Remote Control Commands
R&S
®
AREG100A
192
User Manual 1178.7417.02 ─ 04
:SYSTem:FPReset
Triggers an instrument reset to the original state of delivery.
Example:
SYST:FPR
All instrument settings (also the settings that are not currently
active) are reset to the factory values.
Usage:
Event
Manual operation:
See
9.3
CALibration Subsystem
The
CALibration
subsystem contains the commands needed for performing internal
adjustments. This procedure is triggered by the query commands.
Suffix <hw>
Suffix
Value range
Description
CALibration<hw>
[1]
Optional suffix
Understanding the query response
●
0
: error-free execution of the adjustments
●
1
: indicates that an error occurred; the process has been canceled
......................................................................................... 192
.........................................................................................193
:CALibration<hw>:ALL:INFormation?
............................................................................... 193
........................................................................................ 193
......................................................................................... 194
:CALibration:DATA:FACTory:DATE?
.................................................................................194
:CALibration<hw>:CONTinueonerror
................................................................................ 194
:CALibration:ALL[:MEASure]?
[<Force>]
Starts all internal adjustments that do not need external measuring equipment.
Query parameters:
<Force>
string
Return values:
<Measure>
0 | 1 | OFF | ON
Example:
CAL:ALL:MEAS?
0
Executes the adjustments of all instrument functions. When
completed, it indicates that the adjustment has been performed
successfully.
CALibration Subsystem