 3 0E 06 1 0E 05 2 0E 05 5 0E 05 1 0E 04 2 0E 04 4 0E 04...

Страница 299: ...8741E 01 4 35163427E 01 4 33118686E 01 The FETCh query does not erase measurements from the reading memory You can send the query multiple times to retrieve the same data You can store up to 10 000 me...

Страница 300: ...the measurements from the reading memory use FETCh Use DATA REMove or R to read and erase all or part of the available measurements Use ABORt to return to idle R reading_number Reads and erases measur...

Страница 301: ...E 04 The Read query will not return the unit or count number of the reading Sending READ is similar to sending INITi te followed immediately by FETCh VAL Returns the 1st and 2nd display value Example...

Страница 302: ...on of the switch but cannot change it Return parameter FRON REAR On the RSDM 9060 this query always returns FRON TIME SYNC SERVer Sets the server source for time sync of the 2nd group Parameter server...

Страница 303: ...rameter OFF HOLD DB DBM LIM MXB INV REF CALCulate HOLD REFerence Sets the percentage threshold for the Hold function Parameter NRf 0 01 0 1 1 10 Example CALC HOLD REF 10 Sets the hold percentage to 10...

Страница 304: ...ameter NRf CALCulate AVERage SDEViation Returns the Standard Deviation value Return parameter NRf CALCulate AVERage STATe Turns the statistic calculation function on off Parameter 0 1 ON OFF Example C...

Страница 305: ...F Example CALC LIM UPP DATA 1 0 Sets the upper limit to 1 0 CALCulate LIMit UPPer DATA Returns the upper limit value of the compare function Return parameter NRf CALCulate LIMit STATe Sets the status...

Страница 306: ...the resistance value from the dBm function Return parameter NRf CALCulate SCALe REFerence AUTO Sets the first measurement as the reference value Parameter 0 1 ON OFF Example CALC SCAL REF AUTO ON Sets...

Страница 307: ...function on CALCulate TCHart STATe Returns the trend chart function state Return parameter 0 1 1 ON 0 OFF CALCulate TRANsform HISTogram STATe Turns the histogram function on off Parameter 0 1 ON OFF...

Страница 308: ...orm HISTogram POINts Sets the number of bins between the lower and upper range values for the histogram Parameter NR1 10 20 40 100 200 400 MIN MAX DEF Example CALC TRAN HIST POIN MAX Sets the number o...

Страница 309: ...er range value of the histogram function Return parameter NRf CALCulate TRANsform HISTogram STATe Turns the histogram function on off Parameter 0 1 ON OFF Example CALC TRAN HIST STAT OFF Turns the his...

Страница 310: ...put continuously changes If your application requires autoranging specify DEFault for the resolution or omit the resolution altogether CONFigure VOLTage AC Sets measurement to AC Voltage on the 1st di...

Страница 311: ...ument cannot accurately resolve the integration time especially if the input continuously changes If your application requires autoranging specify DEFault for the resolution or omit the resolution alt...

Страница 312: ...EF Example CONF CAP 10e 7 Sets the Capcitance range to 100nF CONFigure CONTinuity Sets measurement to Continuity on the 1st display Parameter None CONFigure DIODe Sets measurement to Diode on the 1st...

Страница 313: ...specify DEFault for the resolution or omit the resolution altogether CONFigure2 VOLTage AC Sets measurement to AC Voltage on the 2nd display and specifies range Parameter None Range NRf AUTO MIN MAX...

Страница 314: ...display CONFigure2 PERiod Sets measurement to Period on the 2nd display and specifies the range Parameter None Range NRf AUTO MIN MAX DEF Resolution NRf MIN MAX DEF Example CONF2 PER Sets the measure...

Страница 315: ...old Returns the event threshold number Return parameter NR1 Ex 10 DATA REMove reading_number WAIT Reads and erases measurement values from reading memory up to the specified reading_number The measure...

