![RS PRO RSDM-9060 Скачать руководство пользователя страница 299](http://html1.mh-extra.com/html/rs-pro/rsdm-9060/rsdm-9060_user-manual_1496059299.webp)
299
Other Commands
ABORt
Aborts a measurement in progress, returning the instrument to
the trigger idle state.
●
Use this to abort a measurement when the instrument is
waiting for a trigger, or for aborting a long measurement or
series of measurements.
FETCh[X]?
Waits for measurements to complete and copies all available
measurements to the instrument's output buffer. The readings
remain in reading memory.
X = null or 1 indicate 1
st
display value, X = 2 indicate 2
nd
display
value
Example: SAMP:COUN 3
INIT
FETC?
Returns: -4.98748741E-01,-4.35163427E-01,-4.33118686E-01
●
The FETCh? query does not erase measurements from the
reading memory. You can send the query multiple times to
retrieve the same data.
●
You can store up to 10,000 measurements in the reading
memory of the RSDM-9060 or 100,000
measurements on the RSDM-9061. If reading memory
overflows,new measurements overwrite the oldest
measurements stored; the most recent measurements are
always preserved. No error is generated, but the Reading Mem
Ovfl bit (bit 14) is set in the Questionable Data Register's
condition register.
HCOPy:SDUMp:DATA?
Executes TFT LCD screenshot action.
Returns the front panel display image ("screen shot").
Returns a count of data streaming by the image file format of
BMP.
Содержание RSDM-9060
Страница 1: ...User Manual RSDM 906X series Stock number 1919720 RSDM 9061 1919721 RSDM 9060 EN rspro com...
Страница 2: ......
Страница 5: ...5 Dimensions 409 Declaration of Conformity 410...
Страница 23: ...23 DC AC 3 15A Input Current Fuse Holds the current fuse T3 15A 500V 5 20mm For fuse replacement details...
Страница 141: ...141 EOM pulse width timing...
Страница 160: ...160 rspro com SYSTEM FIRMWARE View System Info 161 Firmware Update 162...
Страница 211: ...211 SCREENHOT LOG Capture 212 Save Reading 215...
Страница 219: ...219 DISPLAY SETTING Digit 220 Display 222 Number 222 Bar Meter 223 Trend Chart 227 Histogram 238...
Страница 409: ...409 Dimensions All dimensions are shown in millimeters...
Страница 412: ...412 rspro com...