7 Block codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Block codes for extended host interface
129
Test status blocks
In this section
Block 45: Test status request
Block 45 is used to interrogate the test status. Block 45
has no lines.
Block 46: Test Status Response
This response provides the test status for all loaded tests.
Block 47: Failed Test Status Request
Block 47 contains Line 96 and provides the eventual error
encountered following request with Block 45.
Line code
Description
Iteration
(1)(2)(3)(4)(5)
(1) m is the total number of loaded tests.
(2) x is the number of cassettes for a specific test. The maximum value
is set to 6.
(3) y is the number of calibrators for a specific test. The maximum value
is set to 6.
(4) z is the number of controls for a specific test. The maximum value is
set to 6.
(5) w is the number of diluents for a specific test. The maximum value is
set to 6.
55
Test ID
1 to m
28
Test Status
29
Cassette Status
0 to x
31
Calibrator status
0 to y
33
Control status
0 to z
38
Diluents status
0 to w
y
Block 46
Line code
Description
96
Error Message
y
Block 47