5 Line codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
88
List of line codes
Line code 10: Result Type Selection
Example
10_07<LF>
q
o
Only accepted and not yet transmitted results are sent to
the Host.
o
Due to instrument performance reasons and to
guarantee that all results are transmitted, do not use
collated results requests (i.e., do not use result type
selector 03, 04 and 05) those are only for backwards
compatibility.
o
Collated results requests also give slightly different
output from unspecific results requests, with V3.4 and
higher.
o
Calibration and control results are only transmitted if
enabled.
u
See Known host interface issues (178)
u
See Line code 53: Order ID (96)
u
See Configuring host communication (190)
Description
Format
Range
Range description
Line code
10
Result type selector
12
01
Unspecific result request: Transmit any available result, i.e.
o
Single patient, or
o
Single calibration, or
o
Control result collated by test ID
02
Reserved
03
Transmit patient results, collated by order ID.Results are only
transmitted if all tests belonging to an order are finished and
accepted.
Do not use result
type selector 03,
04, 05 or 06
See the note
below.
04
Transmit patient results, collated by order ID.Blocked tests are
marked with the test blocked flag (CALC flag 255). Their results
are sent later as soon as successfully completed along with any
still-remaining blocked tests.
05
Transmit patient results, collated by order ID.Send all currently
available results (i.e., all accepted but not yet transmitted
results)
06
Transmit new control results collated by control ID.
07
Transmit single patient result.
08
Transmit single calibration result.
09
Transmit control results collated by test ID.
11
Transmit any available result with relevant lot(s) information.
17
Transmit any single patient result with reagent lot information.
18
Transmit any single calibration result with reagent & calibration
info.
19
Transmit any single control result by test ID with reagent &
control info.
y
Line code 10: Result Type Selection