7 Block codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
124
Result blocks
Block 06: Control Result with Lot Information
Block 06 is transmitted to the host if
all
of the following
conditions are met:
• After receiving a result request with result type
selector 11 or 19
• If an accepted and not yet transmitted control result is
available
• If the instrument is configured accordingly
Results always collated by test ID
Result type selector 11 or 19 is used.
• m = 1 (one test ID per result block)
• n = 1 to 6
q
Results of different control orders are not collated.
Line code
Description
Iteration
(1)
(1) w = 1 to 141. The value of w can be different from 1 only if the result
was measured using an absorbance test (Line 06). There are 141
possible lines since each line is produced for every cycle (max 141).
55
Test ID
1 to m
15
Access Date Time
16
CALidx
11
Reagent Lot
Information
02
Control ID
1 to n
12
Control Lot
Information
00
Result Data
06/08/09
Raw Data
1 to w
y
Block 06