8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes for extended host interface
163
Line Code 49: Rack Position
Example
49_H_2-10<LF>
The first character is either the SLOT_ID letter (A-Z) or a
whitespace if the slot is not on board. It is then followed
by [rack number]-[rack position].
Line code 55: Test ID
Example
55_••1_•••2.00_•••3.50<LF>
q
Profile and ratio handling: Use ratio number to order
ratio.
Ratio and individual tests results will be delivered.
The available test no., profile IDs and Ratio IDs can be
found in
Configuration > Test Definitions > ID
.
Samples in the instrument can only be diluted. The
maximum dilution factor is 100.
If the dilution factor is omitted, the 1:1 factor will be used.
Description
Format Range
Line Code
49
Rack position
S10
SlotID_RackNumber-RackPosition
y
Line Code 49: Rack Position
Description
Format Range
Line code
55
Test no.
I3
1 to 999
Optional
If dilution factor is extra information for block
10
Dilution factor
F7
1 to 100
Automatic Rerun
Dilution factor
F7
1 to 100
y
Line code 55: Test ID