R&S
®
NRX
Power MeterUser Manual
1178556602Version 08
(;ÜÅÐ2)
www.allice.de
Allice Messtechnik GmbH
Страница 1: ...R S NRX Power Meter User Manual 1178556602 Version 08 2 www allice de Allice Messtechnik GmbH...
Страница 2: ...444 0041 02 2020 Rohde Schwarz GmbH Co KG M hldorfstr 15 81671 M nchen Germany Phone 49 89 41 29 0 Email info rohde schwarz com Internet www rohde schwarz com Subject to change data without tolerance...
Страница 3: ...ed 17 3 1 Preparing for Use 17 3 1 1 Unpacking and Checking 17 3 1 1 1 Accessory List 17 3 1 2 Operating Conditions 17 3 1 3 Considerations for Test Setup 18 3 1 4 Placing on a Bench Top 18 3 1 5 Moun...
Страница 4: ...erating Concepts 31 4 1 Manual Operation 31 4 1 1 Start Dialog 31 4 1 2 Main Measurement Dialog 32 4 1 3 Status Information 34 4 1 4 Notification Center 35 4 1 5 Selecting the Display Layout 36 4 1 6...
Страница 5: ...plays 71 7 1 Continuous Average 71 7 1 1 Continuous Average Result Display 71 7 1 2 Continuous Average Settings 72 7 2 Burst Average 73 7 2 1 Burst Average Result Display 74 7 2 2 Burst Average Settin...
Страница 6: ...guration Dialog 112 8 Sensor Configuration 114 8 1 Mode Settings 114 8 2 Correction Settings 118 8 3 Filter Settings 121 8 4 Range Settings 125 8 5 NRT Measurement Type 127 8 5 1 NRT Mode Settings 128...
Страница 7: ...rements 175 13 2 2 Preparing an Update 175 13 2 3 Updating the Application Firmware 176 14 Remote Control Commands 177 14 1 Conventions Used in SCPI Command Descriptions 177 14 2 Common Commands 177 1...
Страница 8: ...0 Timeslot 302 14 6 11 Statistics 305 14 6 11 1 Statistics Measurement Settings 305 14 6 11 2 Scaling 308 14 6 11 3 Using Markers 312 14 6 12 NRT 314 14 6 13 Querying Measurement Results 321 14 7 Calc...
Страница 9: ...14 5 Remote Settings 399 14 14 6 Managing Sensors 403 14 14 7 Instrument Information 404 14 14 8 Date and Time Settings 406 14 14 9 Notifications and Errors 408 14 14 9 1 System Event Queue 409 14 14...
Страница 10: ...x Elements 466 15 2 5 Structure of a Command Line 466 15 2 6 Responses to Queries 467 15 3 Command Sequence and Synchronization 468 15 3 1 Preventing Overlapping Execution 468 15 4 Status Reporting Sy...
Страница 11: ...8 8 Operation Upper Limit Fail Status Registers 489 16 Troubleshooting 491 16 1 Displaying Information 491 16 2 Notifications 491 16 2 1 Interpreting Notifications and Their Number 492 16 2 1 1 System...
Страница 12: ...Contents R S NRX 12 User Manual 1178 5566 02 08 www allice de Allice Messtechnik GmbH...
Страница 13: ...NRX only for its designated purpose Observe the operating con ditions and performance limits stated in the data sheet Where do I find safety information Safety information is part of the product docu...
Страница 14: ...d functions It also provides an introduction to remote control a complete description of the remote control commands with programming examples and information on maintenance instrument interfaces and...
Страница 15: ...cense texts of open source software pack ages used in the R S NRX software are provided under System Instrument Info Help Copyrights For further details see Chapter 11 2 4 Help Copyrights on page 164...
Страница 16: ...Welcome R S NRX 16 User Manual 1178 5566 02 08 See also the R S NRX fact sheet at www rohde schwarz com Key Features www allice de Allice Messtechnik GmbH...
Страница 17: ...acking material If the instrument needs to be transported or ship ped later you can use the material to protect the control elements and connectors 3 1 1 1 Accessory List The R S NRX comes with the fo...
Страница 18: ...bles of high quality For example use double shielded RF and LAN cables Always terminate open cable ends Note the EMC classification in the data sheet Do not use USB connecting cables exceeding 5 m 3 1...
Страница 19: ...any object in the vicinity Do not place the R S NRX next to heat generating equipment such as radia tors or other instruments 3 1 5 Mounting in a Rack 1 Order one of the rack adapter kits designed for...
Страница 20: ...witch at the rear to position I On After power up the R S NRX is in standby or ready state depending on the posi tion of the on standby key 2 If the R S NRX is in standby state press the on standby ke...
Страница 21: ...1 Signal source 2 R S power sensor 3 Host Interface connector 4 R S NRP ZK8 5 Sensor connector of the R S NRX 6 R S NRX Use an R S NRP ZK8 cable to connect an R S power sensor to the R S NRX If you us...
Страница 22: ...NRX check that the baud rate setting of the R S NRT Zxx power sensor is 38400 Bd See the manual of the R S NRT Zxx power sensor for details 1 Connect the R S NRT Zxx power sensor between source and l...
Страница 23: ...of available USB devices there is almost no limit to the possi ble expansions In the following useful USB devices are listed exemplarily Memory stick for easy transfer of data to from a computer e g f...
Страница 24: ...ion on page 24 If no option is installed the module bay is closed by a cover Sensor check source R S NRX B1 Used as a power reference for testing the connected power sensors and the cabling The LED of...
Страница 25: ...ger that is charged with static electricity is brought near the touch panel false triggers can occur This behavior is caused by the principle of operation of a PCAP projected capacitive touch panel Fu...
Страница 26: ...values See also Auxiliary Values on page 52 Deletes numbers or text in a field so that you can enter a new value Enter Confirms entries in text fields dialogs and selections in lists Shows a frame ar...
Страница 27: ...of the type A host USB Used to connect USB power sensors External devices like a keyboard mouse or memory stick Further information Chapter 3 1 8 4 USB 2 0 Host Interfaces on page 23 Chapter 3 1 9 Con...
Страница 28: ...rs C and D optional used to connect R S power sensors see Chapter 3 2 2 7 Sen sor Connectors C and D on page 30 3 2 2 1 Trig In Out 2 and Out 1 Trig Out Connectors See 1 in Figure 3 3 The Out 1 Trig O...
Страница 29: ...voltage The range is printed on the type label There is no need to set the voltage manually For more details see Chapter 3 1 6 Connecting to the AC Power Supply on page 19 3 2 2 6 IEC 625 IEEE 488 Int...
Страница 30: ...566 02 08 3 2 2 7 Sensor Connectors C and D See 7 Figure 3 3 Requires 3rd and 4th R S NRP sensor connector R S NRX B4 For more details see Chapter 3 2 1 1 Sensor Connector A and B on page 24 Instrumen...
Страница 31: ...st ele ments on the screen to access the settings belonging to that element topic In graphs use the following gestures Pan put your fingers on the touchscreen and move them while keeping contact Thus...
Страница 32: ...pane In the measurement pane the settings results and status of the active measurements are displayed The layout depends on the selected display layout See Chapter 4 1 5 Selecting the Display Layout o...
Страница 33: ...ivided into touch areas that lead to different settings Figure 4 2 Layout of the main measurement dialog example 1 Settings displayed in the measurement pane 2 Limit values displayed in the measuremen...
Страница 34: ...on page 50 Tapping other areas in the measurement pane can open further dialogs but these dia logs are measurement specific and there is no general rule that applies to all measure ments 4 1 3 Status...
Страница 35: ...resent the symbol of the error message is displayed The symbols used are explained in Table 4 2 The number of all messages is displayed in the color of the most severe message See 6 in Figure 4 1 Tabl...
Страница 36: ...to delete a specific notice tap the bin symbol next to the notice If you want to delete all notices tap the bin symbol in the right corner 4 1 5 Selecting the Display Layout You can split the measure...
Страница 37: ...surement panes you want to display For example if you select 2 panes the measurement display looks as follows Figure 4 3 Two measurement panes Remote command DISPlay LAYout on page 230 DISPlay WINDow...
Страница 38: ...f the panes is not changed To change the position of a measurement pane Touch hold a measurement pane and drag it into the new position The two panes have changed position In the Select Display Layout...
Страница 39: ...ters Tap a parameter to change its value Depending on the selected parameter a numeric or an alphanumeric editor is dis played The numeric editor shows for each parameter the specific value range min...
Страница 40: ...ed user identification and password see Old Password on page 159 In the Notification Center a Notice message shows the file path and name Remote control SYSTem HCOPy on page 258 Further information Ch...
Страница 41: ...nt access disable the VNC server service under VNC on page 159 Prerequisites LAN interface of the external computer is configured for the network R S NRX and the computer are connected using a LAN net...
Страница 42: ...Tour on page 23 The interfaces support the SCPI standard commands for programmable instruments standard version 1999 0 of May 1999 The SCPI standard is based on the IEEE 488 2 standard It defines a s...
Страница 43: ...e R S NRX remains in remote control until you perform one of the following actions Make sure that the R S NRX is free for you to use Press the Esc Local key See Esc Local on page 25 If the manual oper...
Страница 44: ...display to accommodate the number of measurements you want to watch simultaneously see Chapter 4 1 5 Selecting the Display Layout on page 36 If you connect more than 4 power sensors simultaneously the...
Страница 45: ...lematic setting is located See Chapter 5 5 Settings Conflict on page 48 Example Using different ports for the same sensor type 1 Connect an R S NRQ6 to port A 2 Perform a trace measurement 3 Remove th...
Страница 46: ...RQ6 is unchanged but the R S NRP33SN is assigned to measurement pane 2 Further information Chapter 14 3 Addressing Measurements and Sensors on page 186 Suffix usage in remote control 5 3 Performing a...
Страница 47: ...Tap Zero All Sensors See also Chapter 10 Zeroing Sensors on page 135 3 Configure the measurement a Open the Measurement Settings dialog as described in Chapter 4 1 2 Main Measurement Dialog on page 3...
Страница 48: ...lighted and the control elements in the hierarchies above that are leading to this setting are highlighted too Thus you can follow the problem across the hierar chies to solve the settings conflict Th...
Страница 49: ...pe that the sensor supports or assign another sen sor A setting that differs from the preset value is also indicated across the hierarchies by a pencil symbol if the visualization is enabled See Visua...
Страница 50: ...s are described in the following while mea surement specific settings are described in Chapter 7 Measurement Types and Result Displays on page 71 6 1 Display Settings Access Main measurement dialog Di...
Страница 51: ...Reflection Lower Limit State 58 Forward Lower Limit Reflection Lower Limit 58 Forward Upper Limit State Reflection Upper Limit State 59 Forward Upper Limit Reflection Upper Limit 59 Resolution Config...
Страница 52: ...mat with bar chart Graphical Available for time gate timeslot measurements Measured values are plotted over time Remote command CALCulate Measurement DMODe on page 226 Auxiliary Values Available for t...
Страница 53: ...e 235 CALCulate Measurement METer DirectionalChannel LOWer DATA RATio VALue on page 235 CALCulate Measurement METer DirectionalChannel LOWer DATA POWer on page 235 Scale Upper Limit Scaling If Display...
Страница 54: ...UPPer DATA CCDF on page 236 CALCulate Measurement METer DirectionalChannel UPPer DATA RATio RCOefficient on page 236 CALCulate Measurement METer DirectionalChannel UPPer DATA RATio RFRatio on page 237...
Страница 55: ...Culate Measurement TRACe Y SCALe TOP ONE on page 243 CALCulate Measurement TRACe Y SCALe TOP WATT on page 243 Power Div Scaling Available for trace pulse analysis time gate timeslot statistics measure...
Страница 56: ...l RATio RLOSs on page 316 CALCulate Measurement RELative DirectionalChannel RATio SWR on page 317 CALCulate Measurement RELative DirectionalChannel RATio MAGNitude on page 317 SENSe Sensor POWer REFer...
Страница 57: ...ction for the lower limit Remote command CALCulate Measurement LIMit DirectionalChannel LOWer STATe on page 253 Lower Limit Limit Monitor Available for continuous average burst average time gate times...
Страница 58: ...Ss on page 254 CALCulate Measurement LIMit DirectionalChannel UPPer DATA RATio SWR on page 255 CALCulate Measurement LIMit DirectionalChannel UPPer DATA RATio VALue on page 255 Forward Lower Limit Sta...
Страница 59: ...DATA CCDF on page 253 CALCulate Measurement LIMit DirectionalChannel UPPer DATA POWer on page 256 CALCulate Measurement LIMit DirectionalChannel UPPer DATA RATio RCOefficient on page 254 CALCulate Mea...
Страница 60: ...eans that for each partial measurement a new average value is output as a measurement result Thus the measurement result is a moving average of the last partial measurements How many of the partial me...
Страница 61: ...nal TRIGger Measurement IMMediate Internal A Inter nal B Internal C Internal D Receives the trigger signal from the trigger mas ter The trigger master is the sensor connected to port A B C or D See Tr...
Страница 62: ...B or at C Therefore set the dropout time greater than the time elapsed between points D and B and between E and C but smaller than the time elapsed between F and A Thus you ensure that triggering tak...
Страница 63: ...r 65 Jitter Suppression 65 Trigger Master State 65 Trigger Master Port 66 Trigger Synchronize State 66 Trigger Synchronize Port 66 Trigger 2 Input Impedance 66 Trigger Mode Controls the trigger execut...
Страница 64: ...played as dotted red line The trigger time is displayed as small rhomb on the trigger level line Remote command TRIGger undef ALL LEVel on page 263 TRIGger Measurement CHANnel Channel LEVel on page 26...
Страница 65: ...low the set value on a rising slope Rises above the set value on a falling slope Thus you can use this setting to eliminate the effects of noise in the signal for the edge detector of the trigger syst...
Страница 66: ...new measurement is only started after all power sensors have completed their measurements Make sure that the number of repetitions is the same for all power sensors involved in the measurement Otherwi...
Страница 67: ...escribed here apply to the continuous average burst average trace pulse analysis time gate timeslot For the other measurements see Statistics Chapter 7 7 3 Measurement Settings Dialog on page 103 NRT...
Страница 68: ...e a settings conflict is displayed See also Chapter 5 5 Settings Conflict on page 48 Remote command CALCulate Measurement CHANnel Channel SENSe INDex on page 271 SENSe Sensor CATalog on page 270 Prima...
