5 Line codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
84
List of line codes
List of line codes
Line code 00: Result Data
Examples
Normal result without any flag:
00_•8.01_U/l•••_••
0_••0_••0_••0_•0.00_•0.00<LF>
or
">TEST RNG" flag, HIF returns the value and the higher
range limit:
00_10.01_U/l•••_••
0_••0_•31_••0_•1.02_•10.00<LF>
Line code
In
Line code
In
00
Result Data
43
Test State
01
Result Time
44
Cal/CS State
02
Control ID
50
Patient ID
03
Standard Rates
51
Patient Information
04
Calibration Curve
52
Special Order Selection
07
ABS Sample Check
53
Order ID
10
Result Type Selection
54
Order Information
11
Reagent lot information
55
Test ID
12
Control Lot Information
56
Sample name
13
Calibration Lot Information
57
Priority Modification
15
Access Date Time
96
Error Code
16
Result Indexes
97
System Status
40
Service Selection
98
Protocol Version
41
Slot State
99
General Error Code
42
Tube Information
y
List of line codes
Description
Format
Range
Line code
00
Result
F13
+/-9.99.. +/-1.0E-99
Unit
S6
Flag X
I3
0..255
Flag S
I3
0 (not used, currently set to 0)
Flag CALC
I3
0..255
Flag QC
I3
0..255
Range value to flag F13
+/-9.99.. +/-1.0E-99
Range limit
F13
+/-9.99.. +/-1.0E-99
y
Line code 00: Result Data