Страница 316: ...O make use of this cammonad to set output status Parameter NR1 0 255 Boolean 0 1 serial input data strobe pulse Example DIG INT MDOE 4094 DIG INT DATA OUPT 10 1 DIGital INTerface DATA SETup When the...

Страница 317: ...ts the display form of measured value Parameter NUMeric HISTogram TCHart METer Example DISP VIEW HIST Sets display in the histogram mode DISPlay VIEW Returns the display form of measured value Return...

Страница 318: ...AUTO MIN MAX DEF Example MEAS VOLT AC 1 34567684E 04 Returns the AC voltage measurement value as 0 134 mV MEASure CURRent DC Returns the DC current measurement value on the 1st display Parameter None...

Страница 319: ...her MEASure FRESistance Returns the 4 wire resistance measurement value on the 1st display Parameter None Range NRf AUTO MIN MAX DEF Resolution NRf MIN MAX DEF Example MEAS FRES 1 18134472E 06 Sets me...

Страница 320: ...ue on the 1st display Example MEAS DIOD Returns the diode measurement value MEASure TEMPerature Returns the temperature measurement value with the selected probe and type on the 1st display Parameter...

Страница 321: ...e Range NRf AUTO MIN MAX DEF Example MEAS2 VOLT AC MIN 5 11895142E 04 Returns the AC voltage measurement value as 0 5118 mV MEASure2 CURRent DC Returns the DC current measurement value on the 2nd disp...

Страница 322: ...d Commands SENSe FUNCtion X Sets the function for the 1st or 2nd display which X 1 indicate 1st display X 2 indicate 2nd display Parameter 1st VOLT DC VOLT AC CURR DC CURR AC RES FRES FREQ PER TEMP TC...

Страница 323: ...Nt X Sets the digital filter count which X 1 indicate 1st display X 2 indicate 2nd display Parameter NR1 2 100 MIN MAX DEF Example SENS AVER COUN 100 Sets 2nd display digital filter count number to 10...

Страница 324: ...o 0 1 SENSe AVERage WINDow X Returns the digital filter window value Return parameter 0 01 0 1 1 10 NONE SENSe AVERage WINDow METHod X Selects a digital filter window method type which X 1 indicate 1s...

Страница 325: ...y measurement in PLCs power line cycles Where one PLC is equal to 16 6 milliseconds For any NRf parameter the DMM will automatically set the PLC to the closest acceptable PLC value 0 15 0 6 1 Paramete...

Страница 326: ...er delay time in seconds of Continuity measurement Return parameter NRf SENSe CONTinuity ZERO AUTO Sets the auto zero mode to on off or once only of Continuity measurement Parameter 0 1 ON OFF ONCE Ex...

Страница 327: ...of Diode measurement Parameter NRf 0 3600 s MIN MAX DEF Example SENS DIOD TRIG DEL 0 5 Sets the trigger delay time to 500ms of Diode measurement SENSe DIODe TRIGger DELay Returns the trigger delay ti...

Страница 328: ...Voltage measurement in PLCs power line cycles Where one PLC is equal to 16 6 milliseconds Return parameter 0 006 0 0083 0 0125 0 025 0 05 0 15 0 6 1 3 12 SENSe VOLTage DC NULL STATe Sets the relative...

Страница 329: ...Sets the DC voltage Auto range setting on off or once only Parameter 0 1 ON OFF ONCE Example SENS VOLT DC RANG AUTO ON Turns Auto range on for DC voltage measurements SENSe VOLTage DC RANGe AUTO Retur...

Страница 330: ...20 Sets the AC bandwidth to 20Hz SENSe VOLTage AC BANDwidth Returns the AC bandwidth Return parameter NRf Ex 3 00000000E 00 SENSe VOLTage AC NULL STATe Sets the relative function on off for AC Voltag...

Страница 331: ...Se VOLTage AC RANGe Returns the AC Voltage measurement range SENSe VOLTage AC RANGe AUTO Sets the AC voltage Auto range setting on off or once Parameter 0 1 ON OFF ONCE Example SENS VOLT AC RANG AUTO...

