4 Block codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
72
Order blocks
Response to a test addition with default
sample type, if the previously used sample
type was not a default sample type
Response to an incorrect special order
x = 1 to 200
Response to an order or test deletion
Response to a correct order or test deletion
Response to an order or test deletion with
error in line format
Response to an incorrect order or test
deletion with fault in sample order
Response to an incorrect test deletion
Line
code
Description
Iteration
(1)
(1) x = 1 to 200
96
Error Code
1 to x
53
Order ID
55
Test ID
y
Block 19: Response to a test addition with default sample
type, if the previously used sample type was not a default
sample type
Line
code
Description
Iteration
96
Error Code
52
Specific Order Selection
55
Test ID
y
Block 19: Response to an incorrect special order
Line
code
Description
Iteration
96
Error Code
y
Block 19: Response to a correct order or test deletion
Line
code
Description
Iteration
96
Error Code
y
Block 19: Response to an order or test deletion with error in
line format
Line
code
Description
Iteration
96
Error Code
53
Order ID
y
Block 19: Response to an incorrect order or test deletion
with fault in sample order
Line
code
Description
Iteration
96
Error Code
1 to x
55
Test ID
y
Block 19: Response to an incorrect test deletion