Страница 69: ...nsor configuration Alternatively you can use Primary Sensor Config Secondary Sensor Config Parameter Set Quick Setup Selects a parameter set to preconfigure the power sensor Tap Recall Parameter Set t...
Страница 70: ...5566 02 08 Remote command SYSTem STANdard PWSettings on page 347 Recall Parameter Set Quick Setup Loads the parameters set selected under Parameter Set Remote command SYSTem STANdard PRESet on page 34...
Страница 71: ...tics 100 NRT 107 7 1 Continuous Average The power sensor measures the signal average power asynchronously within a defined time interval the so called aperture or sampling window After a trigger event...
Страница 72: ...ative Measurements is set to On or Set Sets the reference value Remote command CALCulate Measurement RELative DirectionalChannel MAGNitude on page 272 CALCulate Measurement RELative DirectionalChannel...
Страница 73: ...tive DirectionalChannel MAGNitude AUTO on page 272 Trigger See Chapter 6 3 Triggering on page 60 7 2 Burst Average The power sensor measures the average burst power of pulsed signals The burst average...
Страница 74: ...rance on page 117 Exclude from Start Exclude from End on page 117 7 2 1 Burst Average Result Display The measurement result is a single scalar value either an absolute value or related to a reference...
Страница 75: ...rement time The length of an individual measurement is deter mined from the ratio of total time and the defined number of measurement points The entire result is called a trace Each trace must be trig...
Страница 76: ...he cursor keys to expand or com press the trace Autoscale Adapts the scaling of the power axis to the trace The scaling of the time axis is adap ted only if the power sensor supports it Info Marker Sh...
Страница 77: ...Trigger Mode on page 63 Trig Source Trigger See Trigger Source on page 64 Trig Slope Trigger See Slope on page 64 Level Trigger See Trigger Level on page 64 Sensor Trigger Trigger See Chapter 6 3 Tri...
Страница 78: ...use the marker as Reference Marker Measure Shows a triangle at the marker position Remote command DISPlay WINDow Window TRACe MARKer Marker MODE on page 279 Position Mode Defines where the marker is p...
Страница 79: ...ode Remote command DISPlay WINDow Window TRACe MARKer Marker POSition TIME on page 283 DISPlay WINDow Window TRACe MARKer Marker POSition RELative TIME on page 283 DISPlay WINDow Window TRACe MARKer M...
Страница 80: ...power ratio in relation to the reference marker Marker result is p Time Diff to Ref Measures the time difference in relation to the reference marker Marker result is t Average Power to Ref Measures t...
Страница 81: ...choose which results you want to display see Chapter 7 4 3 Pulse Analysis Dialog on page 82 7 4 2 Pulse Analysis Settings Access Measurement Settings Measurement Type Pulse Analysis Display 81 Pos Sca...
Страница 82: ...w the trace The R S NRX can display a maximum of 12 measurement results If you select more a warning is displayed 100 Amplitude Time s 0 1 0 2 Pulse width Pulse period Overshoot Pulse top Pulse base U...
Страница 83: ...efaults 89 Time tab Selects the time related measurement results that are displayed below the trace The selections on this tab apply to all traces Pulse Width Time tab Time between the first positive...
Страница 84: ...the pulse in seconds where the edges occur at crossings of the mid threshold During this time the pulse remains at the pulse base level Remote command DISPlay WINDow Window TRACe MEASurement PULSe SEP...
Страница 85: ...al reference power level with falling edge referenced to the delayed trigger event Indicates the stop point of the first power pulse within the analysis window Remote command DISPlay WINDow Window TRA...
Страница 86: ...wer measured within the analysis window Remote command DISPlay WINDow Window TRACe MEASurement POWer MAXimum STATe on page 294 CALCulate Measurement TRACe MEASurement POWer MAX on page 289 Neg Oversho...
Страница 87: ...evel at High Reference Level Remote command DISPlay WINDow Window TRACe MEASurement POWer PULSe HREFerence STATe on page 295 CALCulate Measurement TRACe MEASurement POWer HREFerence on page 288 Trace...
Страница 88: ...fic Reference Levels relate to Configuration tab Selects whether the reference levels are voltage related or power related Remote command DISPlay WINDow Window TRACe MEASurement RRELation on page 298...
Страница 89: ...lse stop time Remote command CALCulate Measurement TRACe MEASurement DEFine DURation REFerence on page 287 Low Reference Level Configuration tab Defines the start of the rising edge and the end of the...
Страница 90: ...he time gate measurement type Measurement Settings Measurement Type Time Gate 2 Select the display format Display Display Format Scalar display Figure 7 3 Time gate scalar digital display The measurem...
Страница 91: ...f you have a channel calculation function running that involves 2 power sensors a legend in the upper right corner shows the assignment of curve to measurement channel In the lower pane the 4 gates ar...
Страница 92: ...ays See Resolution on page 51 Pos Scaling Available for the graphical display Defines the effect of the cursor keys on the displayed trace Pos Shifts the position Press one of the cursor keys to shift...
Страница 93: ...depends on the selected measurement type and display format Time gate scalar display Time Gate Gates Time gate graphical display Tap the lower pane where the gate information is dis played Statistics...
Страница 94: ...CALCulate Measurement POWer TGATe Gate AVG EXCLude MID STATe on page 302 Start of Fence Sets the start time of the fence The start time refers to the start of the gate Remote command CALCulate Measure...
Страница 95: ...the timeslot width to the test signal You can restrict the measurement to the rel evant part of a timeslot by excluding intervals at the start and the end of the timeslot Also you can define an exclus...
Страница 96: ...imeslot scalar digital display The measurement result is a single scalar value It refers to the selected timeslot If Scalar Analog is set as Display Format a bar chart visualizes the measurement resul...
Страница 97: ...ement values Av Average power within gate Pk Peak power within gate Cf Width Borders see t1 t2 t3 t4 t5 If you tap the lower pane the Timeslot Configuration Dialog opens 7 6 2 Timeslot Settings Access...
Страница 98: ...led Selects a border so that you can change the timeslot length and included excluded intervals The selected border is displayed as dashed line t1 Nominal Width t2 Exclude from Start t3 Exclude from E...
Страница 99: ...Nominal Width Sets the length of the timeslot Remote command CALCulate Measurement POWer TSLot AVG WIDTh on page 304 Exclude from Start Defines an interval at the start of the timeslot that is exclud...
Страница 100: ...tics Supported by wideband power sensors The power sensor measures power over time Using the statistics measurement type you can analyze the statistical distribution of the envelope power The duration...
Страница 101: ...aphical display The graph displays the waveform of the selected trace Tap Table to display the measurement results in tabular format The table contains the measurement results for 2 traces The second...
Страница 102: ...tics Timing dialog see Chapter 7 7 5 Statistics Timing Dialog on page 106 dBm dB marker Positions the x marker to a power value The associated measurement result is dis played in the lower pane see Fi...
Страница 103: ...les 104 AWGN 104 Statistics Function Sets the function used for analyzing the statistical distribution of the envelope power CCDF Complementary cumulative distribution function Probability that the en...
Страница 104: ...AWGN Enables or disables the internal additional white Gaussian noise AWGN source If enabled you cannot measure with a second power sensor Remote command CALCulate Measurement STATistics AWGN STATe o...
Страница 105: ...nt STATistics SCALe X POINts on page 310 CALCulate Measurement STATistics SCALe X RANGe on page 311 Scaling of Y Axis Sets linear or logarithmic scaling for the y axis Remote command CALCulate Measure...
Страница 106: ...t2 t3 107 Display Opens the Trace Configuration dialog that contains the scaling functions Start Time on page 54 Time Div on page 54 Power Reference on page 55 Power Div on page 55 Unit on page 55 Po...
Страница 107: ...f Pos Scaling is disabled The same gates are used in the time gate and statistics measurements See t0 t1 t2 t3 on page 92 7 8 NRT Requires the sensor interface for R S NRT R S NRX B9 see Chapter 3 2 1...
Страница 108: ...Settings Measurement Type NRT Forward 108 Average 109 CCDF 109 Peak Envelope Power PEP 109 Absorption Average 109 Crest Factor CF 110 Absorption PEP 110 Burst Average 110 Absorption Burst 110 Reflecti...
Страница 109: ...distribution of spread spectrum signals for example CDMA Remote command CALCulate Measurement CHANnel Channel FEED Channel POWer FORWard CCDFunction Peak Envelope Power PEP Forward Peak power of an a...
Страница 110: ...surement Remote command CALCulate Measurement CHANnel Channel FEED Channel POWer ABSorption PEP Burst Average Forward Average power within a burst The R S NRX determines the average burst power by mul...
Страница 111: ...Channel POWer REVerse Standing Wave Ratio SWR Reflection 1 Reflection coefficient Standing wave ratio 1 Reflection coefficient See also Reflection Coefficient on page 111 Remote command CALCulate Meas...
Страница 112: ...lay Settings on page 50 Autoscale Adapts the scaling of the graphical display Remote command SENSe Sensor POWer REFLection RANGe AUTO on page 319 SENSe Sensor POWer POWer RANGe AUTO on page 319 Trigge...
Страница 113: ...s connected to the sensor interface for R S NRT R S NRX B9 See also Primary Sensor Secondary Sensor on page 68 Trigger Mode See Trigger Mode on page 63 Trigger Source See Trigger Source on page 64 For...
Страница 114: ...he R S NRX These sensors are called primary sensor and secondary sensor Further information Primary Sensor Secondary Sensor on page 68 Chapter 7 Measurement Types and Result Displays on page 71 Chapte...
Страница 115: ...e on page 327 Duty Cycle Available for continuous average measurements Sets the duty cycle for measuring pulse modulated signals The duty cycle defines the percentage of one period during which the si...
Страница 116: ...easurements Sets the width of the sampling window The usage depends on the measurement type Continuous average measurement When measuring modulated signals the measurement can show fluctuation due to...
Страница 117: ...ation period varies or is not precisely known enable smoothing The selected sampling window has to be 5 to 9 times larger than the modulation period so that the fluctuations caused by modulation are s...
Страница 118: ...ou want to correct the influence of an amplifier connected ahead use a negative off set S Parameter corrections Used to compensate for losses and reflections introduced by a two port component that is...
Страница 119: ...ng external signal losses or gains See also Offset corrections on page 118 Remote command CALCulate Measurement CHANnel Channel CORRection OFFSet MAGNitude on page 334 S Parameter List Enables or disa...
Страница 120: ...NDex on page 333 MEMory TABLe SELect on page 393 Edit table name Primary Sensor Offsets Secondary Sensor Offsets dialogs Frequency Dependent Offset Enter the table name Remote command MEMory TABLe MAP...
Страница 121: ...y Sensor Config Filter Measurement Settings Secondary Sensor Config Filter Available for all measurement types Use the averaging filter to reduce fluctuations in the measurement results Such fluctu at...
Страница 122: ...nabled the number of measured values is averaged Averaging reduces the effect of noise so that more reliable results are obtained Remote command CALCulate Measurement CHANnel Channel AVERage STATe on...
Страница 123: ...the autofilter Normal Sets the averaging number so that the intrinsic noise of the power sensor 2 standard deviations does not exceed the specified Noise Content on page 123 Fixed Noise Limits the av...
Страница 124: ...average burst average trace pulse analysis timeslot statis tics measurements Defines how the measurement results are output This is called termination control On Outputs intermediate values to facilit...
Страница 125: ...ettings Access Measurement Settings Primary Sensor Config Range Measurement Settings Secondary Sensor Config Range Available for continuous average burst average trace pulse analysis time gate time sl...
Страница 126: ...nge Available if Range State is set to User Sets the active measurement path in which the power sensor is measuring Remote command CALCulate Measurement CHANnel Channel POWer AVG RANGe VALue on page 3...
Страница 127: ...input attenuation Once Adjusts the input attenuation one time then disables the automatic setting Remote command CALCulate Measurement CHANnel Channel INPut ATTenuation AUTO on page 343 Attenuation Re...
Страница 128: ...ors The power sensor automatically recognizes the duty cycle of the burst series and calculates the average burst power from this duty cycle and the average power Set an appropriate Video Bandwidth Us...
Страница 129: ...2 or from port 2 to port 1 The two ports are indicated on the directional power sensor Remote command CALCulate Measurement CHANnel Channel NRT DIRection on page 363 INPut Sensor PORT SOURce AUTO on p...
Страница 130: ...RPLane on page 366 INPut Sensor PORT POSition on page 367 Offset Considers the transmission loss in a cable that connects the desired measurement point set by Offset Reference Plane and the power sen...
Страница 131: ...Sensor DM STATe on page 365 SENSe Sensor DM STANdard on page 366 WCDMA Chip Rate Available if WCDMA is set under Modulation Sets the chip rate for the WCDMA communication standard Remote command CALC...
Страница 132: ...CHANnel Channel NRT APERture MODE on page 369 Integration Time Available if Integration Time Mode is set to User Defines the integration time for a single measurement Remote command CALCulate Measure...
Страница 133: ...page 134 If you want to save specific measurement settings to reuse at another time save the setup in a file The R S NRX offers 20 setup files for this purpose Access Preset Save Recall Preset dialog...
Страница 134: ...ttings Conflict on page 48 For details on sensor settings see the user manual of the R S power sensor Remote command SYSTem PRESet on page 394 RST on page 183 Save Recall Setup Selects the setup file...
Страница 135: ...ss than 10 dB above the lower measurement range limit Access Zero Zeroing Sensors dialog Figure 10 1 Zeroing Sensors dialog The table below Zero All Sensors shows all connected sensors with Port name...
Страница 136: ...can zero an individual sensor or all sensors at once Tap Zero All Sensors Tap in the row of the sensor you want to zero The status changes from in progress to successful Remote command See Chapter 14...
Страница 137: ...view dialog is divided into the following tabs Connections 137 Instrument Info 153 Hardware Configuration 165 Test 166 Global Settings 167 11 1 Connections Access System Connections See Figure 11 1 On...
Страница 138: ...ic TCP IP configuration assign a valid address information before connecting the R S NRX to the LAN Contact your network administrator to obtain valid IP addresses After integrating the R S NRX into a...
Страница 139: ...e the R S NRX restarts its connection to the network which can take several seconds During this time you cannot address the R S NRX After the restart you can only address the R S NRX using the newly s...
Страница 140: ...ork supports the dynamic host configuration protocol DHCP Static Enables assigning the IP address manually Remote command SYSTem COMMunicate NETWork IPADdress MODE on page 398 SYSTem COMMunicate INET...