Страница 332: ...power line cycles Where one PLC is equal to 16 6 milliseconds Return parameter 0 006 0 0083 0 0125 0 025 0 05 0 15 0 6 1 3 12 SENSe CURRent DC NULL STATe Sets the relative function on off for DC Curre...

Страница 333: ...settings on off or once Parameter 0 1 ON OFF ONCE Example SENS CURR DC RANG AUTO OFF Turns Auto range off for DC current measurement SENSe CURRent DC RANGe AUTO Returns the DC current Auto range setti...

Страница 334: ...measurement Return Parameter 0 1 1 ON 0 OFF SENSe CURRent AC BANDwidth Sets the AC current bandwidth AC filter Parameter NRf 3 20 200 MIN MAX DEF Example SENS CURR AC BAND 3 Sets the AC current bandw...

Страница 335: ...range SENSe CURRent AC RANGe AUTO Sets the AC current Auto range settings on off or once Parameter 0 1 ON OFF ONCE Example SENS CURR AC RANG AUTO OFF Turns Auto range off for AC current measurements...

Страница 336: ...rement SENSe RESistance NPLCycles Returns the integration time for 2 wire resistance measurement in PLCs power line cycles Where one PLC is equal to 16 6 milliseconds Return parameter 0 006 0 0083 0 0...

Страница 337: ...ample SENS RES RANG 1000 Sets the 2 wire resistance range to 1k SENSe RESistance RANGe Returns the 2 wire resistance measurement range SENSe RESistance RANGe AUTO Sets the 2 wire resistance Auto range...

Страница 338: ...SENSe RESistance ZERO AUTO Returns the auto zero mode of 2 wire resistance measurement Return Parameter 0 1 1 ON 0 OFF SENSe FRESistance NPLCycles Sets the integration time for 4 wire resistance measu...

Страница 339: ...2 for 4 wire resistance measurement SENSe FRESistance NULL VALue Returns the current relative value of 4 wire resistance measurement SENSe FRESistance NULL VALue AUTO Sets the relative value auto on...

Страница 340: ...01 SENSe FRESistance RESolution Returns the 4 wire resistance resolution SENSe FRESistance TRIGger DELay Sets the trigger delay time that minimum step is microseconds of 4 wire resistance measurement...

Страница 341: ...ANGe AUTO Sets the Frequency Auto range settings on off or once Parameter 0 1 ON OFF ONCE Example SENS FREQ CURR RANG AUTO ON Turns the Auto range on for the frequency function SENSe FREQuency CURRent...

Страница 342: ...ment as null value SENSe FREQuency NULL VALue AUTO Returns the null value auto state of Frequency measurement SENSe FREQuency TIMeout AUTO Assigns timeout time at the frequency measurement Parameter 0...

Страница 343: ...time gate time for the period function 0 01s 0 1s 1s Parameter NRf 0 01 0 1 1 Example SENS PER APER 0 1 Sets the gate time to 0 1 seconds for the period function SENSe PERiod APERture Returns the ape...

Страница 344: ...1 0 OFF 1 ON SENSe PERiod NULL VALue Sets the relative value for Period measurement Parameter NRf 1 2 1 2 s MIN MAX DEF Example SENS FREQ NULL VAL 1 Sets the relative value to 1s for Period measuremen...

Страница 345: ...easurement range Parameter Range NRf MIN MAX DEF Example SENS PER VOLT RANG DEF Sets the period to the default range SENSe PERiod VOLTage RANGe Returns the period measurement range SENSe PERiod VOLTag...

Страница 346: ...he relative value for Temperature measurement Parameter NRf 1 0e15 1 0e15 MIN MAX DEF Example SENS FREQ NULL VAL 5 Sets the relative value to 5 C for Temperature measurement SENSe TEMPerature NULL VAL...

