377
Replace 3A Input Current Fuse
Preparation
To make sure if 3A input current needs to be
replaced, press the
button to set RSDM-
9060/9061 in Continuity mode and short circuit
the HI input terminal with the 3A input current
terminal.
If the test result shows OPEN, either one of the
fuses requires replacement. The one is accessible
from the rear panel, and the other one is within the
internal part.
If either one of the fuses of 3A input current is
damaged, please first check the one (3.15 A, 500 V)
in the lower-left corner of rear panel.
Step
1.
Turn the RSDM-9060/9061 off
2.
Press and hold the fuse holder in the rear panel
followed by rotating it counterclockwise with a
flat-blade screwdriver.