Страница 141: ...COMMunicate NETWork IPADdress GATeway on page 398 SYSTem COMMunicate INET SELF GATeway ADDRess on page 398 DNS Server IPv4 tab Available if Static is set under Address Mode Sets the DNS server address...
Страница 142: ...a LAN the VISA resource string is required to establish a communication session between the controller and the R S NRX The resource string is a unique identifier composed of the specific IP address of...
Страница 143: ...S NRX based on the standard commands for programmable instruments SCPI 99 Rohde Schwarz NRP2 Rohde Schwarz NRP Emulation for a predecessor Keysight N432A Keysight 1911A Keysight N1912A Keysight E4418...
Страница 144: ...tions tab Available if Customization of IDN is set to User Sets the customized instrument identification string so that you can identify each R S NRX individually Remote command SYSTem IDN ANSWer on p...
Страница 145: ...ure 11 2 Sensor Check Source tab In the upper pane the result display shows the effect of parameter changes on the measurement Signal Output Sensor Check Source tab Disables the output or sets the sig...
Страница 146: ...ated automatically If a new version is available a warning message is displayed in the notification center and the new version is displayed here as shown in Figure 11 3 Tap the info field to update th...
Страница 147: ...nal to the displayed value Limit Violation Available for continuous average burst average time gate timeslot measure ments Out 1 Trig Out BNC connector I O 1 tab Sets the fail voltage that is output i...
Страница 148: ...ble if Mode is set to Analog Out Forw Analog Out Refl Analog Out Enter the measurement value that corresponds to 0 V output voltage Remote command OUTPut RECorder output LIMit LOWer CCDF on page 375 O...
Страница 149: ...1 I O 2 tabs Available if Mode is set to Limit Violation Forw Limit Violation Refl Limit Violation Sets the fail voltage that is output if a measured value causes a limit violation Low 0 V High 3 3 V...
Страница 150: ...ectly or indirectly by USB hub They are assigned the letters E to M R S power sensors that are connected to the LAN interface are not recognized auto matically Add them see To add a LAN power sensor o...
Страница 151: ...Sensor Test 152 Add Sensor Adds a LAN power sensor See To add a LAN power sensor on page 150 Remote command SENSe Sensor ADD on page 403 Sensor Info Access System Connections Sensor Manager Displays...
Страница 152: ...nfo Tap Start Test to start a selftest of the connected power sensor The selftest provides detailed information that you can use for troubleshooting Test Verdict Shows the status of the selftest Remot...
Страница 153: ...cted power sensor see Sensor Info on page 151 On this tab you display and configure the following settings System Info 153 Security Settings 156 Option Settings 161 Help Copyrights 164 11 2 1 System I...
Страница 154: ...on CPLD Version Complex programmable logic device CPLD version Serial Device ID SW Build Version of software build Options Short names of the installed options MAC Address Ethernet hardware address Ho...
Страница 155: ...mat YYYY MM DD Remote command SYSTem DATE on page 406 SYSTem DATE UTC on page 406 SYSTem DATE LOCal on page 406 Time Date and Time Settings Sets the time in the format HH MM SS Remote command SYSTem T...
Страница 156: ...ab 156 Security Password to Unlock Settings 157 USB Storage 157 Volatile Mode 157 Sanitize 157 User Interface 157 LAN Settings tab 158 LAN Services 158 SCPI over LAN 158 Web Server 158 VNC 159 Avahi Z...
Страница 157: ...want to reboot with a defined configuration for a measure ment setup regardless off any settings made manually or by remote control Enabling the volatile mode requires the security password If you cha...
Страница 158: ...rized reading If you use the R S NRX in a remote operation session you can set Display Only or Disabled to make sure that only authorized persons can interrupt the remote session LAN Settings tab Conf...
Страница 159: ...tings tab Enables or disables access using a secure shell SSH a network protocol for secure data communication Software Update LAN Settings tab Enables or disables the software update over LAN Instrum...
Страница 160: ...y used security password The preconfigured password is 123456 Note We recommend that you change the preconfigured password before connecting the R S NRX to a network The security password is required...
Страница 161: ...Manage License Keys tab 162 Enter License Key 163 Import 163 Export 163 Details tab 163 HW Options tab Displays the installed hardware options The hardware options on the front panel you can install...
Страница 162: ...version but are not installed Show Deacti vated On Off Shows or hides deactivated software options These software options have been installed but are not active any more for example because the licens...
Страница 163: ...lled options Enter License Key Manage License Keys tab Enter the license key manually Import Manage License Keys tab For future use Export Manage License Keys tab For future use Details tab Displays a...
Страница 164: ...want to see more information on a specific option tap 11 2 4 Help Copyrights Access System Instrument Info Help Copyrights Gives access to the user manual open source acknowledgement and license info...
Страница 165: ...he down load is displayed in the Notification Center See also Chapter 4 1 4 Notification Cen ter on page 35 Open Source Achnowledgement Displays the open source acknowledgement Open Source Licenses Di...
Страница 166: ...or look ing up the revision of hardware for example when troubleshooting 11 4 Test Access System Test On this tab you can test whether the user interfaces are in working order and create information u...
Страница 167: ...ating information for troubleshooting You can save information for troubleshooting on a memory stick 1 Connect a memory stick to one of the USB interfaces 2 Tap Create R S Support Information The crea...
Страница 168: ...o Thus you can find the setting easily if you want to use a preset value Tabs Position Specifies the position of the tabs in dialogs top or bottom Hide Sensor Overload Message If enabled hides the sen...
Страница 169: ...atory steps 1 Make sure that the most recent firmware version is installed 2 Check the License Keys List whether the license you have purchased is unregis tered If the license is unregistered you need...
Страница 170: ...on nection This chapter contains information on installing updating the firmware on the R S NRX via PC and USB or Ethernet connection Use the Firmware Update program PureFW to load new firmware for th...
Страница 171: ...in vain to find a USB driver for the instrument If this happens the instrument is highlighted by a yellow excla mation mark in the Windows device manager Abort the installation process and install a...
Страница 172: ...pdate The following window should appear The program automatically starts scanning for R S power sensors and meters attached via USB When the scan is completed all recognized power sensors and meters...
Страница 173: ...A network devices automati cally This can be more time consuming than adding the device manually as described above To do this check the setting Including VISA Network Devi ces before you press Rescan...
Страница 174: ...he ellipsis button to browse the file system for it New firmware for the R S NRX generally has an rsu Rohde Schwarz Update extension 5 Select Update to download the new firmware and program it into th...
Страница 175: ...stem and software that supports copying files to the USB flash memory stick USB flash memory stick USB 2 0 or 3 0 with USB A plug FAT32 file system and sufficient space for the firmware file A Rohde S...
Страница 176: ...than one matching Rohde Schwarz update files select the file you want to use for the update The latest version is on top Then press Update to start the update process 3 After copying the Rohde Schwarz...
Страница 177: ...as SCPI con firmed All commands used by the R S NRX follow the SCPI syntax rules Asynchronous commands A command which does not automatically finish executing before the next com mand starts executin...
Страница 178: ...SE 180 ESR 180 GCLS 181 GMC 181 GOPC 181 GWAI 181 IDN 181 IST 181 LMC 182 OPC 182 OPT 182 PMC 182 PRE 182 PSC 183 RCL 183 RMC 183 RST 183 SAV 183 SRE 184 SRQ 184 STB 184 TRG 184 TST 184 WAI 185 XESE 1...
Страница 179: ...e Event GTM Go to local with remote state Usage Event GTR Go to remote Usage Event HFC Hardware flow control Usage Event LLO Local lockout Usage Event NREN Not remote enabled go to local Usage Event C...
Страница 180: ...arameters Macro Parameters for setting and query Label EMC Enable Enables macro command Parameters Enable ESE register Event status enable Sets the event status enable register to the specified value...
Страница 181: ...ons in all internal instruments are finished Usage Query only GWAI Waits for all pending operations in all internal instruments Usage Event IDN Identification Returns a string with information on the...
Страница 182: ...urns a response The response is an advantage because you can query the execution of commands from a controller pro gram before sending new commands Thus preventing overflow of the input queue when too...
Страница 183: ...34 RMC Label Remove macro content Setting parameters Label Usage Setting only RST Sets the instrument to a defined initial state a so called reset The default settings are indicated in the description...
Страница 184: ...meout A generic srq wait command to be used without srq event transport It is sim ply read from the interface Query parameters timeout Return values srq Usage Query only STB Status byte Returns the co...
Страница 185: ...er ESE This register determines which events from the standard event status register ESR are summarized in bit 5 the event summary bit ESB of the status byte Parameters xese expr XESR Reads and clears...
Страница 186: ...e used The first power sensor is connected to port A It is assigned to the first measurement addressed by CALC1 The second power sensor is connected to port C It is assigned to the third measure ment...
Страница 187: ...list but does not start a measurement The query without parameters for example CONF returns the parameters trans fered the last time Since the instrument settings can be changed after sensing a CONFig...
Страница 188: ...ected_value_or_source_list resolution_or_source_list source_list READ Measurement SCALar POWer AVG expected_value_or_source_list resolution_or_source_list source_list MEASure Measurement SCALar POWer...
Страница 189: ...numeric expr source_list expr For the parameter descriptions see Parameter list on page 188 Usage Query only CONFigure Measurement SCALar POWer AVG DIFFerence expected_value_or_source_list resolution_...
Страница 190: ...ery parameters expected_value_or_source_list numeric expr resolution_or_source_list numeric expr source_list expr For the parameter descriptions see Parameter list on page 188 Usage Query only CONFigu...
Страница 191: ...rs Suffix Measurement 1 to 8 Measurement Query parameters expected_value_or_source_list numeric expr resolution_or_source_list numeric expr source_list expr For the parameter descriptions see Paramete...
Страница 192: ...rce_list Relative ratio measured by two power sensor Suffix Measurement 1 to 8 Measurement Query parameters expected_value_or_source_list numeric expr resolution_or_source_list numeric expr source_lis...
Страница 193: ...e_list source_list Reflection coefficient transmission factor of a DUT measured by two power sensors Suffix Measurement 1 to 8 Measurement Query parameters expected_value_or_source_list numeric expr r...
Страница 194: ...tory Number of requested measured values Corresponds to SENSe Sensor POWer AVG BUFFer SIZE expected_value Optional Value that is expected for the measurement resolution Optional Limit up to which the...
Страница 195: ...er AVG DIFFerence buffered_size expected_value_or_source_list resolution_or_source_list source_list FETCh Measurement ARRay POWer AVG DIFFerence buffered_size expected_value_or_source_list resolution_...
Страница 196: ...rameters buffered_size expr expected_value_or_source_list numeric expr resolution_or_source_list numeric expr source_list expr For the parameter descriptions see Parameter list on page 194 Usage Query...
Страница 197: ...y two power sensor with buffering Suffix Measurement 1 to 8 Measurement Query parameters buffered_size expr expected_value_or_source_list numeric expr resolution_or_source_list numeric expr source_lis...
Страница 198: ...d_size expected_value_or_source_list resolution_or_source_list source_list MEASure Measurement ARRay POWer AVG RELative buffered_size expected_value_or_source_list resolution_or_source_list source_lis...
Страница 199: ...ure Measurement ARRay POWer AVG SUM buffered_size expected_value_or_source_list resolution_or_source_list source_list FETCh Measurement ARRay POWer AVG SUM buffered_size expected_value_or_source_list...
Страница 200: ..._or_source_list numeric expr resolution_or_source_list numeric expr source_list expr For the parameter descriptions see Parameter list on page 194 Usage Query only CONFigure Measurement ARRay POWer AV...
Страница 201: ...ptured in the trace measure ments Corresponds to SENSe Sensor TRACe TIME on page 245 source_list Mandatory Primary and secondary sensor The number for each sensor is prece ded by the character The ent...
Страница 202: ...arameter list on page 201 Usage Setting only CONFigure Measurement XTIMe POWer RATio scope_size capture_time source_list FETCh Measurement XTIMe POWer RATio scope_size capture_time source_list READ Me...
Страница 203: ...onal Limit up to which the measurement result should be free of noise Corresponds to SENSe Sensor AVERage COUNt AUTO RESolution source_list Optional Primary and secondary sensor The number for each se...
Страница 204: ...ted_value_or_source_list resolution_or_source_list source_list READ Measurement SCALar POWer BURSt RELative dtolerance start_exclude end_exclude expected_value_or_source_list resolution_or_source_list...
Страница 205: ...t s start_exclude Default unit s end_exclude Default unit s expected_value_or_source_list numeric expr resolution_or_source_list numeric expr source_list expr For the parameter descriptions see Parame...
Страница 206: ...urement SCALar POWer BURSt SUM dtolerance start_exclude end_exclude expected_value_or_source_list resolution_or_source_list source_list READ Measurement SCALar POWer BURSt SUM dtolerance start_exclude...
Страница 207: ...t SCALar POWer BURSt SUM RELative dtolerance start_exclude end_exclude expected_value_or_source_list resolution_or_source_list source_list Sum measured by two power sensors Suffix Measurement 1 to 8 M...
Страница 208: ...ist on page 203 Usage Query only CONFigure Measurement SCALar POWer BURSt RATio RELative dtolerance start_exclude end_exclude expected_value_or_source_list resolution_or_source_list source_list FETCh...
Страница 209: ..._source_list source_list READ Measurement SCALar POWer BURSt SWR dtolerance start_exclude end_exclude expected_value_or_source_list resolution_or_source_list source_list MEASure Measurement SCALar POW...
Страница 210: ...Query parameters dtolerance Default unit s start_exclude Default unit s end_exclude Default unit s expected_value_or_source_list numeric expr resolution_or_source_list numeric expr source_list expr Fo...
Страница 211: ...the measurement Chapter 7 6 Timeslot on page 94 Parameter list For timeslot measurements the following parameters are used tslot_width Mandatory Width of a timeslot Corresponds to SENSe Sensor POWer...
Страница 212: ...ALar POWer TSLot tslot_width no_slots start_exclude end_exclude expected_value_or_source_list resolution_or_source_list source_list READ Measurement SCALar POWer TSLot tslot_width no_slots start_exclu...
Страница 213: ...Default unit s no_slots Default unit s start_exclude Default unit s end_exclude Default unit s expected_value_or_source_list numeric expr resolution_or_source_list numeric expr source_list expr For th...
