Thermistor 2W/4W Setting
Background
The RSDM-9060/9061 supports 2 or 4 wire
Thermistor. It is important to specify the type of
temperature sensor used.
Parameter
Type
Range
Resolution
All
-80~150°C
0.001°C
Procedure
1.
Press the F1 (Probe) key to
enter the
Temperature Probe menu followed by clicking
either the F4 (Therm2W)
or F5
(Therm4W) key
to activate Therm
2W/4W mode. See the figure shown below.
2.
Press the F5 (Type) key
to enter the
Sensor Type menu as the figure shown below.
Click the F1 – F3 key to select a desired sensor
type per your requirement.
3.
The display shows the latest setting. See the
example of the figure below where Thermistor
2W: 10kΩ is currently activated by user.
Probe: Therm2W
Type: 10k
Ω
Temperature
Measurement
Содержание RSDM-9060
Страница 1: ...User Manual RSDM 906X series Stock number 1919720 RSDM 9061 1919721 RSDM 9060 EN rspro com...
Страница 2: ......
Страница 5: ...5 Dimensions 409 Declaration of Conformity 410...
Страница 23: ...23 DC AC 3 15A Input Current Fuse Holds the current fuse T3 15A 500V 5 20mm For fuse replacement details...
Страница 141: ...141 EOM pulse width timing...
Страница 160: ...160 rspro com SYSTEM FIRMWARE View System Info 161 Firmware Update 162...
Страница 211: ...211 SCREENHOT LOG Capture 212 Save Reading 215...
Страница 219: ...219 DISPLAY SETTING Digit 220 Display 222 Number 222 Bar Meter 223 Trend Chart 227 Histogram 238...
Страница 409: ...409 Dimensions All dimensions are shown in millimeters...
Страница 412: ...412 rspro com...