![RS PRO RSDM-9060 Скачать руководство пользователя страница 70](http://html1.mh-extra.com/html/rs-pro/rsdm-9060/html/rs-pro/rsdm-9060/rsdm-9060_user-manual_1496059070.webp)
70
rspro.com
Reference Junction Temperature (SIM Temperature)
Background
(Thermocouple
only)
When a thermocouple is connected to the RSDM-
9060/9061, the temperature difference between the
thermocouple lead and the RSDM-9060/9061 input
terminal should be taken into account and be
cancelled out; otherwise an erroneous temperature
might be added. The value of the reference
junction temperature should be determined by the
user.
Type
Range
Resolution
SIM (simulated) -20°C ~ +80°C 0.01°C
The terminal temperature is manually defined by
user.
Default value: Auto
Thermocouple Setting
Procedure
1.
Press the F1 (Probe) key
to enter the
Temperature Probe menu followed by clicking
the F1 (TCouple) key
to activate
Thermocouple mode. See the figure shown
below.
2.
Press the F5 (Type) key
to enter the
Sensor Type menu as the figure shown below.
Click the F1 – F5 key to select a desired sensor
type per situations.
3.
Press the F6 (More 1/2) key
to enter
the next page with more sensor types available
for selection.
Содержание RSDM-9060
Страница 1: ...User Manual RSDM 906X series Stock number 1919720 RSDM 9061 1919721 RSDM 9060 EN rspro com...
Страница 2: ......
Страница 5: ...5 Dimensions 409 Declaration of Conformity 410...
Страница 23: ...23 DC AC 3 15A Input Current Fuse Holds the current fuse T3 15A 500V 5 20mm For fuse replacement details...
Страница 141: ...141 EOM pulse width timing...
Страница 160: ...160 rspro com SYSTEM FIRMWARE View System Info 161 Firmware Update 162...
Страница 211: ...211 SCREENHOT LOG Capture 212 Save Reading 215...
Страница 219: ...219 DISPLAY SETTING Digit 220 Display 222 Number 222 Bar Meter 223 Trend Chart 227 Histogram 238...
Страница 409: ...409 Dimensions All dimensions are shown in millimeters...
Страница 412: ...412 rspro com...