Страница 214: ...POWer TSLot DIFFerence RELative tslot_width no_slots start_exclude end_exclude expected_value_or_source_list resolution_or_source_list source_list READ Measurement SCALar POWer TSLot DIFFerence RELati...
Страница 215: ...t SCALar POWer TSLot SUM tslot_width no_slots start_exclude end_exclude expected_value_or_source_list resolution_or_source_list source_list MEASure Measurement SCALar POWer TSLot SUM tslot_width no_sl...
Страница 216: ...eters tslot_width Default unit s no_slots Default unit s start_exclude Default unit s end_exclude Default unit s expected_value_or_source_list numeric expr resolution_or_source_list numeric expr sourc...
Страница 217: ...SCALar POWer TSLot RATio RELative tslot_width no_slots start_exclude end_exclude expected_value_or_source_list resolution_or_source_list source_list READ Measurement SCALar POWer TSLot RATio RELative...
Страница 218: ...POWer TSLot SWR tslot_width no_slots start_exclude end_exclude expected_value_or_source_list resolution_or_source_list source_list MEASure Measurement SCALar POWer TSLot SWR tslot_width no_slots star...
Страница 219: ...ters tslot_width Default unit s no_slots Default unit s start_exclude Default unit s end_exclude Default unit s expected_value_or_source_list numeric expr resolution_or_source_list numeric expr source...
Страница 220: ...ut the measurement Chapter 7 7 Statistics on page 100 Parameter list For the calculation functions of the statistics measurement the following parameters are used statistics_size Mandatory Number of t...
Страница 221: ...source_list expr For the parameter descriptions see Parameter list on page 220 Usage Query only CONFigure Measurement STATistics PDF statistics_size capture_time source_list FETCh Measurement STATisti...
Страница 222: ...ts the current measurement If the measurement has been started as a single measurement INITiate Measurement IMMediate the power sensor goes into the idle state However if a continuous measurement is i...
Страница 223: ...measurement is completed the power sensor does not return to the idle state but enters the waiting for trigger state again OFF Ends the continuous measurement mode and sets the power sensor to the id...
Страница 224: ...l times Once the entire measurement is completed a measurement result is available and the power sensor enters the idle state again Use the command only after the continuous measurement mode has been...
Страница 225: ...CALCulate Measurement subsystem Table 14 1 Assigned measurement panes and traces Measurement suffix CALCulate Measurement Assigned measurement panes WINDow Window Assigned traces CALC1 to CALC4 WIND1...
Страница 226: ...ISPlay ERRorlist 229 DISPlay LAYout 230 DISPlay MESSage TEXT CLEar 230 DISPlay MESSage TEXT DATA 230 DISPlay MESSage TYPE 230 DISPlay MESSage STATe 230 DISPlay OVERload STATe 231 DISPlay PIXMap 231 DI...
Страница 227: ...mat RST SDIGital Manual operation See Display Format on page 52 See Info Marker on page 76 See Graph Table on page 102 CALCulate Measurement EXTRemes RESet Saves the currently measured value as the ne...
Страница 228: ...or each point pixel of the selected display type Alias CALCulate Measurement LIMit undef STATe Suffix Measurement 1 to 8 Measurement Parameters state OFF ON RESet RST OFF Manual operation See Max Hold...
Страница 229: ...cimal places e g 1 dBm OI 1 decimal place e g 0 1 dBm OOI 2 decimal places e g 0 01 dBm OOOI 3 decimal places e g 0 001 dBm RST OOI Manual operation See Resolution on page 51 DISPlay BRIGthness bright...
Страница 230: ...max number of lines depends on the message type 9 lines for messages and 7 lines for queries The length of a line depends on the characters used Too long lines are cut off DISPlay MESSage TYPE type S...
Страница 231: ...Queries the display content The return value is a binary block data for example 577110xxxxxx x 577110 block data header xxxxxx x binary format comprising an 8 bit BMP bitmap of the display content Usa...
Страница 232: ...o 234 CALCulate Measurement METer DirectionalChannel LOWer DATA RATio RLOSs 234 CALCulate Measurement METer DirectionalChannel LOWer DATA RATio SWR 235 CALCulate Measurement METer DirectionalChannel L...
Страница 233: ...44 DISPlay WINDow Window ANALog UPPer 244 DISPlay WINDow Window METer UPPer 244 SENSe Sensor TRACe OFFSet TIME 244 SENSe Sensor TRACe TIME 245 CALCulate Measurement METer DirectionalChannel LOWer DATA...
Страница 234: ...ection reverse Parameters value Range 0 0 to 100 0 RST 0 0 Default unit pct Manual operation See Scale Lower Limit on page 53 See Forward Scale Lower Limit Reflection Scale Lower Limit on page 53 CALC...
Страница 235: ...o bargraph display If you enter a value without unit the unit is defined by UNIT Measurement POWer RATio For further information see Chapter 14 6 1 3 Units on page 245 Suffix Measurement 1 to 8 Measur...
Страница 236: ...for the CCDF bargraph display Suffix Measurement 1 to 8 Measurement DirectionalChannel 1 to 2 1 primary sensor 2 secondary sensor or 1 forward 2 reflection reverse Parameters value Range 0 0 to 100 0...
Страница 237: ...arameters value Range 0 0 to 100 0 RST 100 0 Default unit pct Manual operation See Scale Upper Limit on page 53 See Forward Scale Upper Limit Reflection Scale Upper Limit on page 54 CALCulate Measurem...
Страница 238: ...o bargraph display If you enter a value without unit the unit is defined by UNIT Measurement POWer RATio For further information see Chapter 14 6 1 3 Units on page 245 Suffix Measurement 1 to 8 Measur...
Страница 239: ...e delayed trigger The value can be negative so that signal components are displayed before the trigger event Suffix Measurement 1 to 8 Measurement Parameters value Range 15 0 to 15 0 RST 0 0 Default u...
Страница 240: ...400 0 RST 50 0 Default unit dBm Manual operation See Power Span on page 55 CALCulate Measurement TRACe Y SCALe SPAN DBUV value Effective for trace pulse analysis time gate timeslot measurements Sets t...
Страница 241: ...eration See Power Span on page 55 CALCulate Measurement TRACe Y SCALe SPAN WATT value Effective for trace pulse analysis time gate timeslot measurements Sets the power level range Suffix Measurement 1...
Страница 242: ...Power Reference on page 55 CALCulate Measurement TRACe Y SCALe TOP DBUV value Effective for trace pulse analysis time gate timeslot measurements Sets the power reference value The reference value is a...
Страница 243: ...Power Reference on page 55 CALCulate Measurement TRACe Y SCALe TOP WATT value Effective for trace pulse analysis time gate timeslot measurements Sets the power reference value The reference value is a...
Страница 244: ...cale Suffix Window 1 to 4 Measurement pane Parameters value See DISPlay WINDow Window METer LOWer on page 244 SENSe Sensor TRACe OFFSet TIME time Adds an offset to the beginning of the trace sequence...
Страница 245: ...ratio that can be expressed in more than one unit you can enter the value together with the unit and the unit is recognized If you enter a value without unit the unit defined by one of the following...
Страница 246: ...surement POWer RATio unit Sets the output unit for the measured power ratio values For NRT measurements sets the unit of the relative forward power measurement Suffix Measurement 1 to 8 Measurement Pa...
Страница 247: ...Measurement POWer RELative STATe state Alias for CALCulate Measurement RELative DirectionalChannel STATe on page 272 ON corresponds to ON and SET Suffix Measurement 1 to 8 Measurement Parameters stat...
Страница 248: ...Measurement LIMit DirectionalChannel UPPer DATA CCDF 253 CALCulate Measurement LIMit DirectionalChannel UPPer DATA RATio RCOefficient 254 CALCulate Measurement LIMit DirectionalChannel UPPer DATA RAT...
Страница 249: ...ted Suffix Measurement 1 to 8 Measurement undef 1 to n No suffix required Usage Query only CALCulate Measurement LIMit undef FCOunt Queries the number of limit violations The counter is reset if one o...
Страница 250: ...ary sensor 2 secondary sensor or 1 forward 2 reflection reverse Parameters value Range 0 0 to 100 0 RST 0 0 Default unit pct Manual operation See Forward Lower Limit Reflection Lower Limit on page 58...
Страница 251: ...it DirectionalChannel LOWer DATA RATio RLOSs value Sets the lower limit for the return loss Suffix Measurement 1 to 8 Measurement DirectionalChannel 1 to 2 1 primary sensor 2 secondary sensor or 1 for...
Страница 252: ...ary sensor 2 secondary sensor or 1 forward 2 reflection reverse Parameters value Range 180 0 to 180 0 RST 20 0 Default unit dB Manual operation See Lower Limit on page 57 See Forward Lower Limit Refle...
Страница 253: ...t State Reflection Lower Limit State on page 58 CALCulate Measurement LIMit DirectionalChannel UPPer DATA value Sets the upper limit for the measured values Suffix Measurement 1 to 8 Measurement Direc...
Страница 254: ...Manual operation See Upper Limit on page 58 See Forward Upper Limit Reflection Upper Limit on page 59 CALCulate Measurement LIMit DirectionalChannel UPPer DATA RATio RFRatio value Sets the upper limi...
Страница 255: ...ectionalChannel 1 to 2 1 primary sensor 2 secondary sensor or 1 forward 2 reflection reverse Parameters value Range 0 0 to 1e18 RST 2 0 Default unit Manual operation See Upper Limit on page 58 See For...
Страница 256: ...8 Measurement DirectionalChannel 1 to 2 1 primary sensor 2 secondary sensor or 1 forward 2 reflection reverse Parameters value Range 120 0 to 150 0 RST 10 0 Default unit dBm Manual operation See Upper...
Страница 257: ...omit the length the R S NRX uses the last used length Example for REAL 32 format 14 binary float value Example for REAL 64 format 18 binary float value arg1 ASCii 0 to 12 Readable value The digit def...
Страница 258: ...9 TRIGger undef ALL COUNt 259 TRIGger Measurement CHANnel Channel COUNt 259 TRIGger undef ALL DELay AUTO 260 TRIGger Measurement CHANnel Channel DELay AUTO 260 TRIGger undef ALL DELay VALue 260 TRIGge...
Страница 259: ...ment 1 to 8 Measurement Channel 1 to 2 1 primary sensor 2 secondary sensor Parameters stat ON OFF RST ON TRIGger undef ALL COUNt count TRIGger Measurement CHANnel Channel COUNt count Sets the number o...
Страница 260: ...easurement 2 for example enable the R S NRP2 emulation using SYSTem LANGuage NRP2 Suffix Measurement 1 to 8 Measurement Channel 1 to 2 1 primary sensor 2 secondary sensor Parameters stat ON OFF RST OF...
Страница 261: ...Channel 1 to 2 1 primary sensor 2 secondary sensor Parameters dropout Range 0 0 to 10 0 RST 0 0 Default unit s Manual operation See Dropout on page 65 TRIGger Measurement CHANnel Channel EXTernal Por...
Страница 262: ...HYSTeresis hysteresis TRIGger Measurement CHANnel Channel HYSTeresis hysteresis Sets the hysteresis A trigger event occurs if the trigger level Falls below the set value on a rising slope Rises above...
Страница 263: ...hannel LEVel level Effective only if TRIGger Measurement CHANnel Channel SOURce INTernal Sets the trigger threshold for internal triggering derived from the test signal If you enter a value without un...
Страница 264: ...ous triggering so that only one trigger event at a time is executed AUTO Automatically starts a measurement if no trigger event has occurred after 300 ms RST AUTO Manual operation See Trigger Mode on...
Страница 265: ...If port 1 is assigned to measure ment 1 and so on the command works If port 1 is assigned to measurement 2 for example enable the R S NRP2 emulation using SYSTem LANGuage NRP2 Suffix Measurement 1 to...
Страница 266: ...hapter 6 3 2 Trigger Sources on page 61 IMMediate Measures immediately does not wait for trigger condition RST INTernal Manual operation See Trigger Source on page 64 TRIGger Measurement CHANnel Chann...
Страница 267: ...tion using SYSTem LANGuage NRP2 Suffix Measurement 1 to 8 Measurement Channel 1 to 2 1 primary sensor 2 secondary sensor Parameters state ON OFF RST OFF Manual operation See Trigger Synchronize State...
Страница 268: ...o 4 selects the time gate POWer AVERage ON SWEep 1 to 4 POWer PTAVerage ON SWEep 1 to 4 POWer PEAK ON SWEep 1 to 4 Average value in time gate 1 2 3 or 4 Peak to average value in time gate 1 2 3 or 4 P...
Страница 269: ...t only if the first channel suffix is set to 1 Parameters string The availability depends on the measurement type see Table 14 2 RST POWer AVERage Manual operation See Statistics Function on page 103...
Страница 270: ...DMax NONE No additional values are measured MINMax Minima and maxima of the trace are transmitted together with the measured value Usually extreme values are lost due to averaging the measured values...
Страница 271: ...of the sensors that are connected to a sensor port of the R S NRX You can query the connected power sen sors using SENSe Sensor CATalog Suffix Measurement 1 to 8 Measurement Channel 1 to 2 1 primary s...
Страница 272: ...Reflection Reference Value on page 56 See Reference Value on page 72 CALCulate Measurement RELative DirectionalChannel MAGNitude AUTO state Alias for CALCulate Measurement RELative DirectionalChannel...
Страница 273: ...enables the relative measurement RST OFF Manual operation See Forward Relative State Reflection Relative State on page 56 See Relative Measurements on page 73 SENSe Sensor POWer REFerence ref Sets th...
Страница 274: ...OWer AVG BUFFer CLEar Effective for continuous average measurements Clears the contents of the result buffer Suffix Sensor 1 to 128 Configured sensor connected at port A 1 port D 4 R S NRX B9 101 USB...
Страница 275: ...continuous average measurement If enabled all results generated by trigger events are collected until the buffer is filled Thus a higher data rate is achieved You can set the size of the buffer with...
Страница 276: ...and SENSe Sensor BWIDth VIDeo Suffix Sensor 1 to 128 Configured sensor connected at port A 1 port D 4 R S NRX B9 101 USB and LAN port 5 to 100 Usage Query only SENSe Sensor TRACe REALtime state Effec...
Страница 277: ...lay WINDow Window TRACe MARKer Marker POSition POWer RATio DPCT 282 DISPlay WINDow Window TRACe MARKer Marker POSition RELative POWer O 282 DISPlay WINDow Window TRACe MARKer Marker POSition POWer RAT...