Страница 347: ...nt SENSe TEMPerature TRIGger DELay Returns the trigger delay time in seconds of Temperature measurement Return parameter NRf SENSe TEMPerature ZERO AUTO Sets the auto zero mode to on off or once of Te...

Страница 348: ...ter NRf 20 00 20 00 MIN MAX DEF Example SENS TEMP RJUN SIM AUTO OFFS 5 Sets the junction reference temperature adjust value to 5 C SENSe TEMPerature RJUNction SIMulated AUTO OFFSet Returns junction re...

Страница 349: ...SENS TEMP RTD RES REF 100 SENSe TEMPerature RTD RESistance REFerence Returns the 2 wire RTD reference resistance R0 SENSe TEMPerature RTD TYPE Sets the 2 wire RTD sensor type Return parameter Type PT...

Страница 350: ...D sensor to PT100 SENSe TEMPerature FRTD TYPE Returns the 4 wire RTD sensor type Return parameter PT100 D100 F100 PT385 PT3916 USER SENSe TEMPerature THERmistor APARameter Sets the 2 wire Thermistor A...

Страница 351: ...THermistor BPARameter Sets the 4 wire Thermistor B coefficient Parameter NRf 0 0 9 999999 MIN MAX DEF Example SENS TEMP FTH BPAR 0 003425 SENSe TEMPerature FTHermistor BPARameter Returns the 4 wire Th...

Страница 352: ...counts Parameter NRf 1 0 1000000 0 MIN MAX DEF Example TRIG COUN 10 SAMP COUN 10 READ Will returns 100 measurment results Sets the number of trigger counts to 10 The total measurement counts is trigge...

Страница 353: ...ing edge POS TRIGger SLOPe Returns the method of external trigger Return parameter POS NEG TRIGger SOURce Selects the trigger source Parameter IMMediate EXTernal BUS Example TRIG SOUR EXT Sets the tri...

Страница 354: ...place the instrument in the wait for trigger state by sending INITiate or READ A trigger is not accepted from the selected trigger source until the instrument is in the wait for trigger state TRIGger...

Страница 355: ...0FF Turns the buzzer off The key sound of front panel is Not affected by the state The command of SYSTem BEEPer is Not affected by the state SYSTem BEEPer STATe Returns the buzzer state Return paramet...

Страница 356: ...eter 0 1 ON OFF Example SYST CLIC STAT 0FF Turns key sound off SYSTem CLICk STATe Returns the key sound of front panel state Return Parameter 0 1 1 ON 0 OFF SYSTem DATE Sets the date for the instrumen...

Страница 357: ...a message in a large font on the bottom half of the instrument s front panel display Parameter message max length 40 characters Example SYST LAB RSPRO To turn off the message send the following to ch...

Страница 358: ...ory number Last State before power off SYSTem PARameter SAVE Saves the system parameters into 1 of 5 memory slots Parameter NR1 1 5 Example SYST PAR SAVE 1 Saves the system parameters to memory 1 SYST...

Страница 359: ...Parameter NR1 hour minute second Example SYST TIME 16 20 30 Sets the time to 16 20 30 hour 0 23 minute 0 60 second 0 60 SYSTem TIME Returns system time Return parameter Time Ex 16 20 40 000 SYSTem UPT...

Страница 360: ...tion information SYSTem COMMunicate LAN DHCP Returns the DHCP state Return parameter 0 1 0 OFF 1 ON SYSTem COMMunicate LAN DNS X Sets the DNS address which X 1 indicate DNS1 X 2 indicate DNS2 Paramete...

Страница 361: ...12 Hexadecimal characters SYSTem COMMunicate LAN SMASk Sets the subnet mask address Parameter address Example SYST COMM LAN SMAS 255 255 255 0 Sets the subnet mask address to 255 255 255 0 SYSTem COMM...

Страница 362: ...t Returns the telnet prompt message Retrurn parameter string SYSTem COMMunicate LAN TELNet TIMeout Sets the timeout time for auto logout from Telnet communication where unit of time is second Paramete...

