4 Block codes
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Block codes
61
About block codes
The following is the table of contents for the block code
chapter. It also shows the request/response behavior for
host-instrument communication.
Block composition notation
Host (request)
Instrument (response)
Block code Request
Block code
Request
00
Idle Block / Synch Block
00
Idle Block
09
.
Result Request
02
Calibration Result
03
Control Result
04
Patient Result
05
Calibration Result with Lot Information
06
Control Result with Lot Information
07
Patient Result with Lot Information
08
Result Request Response
00
Idle Block, if no result available
10
Order Entry
19
Order Manipulation Response
11
Order Deletion
40
Patient Entry
49
Patient Manipulation Response
41
Patient Deletion
42
Patient Modification
60
Multi Configuration Service Request
61
Slot Configuration
62
Sample Tube Inform.
63
Cal-CS State
69
Service Request Response
90
System Status Request
91
System Status Response
92
Protocol Version Request
93
Protocol Version Data
.-
any Host Request
99
Control Message
y
List of block codes
Line
code
Description
Iteration
55
Test ID
1 to m
(1)
(1) Number of iterations in all lines
05
Access Date Time
16
CALidx
11
Reagent Lot
Information
02
Control ID
1 to n
(2)
(2) Number of iterations in lines 02 and 12
12
Control Lot
Information
y
Notation of block codes