10 Instrument setup
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Instrument setup
191
Configuration Database > Demographics
Configuration Lab > Sample Types
If sample type is enabled in
Configuration Processing
> Result Output
, this
Sample Type Short Name
is
used on the host interface.
q
If sample-type specific racks are not used, this
Sample Type Short Name
has to be included in the
barcode.
Parameter
Value
Patient ID
Selected
All sample orders have to be
assigned to an existing patient,
i.e. prior to downloading
orders, the relevant patient
information has to be
downloaded. The instrument
generally works with patient
data.
Not selected
The instrument generally
works without patient data.
Orders have to be entered
without patient data.
Free Text
Fields
Selected
The instrument allows defining
laboratory specific fields. The
fields and their assigned
names are shown in the
patient or order menu. The
relevant information has to be
transmitted via the host
interface.
Not selected
The free text fields will not be
shown in the patient or order
menu. On the host interface,
the relevant fields within the
Patient
and
Order Entry
blocks may be left empty.
y
Database parameters