Страница 363: ...Tem COMMunicate LAN TCP PORT Returns the TCP port number Retrurn parameter NR1 SYSTem COMMunicate LAN TIMeout Sets the TCP communication timeout time where unit second Parameter NR1 1 60000 Example SY...

Страница 364: ...ent Condition register bits are updated in real time they are neither latched nor buffered This register is read only bits are not cleared when read STATus OPERation ENABle Sets bits in the Operation...

Страница 365: ...y monitors the state of the instrument Condition register bits are updated in real time they are neither latched nor buffered This register is read only bits are not cleared when read STATus QUEStiona...

Страница 366: ...Register ESE Returns the ESER Event Status Enable Register contents Example ESE 130 Returns 130 ESER 10000010 ESE Sets the ESER contents Parameter NR1 0 255 Ex ESE 65 Sets the ESER to 01000001 The sel...

Страница 367: ...MP COUN 100 INIT OPC The difference between OPC and OPC is that OPC sets a status bit when the operation completes and OPC outputs 1 when the operation completes OPC Sets operation complete bit bit0 i...

Страница 368: ...s 1 enables RCL Load the system parameters from 0 of 5 memory locations Parameter NR1 0 5 0 Default settings 1 5 memory number Example RCL 1 Loads the memory 1 system parameters RST Recalls default pa...

Страница 369: ...s the contents of the SBR as 01010001 A condition register continuously monitors the state of the instrument Condition register bits are updated in real time they are neither latched nor buffered This...

Страница 370: ...initions for the Questionable Data Register Note The overload bits are set once per INITiate command If you clear an overload bit it is not set again until a new INITiate is sent Bit Name DecimalDefin...

Страница 371: ...7 Not Used 128 Reserved for future use 8 Not Used 256 Reserved for future use 9 Resistance Overload 512 Only reported as event In Conditon Register this bit always returns 0 Read the Event Register 1...

Страница 372: ...Reserved for future use 7 Not Used 128 Reserved for future use 8 Configurtion Change 256 Instrument configuration has been changed since the last INIT READ or MEASure either from the front panel or f...

Страница 373: ...ery has been read Or both the input and output buffers are full 3 Device Error 8 A device error including a self test error or calibration error occurred an error in the 300 range or any positive erro...

Страница 374: ...et in the Questionable Data Register bits must be enabled see STAT QUES ENAB 4 Message Available 16 Data is available in the instrument s output buffer 5 Standard Event 32 One or more bits are set in...

Страница 375: ...cteristics 394 Frequency and Period Characteristics 395 Measuring Characteristics 395 Operating Characteristics 396 Temperature Characteristics 1 Exclusive of probe errors 396 Capacitance 398 RSDM 906...

Страница 376: ...socket indicates the line voltage is positioned as 240V 3 Pull the fuse holder out of the fuse socket gently as the right figure illustrates 4 Further pull the fuse out of the fuse holder and replace...

Страница 377: ...result shows OPEN either one of the fuses requires replacement The one is accessible from the rear panel and the other one is within the internal part If either one of the fuses of 3A input current is...

Страница 378: ...e is still a damaged fuse in 3A input current follow the instructions in the following section to replace fuse of internal 3A input current Replace internal 10A input fuse RSDM 9061 only To make sure...

Страница 379: ...Power off properly and disconnect all the test leads cables including power cord 2 Disassemble the instrument case in light of the disassembling instructions 3 Make sure the certain fuse to be replac...

Страница 380: ...aware of potential risks to disassemble instrument case Unplug power cord and disconnect external circuit from the instrument before opening the case Some of the electrical connections are dynamic and...

Страница 381: ...pose or recycle it in accord with the applicable regulations 7 Place the new battery CR2032 into the compartment and beware of the polarity is way close to the metal guard plate Gently press the batte...

Страница 382: ...V 1ST Range Auto Range 1ST Speed 5 s 2ND Function Off DCV Ratio Off Filter On Filter Type Move Filter Count 10 Filter Windows 0 10 Filter Method Measure Auto Zero On Input Impedance 10M fixed for DCV...

