Using Power Sensors
R&S
®
AREG100A
73
User Manual 1178.7417.02 ─ 04
For example, to obtain a sufficient low average value, set the aperture time exactly to
one modulation period.
Remote command:
:SENSe<ch>[:POWer]:APERture:TIMe
S-Parameter ← Sensor settings
S-Parameter correction is used to mathematically shift the reference plane to the DUT
by considering the S-parameters for any components connected upstream of the sen-
sor.
The S-Parameter table can be changed with the S-Parameters tool, provided as part of
the free R&S NRP Toolkit software. For more information, refer to the manual of the
connected R&S NRP power sensor.
Remote command:
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe
:SENSe<ch>[:POWer]:CORRection:SPDevice:LIST?
:SENSe<ch>[:POWer]:CORRection:SPDevice:SELect
Enable Logging ← Sensor settings
Activates recording of R&S NRP power sensor readings in a log file.
There is 1 log file per sensor. The log files are created automatically and filled in con-
tinuously. They are text files with predefined filename
SensLog<n>.txt
, where
<n>
indicates the connected sensor. Log files are stored on the internal memory, in the
directory
/var/user/SensorLogging
.
Each log file contains the measured value (2 readings when you work with peak sen-
sors), the sensor type, and the measurement time (timestamp). Logged data is not
overwritten. When a new measurement is started, the collected logging data is appen-
ded in the log file.
Check the used disc space regularly and remove log files to maintain storage capacity.
Note:
The logging function is intended for measurements with long time intervals. It is
suitable source for data reconstructions if the connection to the sensor was interrupted.
Remote command:
:SENSe<ch>[:POWer]:LOGGing:STATe
NRP Power Viewer