Example: Performing a power measurement with a fixed filter
Prerequisite: The sensor is connected to the instrument and mapped to the first sensor
channel.
SENSe1:SOURce RF
//Sensor measures the power of the RF signal
SENSe1:FILTer:TYPE NSRatio
//Selects fixed noise filter mode
SENSe1:FILTer:NSRatio 0.02 DB
//Sets the maximum noise component in the result to 0.02 DB
SENSe1:FILTer:NSRatio:MTIMe 10
//Limits the settling time to 10 seconds.
:SENSe1:APERture:DEFault:STATe 0
// Deactivates the default aperture time of the sensor
:SENSe1:APERture:TIMe 10e-6
// Sets the aperture time to 10 us
SENSe1:UNIT DBM
//Selects unit dBm for the measured value
:INITiate:CONTinuous ON
//Switches the continous power measurement on
:READ?
//Triggers the measurement and displays the results
........................................................................................253
:INITiate<hw>[:POWer]:CONTinuous
............................................................................... 254
............................................................................................255
:SENSe<ch>[:POWer]:APERture:DEFault:STATe
.............................................................. 255
:SENSe<ch>[:POWer]:APERture:TIMe
.............................................................................256
:SENSe<ch>[:POWer]:CORRection:SPDevice:SELect
.......................................................256
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe
........................................................256
:SENSe<ch>[:POWer]:CORRection:SPDevice:LIST?
.........................................................256
:SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?
.................................................................. 257
:SENSe<ch>[:POWer]:FILTer:LENGth[:USER]
.................................................................. 257
:SENSe<ch>[:POWer]:FILTer:NSRatio
............................................................................. 258
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe
.................................................................. 258
:SENSe<ch>[:POWer]:FILTer:SONCe
.............................................................................. 258
:SENSe<ch>[:POWer]:FILTer:TYPE
................................................................................. 259
..................................................................................259
SENSe, READ, INITiate and SLISt Subsystems