Страница 383: ...Simulated Method Auto Simulated junction 23 Auto Simulated ADJ 0 RTD Type PT100 R0 100 Thermistor Type 5k Display Item List Factory Default Parameter Parameter Save Load for Group 1 5 Digit Auto Displ...

Страница 384: ...shold 0 10 Rel Function Off dB Reference Method dBm Reference Resistance 600 dBm Reference Resistance 600 Compare Beep Mode Off Beep Volume Medium Low Limit 1 High Limit 1 MX B M Value 1 B Value 0 Tri...

Страница 385: ...ab Password Enable Display Brightness 60 AutoOff OFF AutoOff Time 30min 1ST Font Color White 2ND Font Color White Math Font Color White Math Off Display Mode Off Antialiasing Off Additional Info All O...

Страница 386: ...Port 3001 Only utilized parameters are listed here due to over amount parameters The rest of the parameters unlisted however can be saved and loaded as well It indicates parameters can be saved and lo...

Страница 387: ...AT II 300V MAX DC1000V AC 750V Line Power Power Supply 100 120 220 240 VAC 10 Power Line Frequency 50 Hz 60 Hz 400 Hz 10 Power Consumption Max 25 VA Environment Operating Environment Full accuracy for...

Страница 388: ...crement of one coefficient per one degree celsius when the range is beyond TCAL 5 C Accuracy Specification It is relevant to the calibration standard Real Time Clock Calendar Set and read year month d...

Страница 389: ...0 0040 010 0 004 0 0008 0 0005 1 000000 k 1 mA 0 002 0 0005 0 008 0 0010 010 0 001 0 0008 0 0001 10 00000 k 100 A 0 002 0 0005 0 008 0 0010 010 0 001 0 0008 0 0001 100 0000 k 10 A 0 002 0 0005 0 008 0...

Страница 390: ...e Diode Test 4 Range 2 24 Hour TCAL 1 C 90 Day TCAL 5 C 1 Year TCAL 5 C Temperature Coefficient C 5 V 0 002 0 030 0 008 0 0300 01 0 03 0 001 0 002 Accuracy Specifications of reading of range DCV Ratio...

Страница 391: ...will pass the set range by 20 except the tests of 1000 V DC 3 A DC 10 A DC and diode 3 This specification applies to 4 wire resistance measurement whilst it requires using REL function for offset on 2...

Страница 392: ...5 0 12 0 05 0 011 0 005 50 kHz 100 kHz 0 55 0 08 0 60 0 08 0 60 0 08 0 060 0 008 100 kHz 300 kHz4 00 0 50 4 00 0 50 4 00 0 50 0 200 0 020 Accuracy Specifications of reading of range True RMS AC Curren...

Страница 393: ...1 10 0 04 1 10 0 04 1 10 0 04 0 100 0 006 5 Hz 10 Hz 0 35 0 04 0 35 0 04 0 35 0 04 0 035 0 006 10 Hz 5 kHz 0 15 0 04 0 15 0 04 0 15 0 04 0 015 0 006 5 kHz 10 kHz 0 35 0 04 0 35 0 04 0 35 0 04 0 030 0...

Страница 394: ...Input Impedance 1 M 2 in parallel with 100 pF Input Protection 750 Vrms on all ranges True RMS AC Current Range Shunt Burden Voltage 100 A 100 0 011 V 1 mA 100 0 11 V 10 mA 1 0 04 V 100 mA 1 0 4 V 1 A...

Страница 395: ...front panel only Due to power factor resulting in temperature rise 2 mA increment per one ampere when input is greater than 5 A rms Frequency and Period Characteristics Frequency Period 1 2 Range Fre...

Страница 396: ...amplitude range is from 10 to 120 and is lower than 750 VAC The input 60 mV for 300 k 1 MHz within 100mV range Temperature Characteristics 1 Exclusive of probe errors RTD Accuracy based on PT100 100...