Страница 278: ...y WINDow Window TRACe MARKer Marker MODE MEASure is set Selects the trace Suffix Window 1 to 4 Measurement pane Marker 1 to 4 Marker M1 to M4 Parameters index 0 No trace selected 1 Trace 1 2 Trace 2 R...
Страница 279: ...NDow Window TRACe MARKer Marker MODE mode Enables or disables the marker Also defines the appearance of the marker Suffix Window 1 to 4 Measurement pane Marker 1 to 4 Marker M1 to M4 Parameters mode O...
Страница 280: ...e marker RPRight Starting from the right border at a power difference of the marker position to the y position of the reference marker PSEarch Measured maximum power MSEarch Measured minimum power RPS...
Страница 281: ...arker M1 to M4 Parameters power Range 100 0 to 300 0 RST 0 0 Default unit dBuV Manual operation See Position on page 79 DISPlay WINDow Window TRACe MARKer Marker POSition RELative POWer DB power DISPl...
Страница 282: ...OSition RELative POWer O power DISPlay WINDow Window TRACe MARKer Marker POSition POWer RATio O power Sets an absolute or relative value for the marker position defined under DISPlay WINDow Window TRA...
Страница 283: ...for the marker position defined under DISPlay WINDow Window TRACe MARKer Marker POSition MODE on page 279 Suffix Window 1 to 4 Measurement pane Marker 1 to 4 Marker M1 to M4 Parameters time Range 15 0...
Страница 284: ...urement TRACe MEASurement POWer LREFerence 289 CALCulate Measurement TRACe MEASurement POWer MAX 289 CALCulate Measurement TRACe MEASurement POWer MIN 289 CALCulate Measurement TRACe MEASurement POWer...
Страница 285: ...TATe 298 DISPlay WINDow Window TRACe MEASurement TRANsition NEGative OCCurrence STATe 298 DISPlay WINDow Window TRACe MEASurement TRANsition NEGative OVERshoot STATe 299 DISPlay WINDow Window TRACe ME...
Страница 286: ...set to the start of the gate in which the pulse analysis is performed See also SENSe Sensor TRACe MID TIME on page 286 Suffix Sensor 1 to 128 Configured sensor connected at port A 1 port D 4 R S NRX B...
Страница 287: ...Algorithm on page 88 CALCulate Measurement TRACe MEASurement DEFine DURation REFerence value Effective for pulse analysis measurements Sets the pulse width pulse start time and pulse stop time Suffix...
Страница 288: ...Default unit pct Manual operation See Low Reference Level on page 89 CALCulate Measurement TRACe MEASurement POWer AVG Effective for pulse analysis measurements Queries the average power during the t...
Страница 289: ...o 8 Measurement Usage Query only Manual operation See Trace Peak on page 86 CALCulate Measurement TRACe MEASurement POWer MIN Effective for pulse analysis measurements Queries the minimum power measur...
Страница 290: ...REFerence Effective for pulse analysis measurements Queries the power level at reference level See CALCulate Measurement TRACe MEASurement DEFine DURation REFerence on page 287 Suffix Measurement 1 to...
Страница 291: ...Effective for pulse analysis measurements Queries the time between the first negative edge and the subsequent positive edge of the pulse in seconds where the edges occur at crossings of the mid thres...
Страница 292: ...ctive for pulse analysis measurements Queries the height of the local minimum before a rising edge divided by the pulse amplitude Negative overshoot 100 x Pulse base power minimum power Pulse amplitud...
Страница 293: ...nt TRANsition POSitive OVERshoot Effective for pulse analysis measurements Queries the height of the local maximum before a falling edge divided by the pulse amplitude Positive overshoot 100 x Max pow...
Страница 294: ...nts Enables or disables the display of the maximum power measured within the analysis window See CALCulate Measurement TRACe MEASurement POWer MAX on page 289 Suffix Window 1 to 4 Measurement pane Par...
Страница 295: ...lysis measurements Enables or disables the display of the power level at high reference level See CALCulate Measurement TRACe MEASurement POWer HREFerence on page 288 Suffix Window 1 to 4 Measurement...
Страница 296: ...pulse analysis measurements Enables or disables the display of the duty cycle of the measured power See CALCulate Measurement TRACe MEASurement PULSe DCYCle on page 290 Suffix Window 1 to 4 Measureme...
Страница 297: ...e value Effective for pulse analysis measurements Displays the number of samples per second See CALCulate Measurement TRACe MEASurement TRANsition SPERiod on page 293 Suffix Window 1 to 4 Measurement...
Страница 298: ...MEASurement TRANsition NEGative DURation STATe value Effective for pulse analysis measurements Enables or disables the display of the fall time of the first detected pulse See CALCulate Measurement T...
Страница 299: ...indow TRACe MEASurement TRANsition POSitive DURation STATe value Effective for pulse analysis measurements Enables or disables the display of the rise time of the first detected pulse See CALCulate Me...
Страница 300: ...RST OFF Manual operation See Pos Overshoot on page 87 14 6 9 Time Gate Further information Chapter 7 5 Time Gate on page 89 CALCulate Measurement POWer TGATe Gate AVG OFFSet TIME 300 CALCulate Measure...
Страница 301: ...unit s Manual operation See Start of Gate on page 94 CALCulate Measurement POWer TGATe Gate AVG EXCLude MID OFFSet TIME value Sets length of the fence Suffix Measurement 1 to 8 Measurement Gate 1 to...
Страница 302: ...is defined by its start time and its length Suffix Measurement 1 to 8 Measurement Gate 1 to 4 Time gate Parameters value RST OFF Manual operation See Fence on page 94 CALCulate Measurement POWer TGATe...
Страница 303: ...EXCLude STARt value Defines an interval at the start of the timeslot that is excluded from the measurement Suffix Measurement 1 to 8 Measurement Parameters value Range 0 0 to 15 0 RST 0 0 Default uni...
Страница 304: ...Measurement 1 to 8 Measurement Parameters value Range 50 0e 9 to 0 1 RST 1 0e 3 Default unit s Manual operation See Nominal Width on page 99 CALCulate Measurement POWer TSLot AVG EXCLude MID OFFSet T...
Страница 305: ...1 1 Statistics Measurement Settings In a statistics measurement you can use commands that combine several setting com mands They are described in Chapter 14 4 6 Statistics Measurements on page 220 For...
Страница 306: ...ensor STATistics PEAK Queries the peak power value calculated during a CCDF or PDF statistics measure ment Suffix Sensor 1 to 128 Configured sensor connected at port A 1 port D 4 R S NRX B9 101 USB an...
Страница 307: ...asurements Enables or disables the internal additional white Gaussian noise AWGN source If enabled you cannot measure with a second power sensor Suffix Measurement 1 to 8 Measurement Parameters value...
Страница 308: ...11 CALCulate Measurement STATistics SCALe X RLEVel ABSolute 311 CALCulate Measurement STATistics SCALe Y SPACing 312 SENSe Sensor STATistics SCALe X MPWidth Queries the greatest level resolution that...
Страница 309: ...0 Default unit Manual operation See Y Maximum on page 105 CALCulate Measurement STATistics CDF SCALe Y LINear PDIVision value Effective if CALCulate Measurement STATistics SCALe Y SPACing LIN is set...
Страница 310: ...easurement STATistics SCALe X POINts value Effective for statistics measurements Sets the measurement result resolution It specifies the number of pixels that are assigned to the logarithmic level ran...
Страница 311: ...with relative power display Sets the lower limit of the level range for the analysis result This level is assigned to the first pixel The level assigned to the last pixel is equal to the level of the...
Страница 312: ...RKer HORizontal DATA 312 CALCulate Measurement STATistics MARKer VERTical DATA POWer 313 CALCulate Measurement STATistics MARKer X POSition ABSolute 313 CALCulate Measurement STATistics MARKer X POSit...
Страница 313: ...measure ment 1 and so on the command works If port 1 is assigned to measurement 2 for example enable the R S NRP2 emulation using SYSTem LANGuage NRP2 Suffix Measurement 1 to 8 Measurement Usage Query...
Страница 314: ...easurement RELative DirectionalChannel RATio RCOefficient 316 CALCulate Measurement RELative DirectionalChannel RATio RFRatio 316 CALCulate Measurement RELative DirectionalChannel RATio RLOSs 316 CALC...
Страница 315: ...rence Value Reflection Reference Value on page 56 See Reference Value on page 72 CALCulate Measurement RELative DirectionalChannel POWer MAGNitude value Sets the reference value for relative ratio mea...
Страница 316: ...alue on page 72 CALCulate Measurement RELative DirectionalChannel RATio RFRatio value Sets the reference value for the ratio of forward reverse power Suffix Measurement 1 to 8 Measurement DirectionalC...
Страница 317: ...rement DirectionalChannel 1 to 2 1 primary sensor 2 secondary sensor or 1 forward 2 reflection reverse Parameters value Range 0 0 to 1 0 RST 0 5 Default unit Manual operation See Forward Reference Val...
Страница 318: ...atically RST ON SENSe Sensor FUNCtion OFF ALL Channel Requires the sensor interface for R S NRT R S NRX B9 Disables all measurement functions for the specified channel Suffix Sensor 101 R S NRT direct...
Страница 319: ...4 R S NRX B9 101 USB and LAN port 5 to 100 Parameters state ON OFF RST ON Manual operation See Autoscale on page 112 SENSe Sensor POWer REFLection RANGe LIMit DETect value SENSe Sensor POWer POWer RA...
Страница 320: ...lower Effective if SENSe Sensor POWer REFLection RANGe AUTO is disabled Sets the lower scale limit for the power indication The entry has no unit The unit corresponds to the output unit selected by UN...
Страница 321: ...Equals CALCulate Measurement RESolution OOOI LOW All other settings RST LOW Manual operation See Resolution on page 51 14 6 13 Querying Measurement Results CALCulate Measurement COUNt DATA 321 CALCul...
Страница 322: ...sured values The mean value is reset if the auxiliary values are reset Suffix Measurement 1 to 8 Measurement Usage Query only CALCulate Measurement PTPeak DATA Queries the peak to peak distance maximu...
Страница 323: ...3 expression parameters for absolute power measurements Channel calculation func tion expression Description Single measurement SENSn Measured value of sensor A B C or D Sum SENSn SENSm Sum of the va...
Страница 324: ...m measures the reflected power With n and m 1 2 3 4 or A B C D Table 14 5 Additional expression parameters for queries expression Description OFF No channel calculation function is selected For SENSn...
Страница 325: ...EQuency FIXed 325 SENSe Sensor FREQuency CW 325 SENSe Sensor FREQuency TRACk 325 SENSe Sensor FREQuency FIXed frequency SENSe Sensor FREQuency CW frequency Sets the carrier frequency of the applied si...
Страница 326: ...AMpling 328 CALCulate Measurement CHANnel Channel SAMPling 328 CALCulate Measurement CHANnel Channel POWer BURSt DTOLerance 329 CALCulate Measurement CHANnel Channel POWer AVG APERture VALue 329 CALCu...
Страница 327: ...rements Enables or disables the duty cycle correction Suffix Measurement 1 to 8 Measurement Channel 1 to 2 1 primary sensor 2 secondary sensor Parameters state RST OFF Manual operation See Duty Cycle...
Страница 328: ...Channel 1 to 2 1 primary sensor 2 secondary sensor Parameters value RST ON Manual operation See Equivalent Time Sampling on page 116 CALCulate Measurement CHANnel Channel SAMPling value Effective for...
Страница 329: ...1 to 8 Measurement Channel 1 to 2 1 primary sensor 2 secondary sensor Parameters value Range 0 0 to 0 3 RST 0 0 Default unit s Manual operation See Dropout Tolerance on page 117 CALCulate Measurement...
Страница 330: ...ensor Parameters value See Smoothing on page 117 RST OFF Manual operation See Smoothing on page 117 CALCulate Measurement TRACe MEASurement TRANsition ESAMpling AUTO STATe value Effective for pulse an...
Страница 331: ...IST Queries the list of the S parameter data sets that have been loaded to the power sen sor The result of the query indicates the consecutive number and mnemonic of each data set Suffix Sensor 1 to 1...
Страница 332: ...0 0 corresponds to an ideal matched source A value of 1 0 corresponds to total reflection Suffix Sensor 1 to 128 Configured sensor connected at port A 1 port D 4 R S NRX B9 101 USB and LAN port 5 to 1...
Страница 333: ...meters value Range 1 to 10 RST 1 Usage Setting only Manual operation See Frequency dependent offset table on page 120 CALCulate Measurement CHANnel Channel CORRection OFFSet TABLe STATe state Enables...
Страница 334: ...or 1 forward 2 reflection reverse Parameters value Range 200 0 to 200 0 RST 0 0 Default unit dB Manual operation See Offset on page 119 See Offset on page 130 CALCulate Measurement CHANnel Channel SGA...
Страница 335: ...de on page 121 14 8 4 Sensor Filters Further information Chapter 8 3 Filter Settings on page 121 Chapter 14 8 8 3 NRT Filter Settings on page 368 SENSe Sensor AVERage COUNt AUTO RESolution 336 SENSe S...
Страница 336: ...noise in the measurement result The setting is only taken into account if CALCulate Measurement CHANnel Channel AVERage COUNt AUTO TYPE RES CALCulate Measurement CHANnel Channel AVERage COUNt AUTO STA...
Страница 337: ...anual operation See Maximum Settling Time on page 123 CALCulate Measurement CHANnel Channel AVERage COUNt AUTO NSRatio value Effective for continuous average burst average measurements if CALCulate Me...
Страница 338: ...ts Sets the autofilter Suffix Measurement 1 to 8 Measurement Channel 1 to 2 1 primary sensor 2 secondary sensor Parameters value RESolution NSRatio RESolution Sets the averaging number so that the int...
Страница 339: ...mary sensor 2 secondary sensor Parameters mode MOVing REPeat If you use SYSTem PRESet instead of RST the RST value dif fers See Table 14 16 MOVing Outputs intermediate values to facilitate early detec...
Страница 340: ...ng filter If enabled the number of measured values is averaged Averaging reduces the effect of noise so that more reliable results are obtained Suffix Measurement 1 to 8 Measurement Channel 1 to 2 1 p...
Страница 341: ...2 secondary sensor Parameters state RST ON Manual operation See Moving Average State on page 124 CALCulate Measurement CHANnel Channel TRACe AVERage TCONtrol ENUM mode Effective for trace measurements...
