Remote Control Commands
R&S
®
AREG100A
256
User Manual 1178.7417.02 ─ 04
Example:
See
Example "Performing a power measurement with a fixed fil-
Manual operation:
See
:SENSe<ch>[:POWer]:APERture:TIMe
<ApTime>
Defines the aperture time (size of the acquisition interval) for the corresponding sensor.
Parameters:
<ApTime>
float
Range:
depends on connected power sensor
Increment: 1E-9
*RST:
depends on connected power sensor
Example:
See
Example "Performing a power measurement with a fixed fil-
Manual operation:
See
:SENSe<ch>[:POWer]:CORRection:SPDevice:SELect
<Select>
Several S-parameter tables can be stored in a sensor. The command selects a loaded
data set for S-parameter correction for the corresponding sensor.
Parameters:
<Select>
float
*RST:
0
Manual operation:
See
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe
<State>
Activates the use of the S-parameter correction data.
Note:
If you use power sensors with attenuator, the instrument automatically activates
the use of S-parameter data.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
SENSe1:POWer:CORRection:SPDevice:STATe 1
Activates the use of the S-parameters correction data.
Manual operation:
See
:SENSe<ch>[:POWer]:CORRection:SPDevice:LIST?
Queries the list of the S-parameter data sets that have been loaded to the power sen-
sor.
SENSe, READ, INITiate and SLISt Subsystems