![RS PRO RSDM-9060 Скачать руководство пользователя страница 330](http://html1.mh-extra.com/html/rs-pro/rsdm-9060/rsdm-9060_user-manual_1496059330.webp)
330
rspro.com
[SENSe:]VOLTage[:DC]:TRIGger:DELay?
Returns the trigger delay time in seconds of DC Voltage
measurement.
Return parameter: <NRf>
[SENSe:]VOLTage[:DC]:ZERO:AUTO
Sets the auto zero mode to on, off or once of DC Voltage
measurement.
Parameter: 0 | 1 | ON | OFF | ONCE
Example: SENS:VOLT:DC:ZERO:AUTO ONCE
Sets the auto zero to once.
[SENSe:]VOLTage[:DC]:ZERO:AUTO?
Returns the auto zero mode of DC Voltage measurement.
Return Parameter: 0 | 1, 1=ON, 0=OFF
[SENSe:]VOLTage:AC:BANDwidth
Sets the AC bandwidth (AC filter).
Parameter: <NRf> (3 | 20 | 200) | MIN | MAX | DEF
Example: SENS:VOLT:AC:BAND 20
Sets the AC bandwidth to 20Hz.
[SENSe:]VOLTage:AC:BANDwidth?
Returns the AC bandwidth.
Return parameter: <NRf>, Ex: 3.0000
[SENSe:]VOLTage:AC:NULL[:STATe]
Sets the relative function on/off for AC Voltage measurement.
Parameter: 0 | 1 | ON | OFF
Example: SENS:VOLT:AC:NULL:STAT ON
Turns the relative function on for AC Voltage measurements.
[SENSe:]VOLTage:AC:NULL[:STATe]?
Returns the relative function state of AC Voltage measurement.
Return parameter: 0 | 1, 0=OFF, 1=ON
[SENSe:]VOLTage:AC:NULL:VALue
Sets the relative value for AC Voltage measurement.
Parameter: <NRf> (-1200.0~1200.0 V) | MIN | MAX | DEF
Example: SENS:VOLT:AC:NULL:VAL 1
Sets the relative value to 1V for AC Voltage measurement.
[SENSe:]VOLTage:AC:NULL:VALue?
Returns the current relative value of AC Voltage measurement.
Содержание RSDM-9060
Страница 1: ...User Manual RSDM 906X series Stock number 1919720 RSDM 9061 1919721 RSDM 9060 EN rspro com...
Страница 2: ......
Страница 5: ...5 Dimensions 409 Declaration of Conformity 410...
Страница 23: ...23 DC AC 3 15A Input Current Fuse Holds the current fuse T3 15A 500V 5 20mm For fuse replacement details...
Страница 141: ...141 EOM pulse width timing...
Страница 160: ...160 rspro com SYSTEM FIRMWARE View System Info 161 Firmware Update 162...
Страница 211: ...211 SCREENHOT LOG Capture 212 Save Reading 215...
Страница 219: ...219 DISPLAY SETTING Digit 220 Display 222 Number 222 Bar Meter 223 Trend Chart 227 Histogram 238...
Страница 409: ...409 Dimensions All dimensions are shown in millimeters...
Страница 412: ...412 rspro com...