Страница 342: ...ACe AVERage STATe value Effective for trace measurements Enables or disables the averaging filter If enabled the number of measured values is averaged Averaging reduces the effect of noise so that mor...
Страница 343: ...ation Chapter 8 4 Range Settings on page 125 CALCulate Measurement CHANnel Channel INPut ATTenuation AUTO 343 CALCulate Measurement CHANnel Channel INPut ATTenuation VALue 344 CALCulate Measurement CH...
Страница 344: ...O OFF is set Sets the input attenuation Suffix Measurement 1 to 8 Measurement Channel 1 to 2 1 primary sensor 2 secondary sensor Parameters value Only two values are possible 0 0 dB and 30 0 dB The en...
Страница 345: ...value Reduces the transition range between the measurement paths 0 1 and 1 2 by the set value the so called cross over level Thus you can improve the measurement accuracy for signals with a high peak...
Страница 346: ...Tem STANdard CATalog Queries the supported standards Use the returned strings for SYSTem STANdard PRESet Usage Query only SYSTem STANdard PRESet string Configures the power sensor and trigger settings...
Страница 347: ...ttings state Specifies whether the display settings are kept unchanged when using SYSTem STANdard PRESet Parameters state ON Only configures the power sensor OFF Configures the power sensor and the di...
Страница 348: ...er BURSt DTOLerance 577 s Exclude time start CALCulate Measurement TSLot TIMing EXCLude STARt 20 s Exclude time end CALCulate Measurement TSLot TIMing EXCLude STOP 30 s Timeslot count CALCulate Measur...
Страница 349: ...APERture VALue 10 ms Duty cycle CALCulate Measurement CHANnel Channel CORRection DCYCle VALue 3 7 Duty cycle correction CALCulate Measurement CHANnel Channel CORRection DCYCle STATe ON Smoothing CALC...
Страница 350: ...el Channel SLOPe POS Trigger holdoff TRIGger Measurement CHANnel Channel HOLDoff 39 9 ms Trigger dropout time TRIGger Measurement CHANnel Channel DTIMe 0 s Trigger hysteresis TRIGger Measurement CHANn...
Страница 351: ...POWer TSLot AVG COUNt 15 Timeslot fence CALCulate Measurement POWer TSLot AVG EXCLude MID STATe OFF Timeslot width CALCulate Measurement POWer TSLot AVG WIDTh 666 667 s Timegate offset CALCulate Meas...
Страница 352: ...e CALCulate Measurement CHANnel Channel POWer BURSt DTOLerance 666 667 s Exclude time start CALCulate Measurement TSLot TIMing EXCLude STARt 15 s Exclude time end CALCulate Measurement TSLot TIMing EX...
Страница 353: ...Nnel Channel POWer AVG APERture VALue 5 ms Duty cycle CALCulate Measurement CHANnel Channel CORRection DCYCle VALue 13 250 Duty cycle correction CALCulate Measurement CHANnel Channel CORRection DCYCle...
Страница 354: ...uetooth DH1 Sensor parameter Remote control command Value Sensor mode CALCulate Measurement TYPE TRACe Aperture time CALCulate Measurement CHANnel Channel POWer AVG APERture VALue 1 25 ms Duty cycle C...
Страница 355: ...TRIGger Measurement CHANnel Channel HYSTeresis 3 dB Table 14 13 Bluetooth DH3 Sensor parameter Remote control command Value Sensor mode CALCulate Measurement TYPE TRACe Aperture time CALCulate Measur...
Страница 356: ...gger dropout time TRIGger Measurement CHANnel Channel DTIMe 0 s Trigger hysteresis TRIGger Measurement CHANnel Channel HYSTeresis 3 dB Table 14 14 Bluetooth DH5 Sensor parameter Remote control command...
Страница 357: ...INT Trigger slope TRIGger Measurement CHANnel Channel SLOPe POS Trigger holdoff TRIGger Measurement CHANnel Channel HOLDoff 7 480 ms Trigger dropout time TRIGger Measurement CHANnel Channel DTIMe 0 s...
Страница 358: ...ALue 120 dB Upper power ratio limit CALCulate Measurement METer DirectionalChannel UPPer DATA RATio VALue 60 dB Relative measurement status CALCulate Measurement RELative DirectionalChannel STATe OFF...
Страница 359: ...TPut STATe 359 CALCulate Measurement CHANnel Channel ROSCillator SOURce 360 SENSe Sensor BANDwidth RESolution TYPE AUTO STATe 360 SENSe Sensor BANDwidth RESolution TYPE VALue 360 SENSe Sensor BANDwidt...
Страница 360: ...ondary sensor Parameters source HOST INTernal REFio RST INTernal SENSe Sensor BANDwidth RESolution TYPE AUTO STATe state Effective for R S frequency selective power sensors If enabled sets the filter...
Страница 361: ...N port 5 to 100 Parameters value Range 10 0 to 400 0e6 RST 25 0e6 Default unit Hz 14 8 8 NRT Measurement Type NRT Mode Settings 361 NRT Correction Settings 365 NRT Filter Settings 368 14 8 8 1 NRT Mod...
Страница 362: ...te video bandwidth using CALCulate Measurement CHANnel Channel NRT VBWidth VALue USER Define the duty cycle by CALCulate Measurement CHANnel Channel NRT BURSt PERiod CALCulate Measurement CHANnel Chan...
Страница 363: ...ement CHANnel Channel NRT CCDF THReshold value Sets the threshold for the complementary cumulative distribution function CCDF If you enter a value without unit the unit is defined by UNIT Measurement...
Страница 364: ...HOLD TIME value Sets the hold time of the peak hold circuit of the power sensor Suffix Measurement 1 to 8 Measurement Channel 1 to 2 1 primary sensor 2 secondary sensor or 1 forward 2 reflection reve...
Страница 365: ...ther information Chapter 8 5 2 NRT Correction Settings on page 129 SENSe Sensor DM STATe 365 SENSe Sensor DM STANdard 366 CALCulate Measurement CHANnel Channel CORRection OFFSet RPLane 366 CALCulate M...
Страница 366: ...RRection OFFSet RPLane plane Selects the power sensor port to which the measurement results are referred to Suffix Measurement 1 to 8 Measurement Channel 1 to 2 1 primary sensor 2 secondary sensor or...
Страница 367: ...eflection reverse Parameters value Range 0 0 to 8 2e6 RST 1 0e6 Default unit Hz Manual operation See WCDMA Chip Rate on page 131 INPut Sensor PORT OFFSet offs Considers the transmission loss in a cabl...
Страница 368: ...nnel NRT APERture MODE 369 CALCulate Measurement CHANnel Channel NRT APERture VALue 370 CALCulate Measurement CHANnel Channel NRT VBWidth VALue 370 SENSe Sensor BWIDth VIDeo FNUMber 371 SENSe Sensor B...
Страница 369: ...T 4 Manual operation See Filter Length on page 122 See Averaging Count on page 132 CALCulate Measurement CHANnel Channel AVERage COUNt ENUM value See CALCulate Measurement CHANnel Channel AVERage COUN...
Страница 370: ...1 to 8 Measurement Channel 1 to 2 1 primary sensor 2 secondary sensor Parameters value Range 0 005 to 0 111 RST 0 037 Default unit s Manual operation See Integration Time on page 132 CALCulate Measure...
Страница 371: ...nge 0 to 2 RST 0 Manual operation See Video Bandwidth on page 132 14 9 Configuring the Test Generator If the sensor check source R S NRX B1 is installed you can use it as a power refer ence for testin...
Страница 372: ...e 146 SOURce PULM STATe state Requires the sensor check source R S NRX B1 Effective if the output signal is enabled using SOURce OUTPut STATe or OUTPut SOURce STATe Sets the signal type of the output...
Страница 373: ...7 OUTPut RECorder output LIMit LOWer RATio VALue 378 OUTPut RECorder output LIMit UPPer CCDF 378 OUTPut RECorder output LIMit UPPer VALue 379 OUTPut RECorder output LIMit UPPer POWer 379 OUTPut RECord...
Страница 374: ...set Sets the measurement that is monitored Parameters index Manual operation See Measurement for Limit Output on page 149 OUTPut MODE output mode Sets the functionality of the Out 1 Trig Out and Trig...
Страница 375: ...NC connectors at the rear 1 Out 1 Trig Out 2 Trig In Out 2 Parameters index Manual operation See Measurement for Recorder Output on page 148 OUTPut RECorder output LIMit LOWer CCDF value Effective if...
Страница 376: ...output is set to REC FREC or RREC Sets the power measurement value that corresponds to 0 V output voltage If you enter a value without unit the unit is defined by UNIT Measurement POWer VALue For furt...
Страница 377: ...100 0 RST 0 0 Default unit pct Manual operation See 0 V Equivalent on page 148 OUTPut RECorder output LIMit LOWer RATio RLOSs value Effective if OUTPut MODE output is set to REC FREC or RREC Sets the...
Страница 378: ...value without unit the unit is defined by UNIT Measurement POWer RATio For further information see Chapter 14 6 1 3 Units on page 245 Suffix output 1 to 2 BNC connectors at the rear 1 Out 1 Trig Out 2...
Страница 379: ...ECorder output LIMit UPPer POWer value Effective if OUTPut MODE output is set to REC FREC or RREC Sets the power measurement value that corresponds to 2 5 V output voltage If you enter a value without...
Страница 380: ...2 BNC connectors at the rear 1 Out 1 Trig Out 2 Trig In Out 2 Parameters value Range 0 0 to 100 0 RST 100 0 Default unit pct Manual operation See 2 5 V Equivalent on page 148 OUTPut RECorder output L...
Страница 381: ...C Sets the power ratio measurement value that corresponds to 2 5 V output voltage If you enter a value without unit the unit is defined by UNIT Measurement POWer RATio For further information see Chap...
Страница 382: ...fix Sensor 1 to 128 Configured sensor connected at port A 1 port D 4 R S NRX B9 101 USB and LAN port 5 to 100 Parameters auto ON OFF ONCE LFR UFR ON Return value if a calibration is in progress OFF Re...
Страница 383: ...o n No suffix required Parameters auto ON OFF ONCE LFR UFR CALibration Sensor ZERO FAST AUTO auto Effective for trace measurements Performs fast zeroing Since the commands are processed very quickly t...
Страница 384: ...sensor equipped with an additional test heater Queries the power difference in W when measuring with the external heater enabled and disabled Suffix Sensor 1 to 128 Configured sensor connected at port...
Страница 385: ...sensor connected at port A 1 port D 4 R S NRX B9 101 USB and LAN port 5 to 100 Usage Query only DIAGnostic INFO OTIMe Queries the count of the built in elapsed time meter The count is always output i...
Страница 386: ...firmware version number If an error has occurred the response also contains the error message in plain text Suffix Sensor 1 to 128 Configured sensor connected at port A 1 port D 4 R S NRX B9 101 USB a...
Страница 387: ...ABLe GAIN MAGNitude 392 MEMory TABLe MAP 392 MEMory TABLe MOVE 392 MEMory TABLe RESet 393 MEMory TABLe SELect 393 MEMory CATalog STATe Queries information on the available setups stored on the R S NRX...
Страница 388: ...up 1 STAT 1212479 Example for table Table 2 TABL 84 Usage Query only MEMory CLEar TABLe Deletes the content of the selected correction table Use MEMory TABLe SELect to select the table Notice Once sen...
Страница 389: ...onse Usage Query only MEMory FREE ALL Queries the used and remaining disk space for setups and frequency dependent cor rection tables Combines the information queried by MEMory FREE STATe MEMory FREE...
Страница 390: ...f the setup 0 Factory set setup cannot be changed 1 to 19 Available memory locations Example MEM STAT MAP test 5 Example MEM STAT MAP test Manual operation See Save Recall Setup on page 134 See Setup...
Страница 391: ...a maximum of 2 digits after the decimal point separated by commas Values with more than 2 decimal places are rounded Default unit Hz Example MEM TABL FREQ 50 00 60 70 3456 Sets 3 frequency values 50...
Страница 392: ...table name on page 120 MEMory TABLe MAP register_name register_number Assigns a name to the correction table stored in the memory location Take care to use recognizable names The R S NRX does not chec...
Страница 393: ...TABLe GAIN MAGNitude MEMory TABLe GAIN POINts MEMory TABLe FREQuency POINts MEMory TABLe FREQuency MEMory TABLe DATA MEMory TABLe DATA POINts Alternatively you can use CALCulate Measurement CHANnel C...
Страница 394: ...Culate Measurement CHANnel Channel AVERage TCONtrol ENUM REPeat MOVing CALCulate Measurement CHANnel Channel TRACe AVERage TCONtrol ENUM REPeat MOVing INITiate Measurement CONTinuous OFF ON Usage Even...
Страница 395: ...d the file to the R S NRX for updating the firmware your application has to assemble a memory block containing SYST FWUP block_data The block_data are definite length arbitrary block data as described...
Страница 396: ...irect copy of the binary rsu file in the following format Single digit indicating how many digits follow to specify the size of the binary file Number that specifies the size of the binary file Binary...
Страница 397: ...ss of the R S NRX Parameters IPaddress Consists of four blocks separated by dots consisting of up to 3 digits Example 108 0 0 255 Range 0 to 255 for each block Manual operation See IPv4 Address on pag...
Страница 398: ...the dynamic host configuration protocol DHCP STATic Enables assigning the IP address manually RST AUTO Manual operation See Address Mode on page 140 SYSTem COMMunicate INET SELF SUBNetmask ADDRess ne...
Страница 399: ...ion to the R S NRX i e terminates the connection and sets it up again Usage Event SYSTem COMMunicate NETWork STATus Queries the network configuration state Usage Query only 14 14 5 Remote Settings SYS...
Страница 400: ...Long or short form of the SCPI command See also Chap ter 15 2 2 Syntax for Device Specific Commands on page 462 Return values Syntax dblock Example SYST HELP SYNT SYST KLOC Query SYSTem KLOCk boolean...
Страница 401: ...tomatic instrument identification USER Customized identification string Define the string using SYSTem IDN ANSWer RST AUTO Manual operation See Customization of IDN on page 143 SYSTem LANGuage languag...
Страница 402: ...tion See Custom OPT String on page 144 SYSTem OPT AUTO status Enables or disables the automatic instrument identification for OPT Parameters status ON OFF RST 1 Manual operation See Customization of O...
Страница 403: ...t A 1 port D 4 R S NRX B9 101 USB and LAN port 5 to 100 Setting parameters sensor Hostname of the power sensor Example ADD NRQ6 101435 Usage Setting only Manual operation See Add Sensor on page 151 SY...
Страница 404: ...14 01 Sensor Name NRP33SN 900004 Serial 900444 Stock Number 1419 7777K02 Technology 3 Path Diode TestLimit 0 160 dB TestLimit pd 0 160 dB Type NRP33SN Uptime 904 Response Usage Query only Manual opera...
Страница 405: ...rint Return values XmlDeviceFootprint dblock Usage Query only SYSTem INFO TERMchar termination Selects the termination characters for information returned by SYSTem INFO INFO SENSe Sensor CATalog SENS...
Страница 406: ...14 8 Date and Time Settings SYSTem DATE 406 SYSTem DATE UTC 406 SYSTem DATE LOCal 406 SYSTem TIME 407 SYSTem TIME UTC 407 SYSTem TIME LOCal 407 SYSTem TIME DSTime MODE 407 SYSTem TIME DSTime RULE 407...
Страница 407: ...em TIME UTC is changed accordingly Parameters hour hh minute mm second ss Manual operation See Time on page 155 SYSTem TIME DSTime MODE dst Enables or disables the automatic clock adjustment for dayli...
Страница 408: ...e SET Sets the start time for an absolute timer Return values year YYYY month MM day DD hour hh min mm sec ss msec SYSTem TZONe hour minute Sets the offset of the local time to the UTC time due to the...
Страница 409: ...ype is not included in the description of the event SYSTem SERRor ALL 409 SYSTem SERRor REMove 409 SYSTem SERRor ALL Queries all events from the system event queue Returns a comma separated list of un...
Страница 410: ...error number and a short description of the error 0 means that no errors have occurred since the error queue was last read out Usage Query only 14 14 9 3 SCPI Communication Error Queue This list conta...
Страница 411: ...Usage Query only SYSTem ERRor CODE NEXT Queries the SCPI communication error queue for the oldest entry and removes it from the queue Returns the error number but no error description Return values E...
Страница 412: ...ue was last read out ErrorDescription Short error description Usage Query only 14 14 10 Locking SYSTem KLOCk klock Disables or enables the front panel keyboard of the instrument including the LOCAL ke...
Страница 413: ...rolling the Positive Transition Part 416 14 15 1 General Status Register Commands STATus PRESet 413 STATus QUEue NEXT 413 STATus PRESet Resets the edge detectors and ENABle parts of all registers to a...
Страница 414: ...us QUEStionable MEASure RegisterIndex CONDition STATus QUEStionable POWer SUMMary CONDition STATus QUEStionable WINDow RegisterIndex CONDition Suffix RegisterIndex 1 to 8 Register Usage Query only 14...
Страница 415: ...ion RegisterIndex ENABle value STATus OPERation ULFail RegisterIndex SUMMary ENABle value STATus QUEStionable BIT bitno ENABle RegisterBit STATus QUEStionable CALibration SUMMary ENABle value STATus Q...
Страница 416: ...TATus DEVice PTRansition value STATus OPERation BIT bitno PTRansition RegisterBit STATus OPERation CALibrating SUMMary PTRansition value STATus OPERation LLFail RegisterIndex SUMMary PTRansition value...
Страница 417: ...S NRX K301 option 14 16 1 R S NRP2 Compatibility This chapter describes all R S NRP2 remote commands that are still functional but not recommended to use if you start afresh If you want to reuse progr...
Страница 418: ...Sition Y value Available for compatibility Recommended R S NRX command CALCulate Measurement STATistics CDF MARKer Y POSition CALCulate Measurement STATistics PDF MARKer Y POSition The command is not...
Страница 419: ...r RATio 422 DISPlay WINDow Window NUMeric Numeric RESolution 422 DISPlay WINDow Window TRACe LOWer 423 DISPlay WINDow Window TRACe UPPer 423 DISPlay WINDow Window TSLot 424 DISPlay ILLumination state...
Страница 420: ...NRX command CALCulate Measurement METer DirectionalChannel LOWer DATA RATio VALue on page 235 Suffix Window 1 to 8 Measurement Parameters value DISPlay WINDow Window ANALog UPPer POWer value Availabl...
Страница 421: ...ow Window FORMat format Available for compatibility Recommended R S NRX command CALCulate Measurement DMODe The command is only fully R S NRP2 compatible if SYSTem LANGuage NRP2 is set Suffix Window 1...
Страница 422: ...t METer DirectionalChannel UPPer DATA POWer on page 238 Suffix Window 1 to 4 Measurement pane Parameters value DISPlay WINDow Window METer UPPer RATio value Available for compatibility Recommended R S...
Страница 423: ...SCALe TOP DBM CALCulate Measurement TRACe Y SCALe TOP DBUV CALCulate Measurement TRACe Y SCALe TOP DPCT CALCulate Measurement TRACe Y SCALe TOP ONE CALCulate Measurement TRACe Y SCALe TOP WATT Suffix...
Страница 424: ...TSLot AVG SELection Suffix Window 1 to 8 Measurement Parameters slot Range 1 to 4 RST 1 14 16 1 3 OUTPut Commands OUTPut RECorder output STATe 424 OUTPut RECorder output FEED VALue 425 OUTPut ROSCill...
Страница 425: ...or compatibility Recommended R S NRX command OUTPut SOURce STATe Parameters state OUTPut TRIGger STATe state Available for compatibility Recommended R S NRX command OUTPut MODE output Parameters state...
Страница 426: ...ERage COUNt AUTO SLOT 429 SENSe Sensor AVERage COUNt AUTO TYPE 429 SENSe Sensor AVERage COUNt AUTO STATe 429 SENSe Sensor AVERage COUNt ENUM 430 SENSe Sensor AVERage COUNt VALue 430 SENSe Sensor AVERa...
Страница 427: ...ensor POWer TSLot AVG EXCLude MID OFFSet TIME 441 SENSe Sensor POWer TSLot AVG EXCLude MID TIME 441 SENSe Sensor POWer TSLot AVG EXCLude MID STATe 441 SENSe Sensor POWer CCDFunction REFerence 442 SENS...
Страница 428: ...TRACe MEASurement TRANsition ESAMpling AUTO STATe 454 SENSe Sensor TRACe MEASurement TRANsition NEGative DURation 454 SENSe Sensor TRACe MEASurement TRANsition NEGative OCCurrence 455 SENSe Sensor TR...
Страница 429: ...ted at port A 1 port D 4 R S NRX B9 101 USB and LAN port 5 to 100 Parameters slot Range 1 to 128 RST 1 SENSe Sensor AVERage COUNt AUTO TYPE type Available for compatibility Recommended R S NRX command...
Страница 430: ...ers value E1 E2 E4 E8 E16 E32 E64 E128 E256 RST E4 SENSe Sensor AVERage COUNt VALue count Available for compatibility Recommended R S NRX command CALCulate Measurement CHANnel Channel AVERage COUNt VA...
Страница 431: ...Wer SENSe Sensor AVERage STATe state Available for compatibility Recommended R S NRX command CALCulate Measurement CHANnel Channel AVERage STATe Suffix Sensor 1 to 128 Configured sensor connected at p...
Страница 432: ...ERiod Suffix Sensor 1 to 128 Configured sensor connected at port A 1 port D 4 R S NRX B9 101 USB and LAN port 5 to 100 Parameters value Range 0 0 to 1 0 RST 0 1 Default unit s SENSe Sensor BURSt WIDTh...
Страница 433: ...and LAN port 5 to 100 Parameters state ON OFF RST OFF SENSe Sensor CORRection DCYCle INPut MAGNitude duty_cycle Available for compatibility Recommended R S NRX command CALCulate Measurement CHANnel C...
Страница 434: ...tibility Recommended R S NRX command CALCulate Measurement CHANnel Channel CORRection OFFSet TABLe STATe Suffix Sensor 1 to 128 Configured sensor connected at port A 1 port D 4 R S NRX B9 101 USB and...
Страница 435: ...r 1 to 128 Configured sensor connected at port A 1 port D 4 R S NRX B9 101 USB and LAN port 5 to 100 Parameters offset Range 200 00 to 200 00 RST 0 0 Default unit dB SENSe Sensor DATA function Availab...
Страница 436: ...ort 5 to 100 Parameters function RST POWer AVG SENSe Sensor INFormation argument Available for compatibility Recommended R S NRX command SYSTem SENSor Sensor INFO Suffix Sensor 1 to 128 Configured sen...
Страница 437: ...ent CHANnel Channel JITTer METHod Suffix Sensor 1 to 128 Configured sensor connected at port A 1 port D 4 R S NRX B9 101 USB and LAN port 5 to 100 Parameters method COMPensate MEASure NONE RST COMPens...
Страница 438: ...Configured sensor connected at port A 1 port D 4 R S NRX B9 101 USB and LAN port 5 to 100 Parameters tolerance Range 0 00 to 0 30 RST 1 000e 6 Default unit s SENSe Sensor POWer NCORrection STATe state...
Страница 439: ...ers time Default unit s SENSe Sensor POWer TGATe Gate TIME time Available for compatibility Recommended R S NRX command CALCulate Measurement POWer TGATe Gate AVG TIME Suffix Sensor 1 to 128 Configure...
Страница 440: ...al Default unit s SENSe Sensor POWer TGATe EXCLude MID STATe state Available for compatibility Recommended R S NRX command CALCulate Measurement POWer TGATe Gate AVG EXCLude MID STATe Suffix Sensor 1...
Страница 441: ...ement POWer TSLot AVG EXCLude MID OFFSet TIME Suffix Sensor 1 to 128 Configured sensor connected at port A 1 port D 4 R S NRX B9 101 USB and LAN port 5 to 100 Parameters time Range 0 00 to 0 10 RST 0...
Страница 442: ...n see Chapter 14 6 1 3 Units on page 245 Suffix Sensor 1 to 128 Configured sensor connected at port A 1 port D 4 R S NRX B9 101 USB and LAN port 5 to 100 Parameters ref Range 290 0 to 110 0 RST 0 0 De...
Страница 443: ...vel Range 20 00 to 0 00 RST 0 00 Default unit dB SENSe Sensor RANGe VALue range Available for compatibility Recommended R S NRX command CALCulate Measurement CHANnel Channel POWer AVG RANGe VALue Suff...
Страница 444: ...ROSCillator SOURce value Available for compatibility Recommended R S NRX command CALCulate Measurement CHANnel Channel ROSCillator SOURce Suffix Sensor 1 to 128 Configured sensor connected at port A...
Страница 445: ...PHASe Suffix Sensor 1 to 128 Configured sensor connected at port A 1 port D 4 R S NRX B9 101 USB and LAN port 5 to 100 Parameters phase Range 360 0 to 360 0 RST 0 0 Default unit degree SENSe Sensor S...
Страница 446: ...NSe Sensor STATistics SCALe X RANGe range Available for compatibility Recommended R S NRX command CALCulate Measurement STATistics SCALe X RANGe Suffix Sensor 1 to 128 Configured sensor connected at p...
Страница 447: ...0 01 Default unit s SENSe Sensor STATistics EXCLude MID OFFSet TIME time Available for compatibility Recommended R S NRX command CALCulate Measurement POWer TGATe Gate AVG EXCLude MID OFFSet TIME Suf...
Страница 448: ...de STOP exclude_stop Available for compatibility Recommended R S NRX command CALCulate Measurement TSLot TIMing EXCLude STOP Suffix Sensor 1 to 128 Configured sensor connected at port A 1 port D 4 R S...
Страница 449: ...Measurement CHANnel Channel TRACe AVERage STATe Suffix Sensor 1 to 128 Configured sensor connected at port A 1 port D 4 R S NRX B9 101 USB and LAN port 5 to 100 Parameters state ON OFF RST ON SENSe S...
Страница 450: ...to 100 Parameters value Range Depends on sensor RST Depends on sensor Default unit pct SENSe Sensor TRACe MEASurement DEFine TRANsition HREFerence value Available for compatibility Recommended R S NR...
Страница 451: ...B9 101 USB and LAN port 5 to 100 Usage Query only SENSe Sensor TRACe MEASurement POWer HREFerence Available for compatibility Recommended R S NRX command CALCulate Measurement TRACe MEASurement POWer...
Страница 452: ...for compatibility Recommended R S NRX command CALCulate Measurement TRACe MEASurement POWer MIN If you enter a value without unit the unit is defined by SENSe Sensor UNIT POWer VALue For further infor...
Страница 453: ...a value without unit the unit is defined by SENSe Sensor UNIT POWer VALue For further information see Chapter 14 6 1 3 Units on page 245 Suffix Sensor 1 to 128 Configured sensor connected at port A 1...
Страница 454: ...tion Suffix Sensor 1 to 128 Configured sensor connected at port A 1 port D 4 R S NRX B9 101 USB and LAN port 5 to 100 Usage Query only SENSe Sensor TRACe MEASurement TRANsition ESAMpling AUTO STATe va...
Страница 455: ...tive OVERshoot Available for compatibility Recommended R S NRX command CALCulate Measurement TRACe MEASurement TRANsition NEGative OVERshoot Suffix Sensor 1 to 128 Configured sensor connected at port...
Страница 456: ...port A 1 port D 4 R S NRX B9 101 USB and LAN port 5 to 100 Usage Query only SENSe Sensor TRACe MEASurement TRANsition SPERiod Available for compatibility Recommended R S NRX command CALCulate Measurem...
Страница 457: ...protocol IVI 6 1 TCPIP host address hislip0 INSTR VISA Chapter 15 1 2 3 HiSLIP Protocol on page 460 Socket communica tion SCPI raw TCPIP host address LAN device name port SOCKET Chapter 15 1 2 4 Sock...
Страница 458: ...a communication session with the sensor It is based on the sensor address and some instrument and vendor specific information The syntax of the used USB resource string is USB vendor ID product ID ser...
Страница 459: ...ddress is assigned by a DHCP server when the sensor requests one Alternatively the IP address is deter mined with a procedure called Zeroconf You can also assign a LAN device name which defines the pr...
Страница 460: ...as the network transport layer The TCP IP network protocol and the associated network services are preconfigured TCP IP ensures connection oriented communication where the order of the exchanged mess...
Страница 461: ...s are established on a specially defined port The socket address is a combination of the IP address or hostname of the sensor and the number of the port configured for remote control The power sensors...
