![RS PRO RSDM-9060 Скачать руководство пользователя страница 352](http://html1.mh-extra.com/html/rs-pro/rsdm-9060/rsdm-9060_user-manual_1496059352.webp)
352
rspro.com
TRIGger Commands
SAMPle:COUNt
Sets the number of samples.
Parameter: <NRf>(1.0 ~ 1000000.0) | MIN | MAX | DEF
Example: TRIG:COUN 10
SAMP:COUN 10
INIT
FETC?
Will returns 100 measurment results.
Sets the number of samples to 10.
●
The total measurement counts is trigger count multiplication
sample count.
SAMPle:COUNt?
Returns the number of samples.
Return parameter: <NRf>
TRIGger:COUNt
Sets the number of trigger counts.
Parameter: <NRf>(1.0 ~ 1000000.0) | MIN | MAX | DEF
Example: TRIG:COUN 10
SAMP:COUN 10
READ?
Will returns 100 measurment results.
Sets the number of trigger counts to 10.
●
The total measurement counts is trigger count multiplication
sample count.
TRIGger:COUNt?
Returns the number of trigger counts.
Return parameter: <NRf>
TRIGger:DELay
Sets the trigger delay time that minimum step is microseconds
in all of the function.
Parameter: <NRf> (0 ~ 3600 s) | MIN | MAX | DEF
Example: TRIG:DEL MAX
Sets the trigger delay time to the maximum.
TRIGger:DELay?
Returns the trigger delay time in seconds of current function.
Return parameter: <NRf>
Содержание RSDM-9060
Страница 1: ...User Manual RSDM 906X series Stock number 1919720 RSDM 9061 1919721 RSDM 9060 EN rspro com...
Страница 2: ......
Страница 5: ...5 Dimensions 409 Declaration of Conformity 410...
Страница 23: ...23 DC AC 3 15A Input Current Fuse Holds the current fuse T3 15A 500V 5 20mm For fuse replacement details...
Страница 141: ...141 EOM pulse width timing...
Страница 160: ...160 rspro com SYSTEM FIRMWARE View System Info 161 Firmware Update 162...
Страница 211: ...211 SCREENHOT LOG Capture 212 Save Reading 215...
Страница 219: ...219 DISPLAY SETTING Digit 220 Display 222 Number 222 Bar Meter 223 Trend Chart 227 Histogram 238...
Страница 409: ...409 Dimensions All dimensions are shown in millimeters...
Страница 412: ...412 rspro com...