![RS PRO RSDM-9060 Скачать руководство пользователя страница 324](http://html1.mh-extra.com/html/rs-pro/rsdm-9060/rsdm-9060_user-manual_1496059324.webp)
324
rspro.com
[SENSe:]AVERage:TCONtrol[X]
Selects the digital filter type, which X = 1 indicate 1
st
display, X
= 2 indicate 2
nd
display.
Parameter: MOV | REP
Example: SENS:AVER:TCON MOV
Sets 1
st
display digital filter type to the moving filter.
[SENSe:]AVERage:TCONtrol[X]?
Returns the digital filter type.
Return parameter: MOV (moving) | REP (repeating)
[SENSe:]AVERage:WINDow[X]
Selects a digital filter window, which X = 1 indicate 1
st
display, X
= 2 indicate 2
nd
display.
Parameters: 0.01 | 0.1 | 1 | 10 | NONE
Example: SENS:AVER:WIND 0.1
Sets 1
st
display digital filter window to 0.1%
[SENSe:]AVERage:WINDow[X]?
Returns the digital filter window value.
Return parameter: 0.01 | 0.1 | 1 | 10 | NONE
[SENSe:]AVERage:WINDow:METHod[X]
Selects a digital filter window method type, which X = 1
indicate 1
st
display, X = 2 indicate 2
nd
display.
Parameters: Measure | Range
Example: SENS:AVER:WIND:METH Measure
Sets 1
st
display digital filter window method to the measure
type
[SENSe:]AVERage:WINDow:METHod[X]?
Returns the digital filter window method type.
Return parameter: Measure | Range
SENSe CAPacitance Commands
[SENSe:]CAPacitance:CABLe:CALibratoin
It is used like Relative function before capacitance
measurement, (only be used at range 1nF,10nF)
Parameter: [None]
Example: CONF:CAP 1e-9
SENS:CAP:CABL:CAL
Makes test lead to zero before capacitance measurement.
Содержание RSDM-9060
Страница 1: ...User Manual RSDM 906X series Stock number 1919720 RSDM 9061 1919721 RSDM 9060 EN rspro com...
Страница 2: ......
Страница 5: ...5 Dimensions 409 Declaration of Conformity 410...
Страница 23: ...23 DC AC 3 15A Input Current Fuse Holds the current fuse T3 15A 500V 5 20mm For fuse replacement details...
Страница 141: ...141 EOM pulse width timing...
Страница 160: ...160 rspro com SYSTEM FIRMWARE View System Info 161 Firmware Update 162...
Страница 211: ...211 SCREENHOT LOG Capture 212 Save Reading 215...
Страница 219: ...219 DISPLAY SETTING Digit 220 Display 222 Number 222 Bar Meter 223 Trend Chart 227 Histogram 238...
Страница 409: ...409 Dimensions All dimensions are shown in millimeters...
Страница 412: ...412 rspro com...