Страница 397: ...and test lead error will be constrained by the adopted test lead The test lead accuracy adder covers all errors of measurements and ITS 90 temerature change Capacitance 1 Range 24 Hour TCAL 1 C 90 Da...

Страница 398: ...is selected The charge and discharge time is used to calculate the capacitance of Cx if the selected capacitance range is equal to or less than 10 nF Only the charge time is used to calculate the capa...

Страница 399: ...0070 014 0 007 0 0006 0 0005 1 000000 k 1 mA 0 003 0 00080 011 0 0010 014 0 001 0 0006 0 0001 10 00000 k 100 A 0 003 0 00050 011 0 0010 014 0 001 0 0006 0 0001 100 0000 k 10 A 0 003 0 00050 011 0 001...

Страница 400: ...AL 5 C Temperature Coefficient C 5 V 0 003 0 030 0 011 0 0300 014 0 030 0 0010 0 0020 Accuracy Specifications of reading of range DCV Ratio 5 Accuracy Specification DC Input accuracy DC Reference accu...

Страница 401: ...range of measurement will pass the set range by 20 except the tests of 1000 V DC 3 A DC 10 A DC and diode 3 This specification applies to 4 wire resistance measurement whilst it requires using REL fu...

Страница 402: ...05 0 12 0 05 0 011 0 005 50 kHz 100 kHz 0 55 0 08 0 60 0 08 0 60 0 08 0 060 0 008 100 kHz 300 kHz4 00 0 50 4 00 0 50 4 00 0 50 0 200 0 020 Accuracy Specifications of reading of range True RMS AC Curre...

Страница 403: ...al Crest Factor Errors non sine wave Crest Factor Error of reading 1 2 0 05 2 3 0 15 3 4 0 30 4 5 0 40 Additional Low Frequency Errors of reading Frequency Speed 1 s 3 Hz 5 s 20 Hz 20 s 200 Hz 10 Hz 2...

Страница 404: ...tion It will be available after 60 minutes of warm up sine wave as well as 1 s speed rate 2 The entire range of measurement will pass the set range by 20 except the tests of 750 VAC 3 A AC and 10 A AC...

Страница 405: ...function Voltage Ranges 100 mVrms full scale to 750 Vrms Auto or manual ranging Settling Considerations Errors will occur when attempting to measure the frequency or period of an input following a dc...

Страница 406: ...0 001 C 0 08 C 0 005 C C 20 C 20 C 0 001 C 0 06 C 0 005 C C 20 C 100 C 0 001 C 0 08 C 0 005 C C 100 C 300 C 0 001 C 0 12 C 0 007 C C 300 C 600 C 0 001 C 0 22 C 0 009 C C Thermocouples Accuracy based o...

Страница 407: ...C Temperature Coefficient C 1 000 nF 2 00 2 00 2 00 2 00 2 00 2 00 0 05 0 01 10 00 nF 2 00 1 00 2 00 1 00 2 00 1 00 0 05 0 01 100 0 nF 2 00 0 40 2 00 0 40 2 00 0 40 0 05 0 01 1 000 F 2 00 0 40 2 00 0...

Страница 408: ...ed to calculate the capacitance of Cx if the selected capacitance range is equal to or less than 10 nF Only the charge time is used to calculate the capacitance of Cx if the selected capacitance range...

Страница 409: ...409 Dimensions All dimensions are shown in millimeters...

Страница 410: ...Electrical equipment for measurement control and laboratory use EMC requirements 2013 Conducted Radiated Emission EN 55011 2016 A1 2017 Class A Electrical Fast Transients EN 61000 4 4 2012 Current Har...

Страница 411: ...ea Business Centre Phase 2 NO 618 Yan an Eastern Road Shanghai 200001 China Europe RS Components Ltd PO Box 99 Corby Northants NN17 9RS United Kingdom Japan RS Components Ltd West Tower 12th Floor Yok...

Страница 412: ...412 rspro com...

Отзывы: