5 Line codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes
85
or
"<TEST RNG" flag, HIF returns the value and the lower
range limit:
00_•3.01_U/l•••_••
0_••0_•30_••0_•1.02_•5.00<LF>
q
Result will contain the concentration value, even if
the flag is ">TEST RNG" or "<TEST RNG".
Range limit
• Contains the upper limit in the case that we get a
">TEST RNG" flag.
• Contains the lower limit in the case that we get a
"<TEST RNG" flag.
• Contains 0.00 in any other case (flag or not).
Line code 01: Result Time
Example
01_22:07:59<LF>
Line code 02: Control ID
Example
02_ROC-N<LF>
Description
Format
Range
Line code
01
Time
T1
hh:mm:ss
y
Line code 01: Result Time
Description
Format
Range
Line code
02
QC-name
S5
y
Line code 02: Control ID