Страница 462: ...ature a long form and a short form The short form is marked by upper case letters here to distinguish it from the long form which constitutes the com plete word Either the short form or the long form...
Страница 463: ...nd INIT IMM is equivalent to INIT Parameters Parameters must be separated from the header by a white space If several parame ters are specified in a command they are separated by a comma For a descrip...
Страница 464: ...r set unit is used See also Chapter 14 6 1 3 Units on page 245 If you need decimal multiples and submultiples of a unit you can use SCPI prefixes see Table 15 2 Because SCPI uses only capital letters...
Страница 465: ...nfinite and the representation of missing values Boolean parameters Boolean parameters represent two states The ON state logically true is represen ted by ON or a numeric value 1 The OFF state logical...
Страница 466: ...or other considerations prevent segmentation of the data into blocks of definite length 15 2 4 Overview of Syntax Elements The following table provides an overview of the syntax elements The colon se...
Страница 467: ...s as follows TRIG LEV 0 1E 3 DEL 3E 3 A new command line always begins with the complete path Example TRIG LEV 0 1E 3 TRIG DEL 3E 3 15 2 6 Responses to Queries A query is defined for each setting comm...
Страница 468: ...the ESE Setting bit 5 in the SRE Waiting for service request SRQ OPC Stops command processing until 1 is returned Occurs when all pending operations are completed Send OPC directly after the command w...
Страница 469: ...hapter 15 4 3 Status Byte STB and Service Request Enable Register SRE on page 472 2 Chapter 15 4 5 Device Status Register on page 473 3 Chapter 15 4 6 Questionable Status Register on page 475 4 Chapte...
Страница 470: ...register is always m 1 The suffix of the CALC command corresponds to the measurement channel m x Bit 0 summarizes the status of the next lower status register 15 4 2 Structure of a SCPI Status Registe...
Страница 471: ...set NTR bit 0 The EVENt bit is not set This part can be written into and read as required Its contents are not affected by reading EVENt status register part The EVENt part indicates whether an event...
Страница 472: ...equest triggers an interrupt in the controller configured for this purpose and can be further processed by the controller Set and read the service request enable register using SRE See Figure 15 1 Tab...
Страница 473: ...to 1 A set bit denotes that an action is being performed by the instrument Information on the type of action can be obtained by querying the operation sta tus register Chapter 15 4 8 Operation Status...
Страница 474: ...ter Querying the register STATus DEVice CONDition STATus DEVice EVENt Table 15 5 Used device status bits and their meaning Bit no Short description Bit is set if 1 Sensor A connected 2 Sensor B connec...
Страница 475: ...nt Summary 0 0 0 Questionable Calibration Summary POST Failure 0 0 0 0 Warning 0 Figure 15 4 Questionable status register Querying the register STATus QUEStionable CONDition STATus QUEStionable EVENt...
Страница 476: ...er Sensor D Power NRT Sensor Power Other Sensor Power 0 0 0 0 0 0 0 0 0 Figure 15 5 Questionable power status register Querying the register STATus QUEStionable POWer SUMMary CONDition STATus QUEStion...
Страница 477: ...onable measurement status registers Querying the register STATus QUEStionable MEASure RegisterIndex CONDition STATus QUEStionable MEASure RegisterIndex SUMMary EVENt STATus QUEStionable WINDow Registe...
Страница 478: ...r A Calibration Sensor B Calibration Sensor C Calibration Sensor D Calibration NRT Sensor Calibration Other Sensor Calibration 0 0 0 0 0 0 0 0 0 Figure 15 7 Questionable calibration status register Qu...
Страница 479: ...C is received 2 Query error The controller wants to read data from the instrument but has not sent a query or it sends new commands to the instrument before it retrieves existing requested data A freq...
Страница 480: ...ation CONDition STATus OPERation EVENt Table 15 11 Used operation status bits and their meaning Bit no Short description Bit is set if 0 Operation CAL summary Summary of Operation Calibrating Status R...
Страница 481: ...r calibrating Other Sensor calibrating 0 0 0 0 0 0 0 0 0 Figure 15 10 Operation calibrating status register Querying the register STATus OPERation CALibrating SUMMary CONDition STATus OPERation CALibr...
Страница 482: ...0 0 0 Figure 15 11 Operation measuring status register Querying the register STATus OPERation MEASuring SUMMary CONDition STATus OPERation MEASuring SUMMary EVENt Table 15 13 Used operation measuring...
Страница 483: ...ement m 6 measuring Measurement m 7 measuring Measurement m 8 measuring 0 0 0 0 0 0 0 Figure 15 12 Operation measuring extended status registers Querying the register STATus OPERation MEASuring EXTens...
Страница 484: ...9 10 11 12 13 14 15 Extension Summary Sensor A wait for trigger Sensor B wait for trigger Sensor C wait for trigger Sensor D wait for trigger NRT Sensor wait for trigger Other Sensor wait for trigger...
Страница 485: ...sure ment channel has been waiting for a trigger event since the last query For information on index m see Chapter 15 4 1 Hierarchy of the Status Registers on page 469 0 1 2 3 4 5 6 7 8 9 10 11 12 13...
Страница 486: ...surement m 4 wait for trigger Channel 4 5 Measurement m 5 wait for trigger Channel 5 6 Measurement m 6 wait for trigger Channel 6 7 Measurement m 7 wait for trigger Channel 7 8 Measurement m 8 wait fo...
Страница 487: ...is set if a sensor is initializing 1 Sensor A initializing Sensor A 2 Sensor B initializing Sensor B 3 Sensor C initializing Sensor C 4 Sensor D initializing Sensor D 5 NRT sensor initializing R S NR...
Страница 488: ...s OPERation LLFail RegisterIndex SUMMary EVENt Table 15 18 Used operation lower limit fail status bits and their meaning Bit no Short description Bit is set if measured values of a measurement channel...
Страница 489: ...r Limit Fail Measurement m 7 Upper Limit Fail Measurement m 8 Upper Limit Fail 0 0 0 0 0 0 0 Figure 15 17 Operation upper limit fail status registers Querying the register STATus OPERation ULFail Regi...
Страница 490: ...ort description Bit is set if measured values of a measurement channel exceed the upper limit value or for the summary 7 Measurement m 7 upper limit fail Channel 7 8 Measurement m 8 upper limit fail C...
Страница 491: ...uding the installed hardware and software options is availa ble under Chapter 11 2 Instrument Info on page 153 The hardware configuration is provided separately under Chapter 11 3 Hardware Configurati...
Страница 492: ...fications 1004 firmware update error 492 1005 settings conflict 492 1007 target descriptor error 492 1008 temperature alert 492 1009 fan failure alert 493 1004 firmware update error Firmware update fa...
Страница 493: ...493 2003 sensor settings conflict 493 2005 sensor underrange 493 2006 sensor overrange 494 2007 sensor overload 494 2008 sensor protocol minor mismatch 494 2009 sensor protocol major mismatch 494 200...
Страница 494: ...ate of the power sensor firmware is recommended 2009 sensor protocol major mismatch The firmware version of the connected power sensor is not supported any more Solution Update the firmware of the con...
Страница 495: ...a software option See SW Options tab on page 161 4002 remote language not allowed NRX K301 option missing For the feature you want to use the NRX KS emulation mode R S NRX K301 option is required The...
Страница 496: ...and save the information for troubleshooting See Creating information for troubleshooting on page 167 Attach the archive file to an email in which you describe the problem 16 5 Contacting Customer Su...
Страница 497: ...DNS Domain name system E EMC Electromagnetic compatibility EMI Electromagnetic interference G GPIB General purpose interface bus H HiSLIP High speed LAN instrument protocol I IDN Instrument identifica...
Страница 498: ...ual 1178 5566 02 08 SSH Secure shell SWR Standing wave ratio U USB Universal serial bus V VISA Virtual instrument software architecture VNC Virtual network computing W WCDMA Wideband code division mul...
Страница 499: ...ude MID TIME 441 SENSe Sensor POWer TSLot AVG EXCLude MID STATe 441 SENSe Sensor ADD 403 SENSe Sensor AUXiliary 270 SENSe Sensor AVERage COUNt AUTO MTIMe 428 SENSe Sensor AVERage COUNt AUTO NSRatio 42...
Страница 500: ...ion ON 436 SENSe Sensor INFormation 436 SENSe Sensor INPut ATTenuation AUTO 436 SENSe Sensor INPut ATTenuation VALue 437 SENSe Sensor INTernal TRIGger JITTer METHod 437 SENSe Sensor POWer CCDFunction...
Страница 501: ...Sensor TRACe MEASurement OFFSet TIME 286 SENSe Sensor TRACe MEASurement POWer AVG 451 SENSe Sensor TRACe MEASurement POWer HREFerence 451 SENSe Sensor TRACe MEASurement POWer LREFerence 451 SENSe Sen...
Страница 502: ...GMC 181 GOPC 181 GWAI 181 IDN 181 IST 181 LMC 182 OPC 182 OPT 182 PMC 182 PRE 182 PSC 183 RCL 183 RMC 183 RST 183 SAV 183 SRE 184 SRQ 184 STB 184 TRG 184 TST 184 WAI 185 XESE 185 XESR 185 XPRE 185 XSR...
Страница 503: ...ionalChannel UPPer DATA RATio VALue 255 CALCulate Measurement LIMit undef BEEP STATe 417 CALCulate Measurement LIMit undef CLEar AUTO 248 CALCulate Measurement LIMit undef CLEar IMMediate 248 CALCulat...
Страница 504: ...tics PDF SCALe Y TOP 309 CALCulate Measurement STATistics POWer AVG DATA 306 CALCulate Measurement STATistics SAMPles MINimum 307 CALCulate Measurement STATistics TGATe SELection 307 CALCulate Measure...
Страница 505: ...Ce Y SCALe TOP DBM 242 CALCulate Measurement TRACe Y SCALe TOP DBUV 242 CALCulate Measurement TRACe Y SCALe TOP DPCT 242 CALCulate Measurement TRACe Y SCALe TOP ONE 243 CALCulate Measurement TRACe Y S...
Страница 506: ...ment CHANnel Channel TRACe AVERage COUNt VALue 340 CALCulate Measurement CHANnel Channel TRACe AVERage TCONtrol AUTO 341 CALCulate Measurement CHANnel Channel TRACe AVERage TCONtrol ENUM 341 CALCulate...
Страница 507: ...05 CONFigure Measurement SCALar POWer BURSt RATio RELative 208 CONFigure Measurement SCALar POWer BURSt RATio 207 CONFigure Measurement SCALar POWer BURSt REFLection 210 CONFigure Measurement SCALar P...
Страница 508: ...AVALue 421 DISPlay WINDow Window FORMat 421 DISPlay WINDow Window METer AUTO 243 DISPlay WINDow Window METer LOWer 244 DISPlay WINDow Window METer LOWer POWer 421 DISPlay WINDow Window METer LOWer RAT...
Страница 509: ...Surement TRANsition NEGative DURation STATe 298 DISPlay WINDow Window TRACe MEASurement TRANsition NEGative OCCurrence STATe 298 DISPlay WINDow Window TRACe MEASurement TRANsition NEGative OVERshoot S...
Страница 510: ...FETCh Measurement SCALar POWer AVG RATio 191 FETCh Measurement SCALar POWer AVG REFLection 193 FETCh Measurement SCALar POWer AVG RELative 189 FETCh Measurement SCALar POWer AVG RLOSs 193 FETCh Measur...
Страница 511: ...SCALar POWer TSLot RATio RELative 217 MEASure Measurement SCALar POWer TSLot RATio 216 MEASure Measurement SCALar POWer TSLot REFLection 219 MEASure Measurement SCALar POWer TSLot RELative 213 MEASur...
Страница 512: ...o VALue 378 OUTPut RECorder output LIMit LOWer VALue 376 OUTPut RECorder output LIMit UPPer CCDF 378 OUTPut RECorder output LIMit UPPer POWer 379 OUTPut RECorder output LIMit UPPer RATio RCOefficient...
Страница 513: ...3 READ Measurement SCALar POWer TSLot RATio RELative 217 READ Measurement SCALar POWer TSLot RATio 216 READ Measurement SCALar POWer TSLot REFLection 219 READ Measurement SCALar POWer TSLot RELative 2...
Страница 514: ...4 STATus OPERation MEASuring SUMMary CONDition 414 STATus OPERation MEASuring SUMMary ENABle 415 STATus OPERation MEASuring SUMMary NTRansition 415 STATus OPERation MEASuring SUMMary PTRansition 416 S...
Страница 515: ...er SUMMary NTRansition 416 STATus QUEStionable POWer SUMMary PTRansition 416 STATus QUEStionable POWer SUMMary EVENt 415 STATus QUEStionable PTRansition 416 STATus QUEStionable WINDow RegisterIndex CO...
Страница 516: ...Tem HELP SYNTax 400 SYSTem IDN ANSWer 400 SYSTem IDN AUTO 401 SYSTem IDN MODE 401 SYSTem INFO TERMchar 405 SYSTem INFO INFO 405 SYSTem KLOCk 412 SYSTem LANGuage 401 SYSTem LOCK SHARed STRing 412 SYSTe...
Страница 517: ...ANnel Channel JITTer METHod 263 TRIGger Measurement CHANnel Channel LEVel 263 TRIGger Measurement CHANnel Channel MASTer PORT 264 TRIGger Measurement CHANnel Channel MASTer STATe 265 TRIGger Measureme...
Страница 518: ...72 Control Measurement 59 Cursor keys 27 Customer support 496 Collecting information 496 D Data sheets 15 Date and Time dialog 155 Default settings 134 Delete key 26 Details tab 163 Device status regi...
Страница 519: ...el 44 Setup 46 Starting 46 Measurement Main Configuration dialog 112 Measurement panes Swapping 38 Measurement results 71 Controlling 60 Measurement settings 50 71 Measurement Settings dialog 67 Stati...
Страница 520: ...57 Changing to 42 Command syntax 462 Common commands 177 Emulation 417 Ethernet interface 458 HiSLIP protocol 460 Interfaces 457 Leaving 43 Protocols 457 Query 467 Response 467 SCPI command structure...
Страница 521: ...peration sense status register 486 Operation status register 480 Operation trigger extended status register 485 Operation trigger status register 484 Operation upper limit fail status register 489 Par...
Страница 522: ...ng back 31 Main measurement dialog 32 Measurement panes 38 Notification center 35 Parameters 39 Restricting access 40 Screenshots 40 Settings conflict 48 Start dialog 31 Status information 34 User int...