Network Operation and Remote Control
R&S
®
AREG100A
156
User Manual 1178.7417.02 ─ 04
The dialog shows all specified resource strings of the supported remote control
interfaces.
8.7
How to Change the GPIB Instrument Address
To control the instrument remotely via the GPIB bus, it must be addressed using the
GPIB address. The remote control address is factory-set to 28, but it can be changed if
it does not fit in the network environment. For remote control, addresses 0 through 30
are allowed. The GPIB address is maintained after a reset of the instrument settings.
To set the GPIB address:
1. Select "System Config > Remote Access > GPIB Address".
2. Select "GPIB Channel Address" and enter a value between 0 and 30.
Risk of losing remote connection
If the remote access and network settings had been configured to values different to
the default, executing a factory preset via remote control terminates the connection to
the instrument.
8.8
How to Set Up a Remote Control Connection
This section guides you through the steps required to set up remote control connec-
tions of the available interfaces.
●
Chapter 8.8.1, "Establishing a Remote Control Connection over the LXI Browser
Interface"
●
Establishing a Remote Control Connection over LAN Using VXI-11 Protocol
●
Setting Up a Remote Control Connection over LAN Using Socket Communication
●
Setting Up a Remote Control Connection over GPIB
●
Setting Up a Remote Control Connection over USB
The instrument and the controller have to be connected with the suitable cable and
switched on.
A remote control program must open a connection to the instrument, before it can send
commands to and receive device responses from the instrument.
How to Set Up a Remote Control Connection