Flag CALC (result error)
Flag CALC (result error)
Flag CALC provides information about the checks
performed during the calculation of the result.
q
A flag with a higher priority (lower number)
supersedes all other flags.
Flag value
Priority
Name
Meaning
0
No flag
1
1
HIGH ABS
High absorbency
2
2
DARK ERR
Dark error
3
3
OUTLIERS
Outliers
4
4
FP UNSTAB
FP unstable
5
5
POL ERR
Polarization error
6
6
LOW SIGNAL
Low signal intensity
7
7
NO FLUID
No sample
8
8
NO ISECAL
No calibrator bottle
9
9
SEG FLUID
No end of sample segment detected
10
10
SEG ISECAL
No end of standard segment detected
11
11
AIR FLUID
Not enough sample / too many bubbles
12
12
AIR ISECAL
Not enough standard / too many bubbles
13
13
ISE UNSTAB
ISE measurement unstable
14
14
< REAG RNG
Low reagent range exceeded
15
15
> REAG RNG
High reagent range exceeded
16
16
EP UNSTAB
Endpoint unstable
17
17
HIGH ACT
High activity (please refer to following note)
18
18
LOW ACT
Low activity (please refer to following note)
19
19
CURVE DIR
Curve direction
20
20
NON LINEAR
Non-linear kinetic
21
21
< KIN READ
Insufficient kinetic readings
22
22
HIGH BKG
High background noise (please refer to following note)
23
23
> REPL DEV
Replicate deviation check failed
24
24
CALC ERROR
Calculation error (please refer to following note)
25
25
NON MONO
Curve not monotone
26
26
OUT OF RNG
Curve / slope out of range
27
27
> STD DEV
Standard deviation exceeded
28
28
SOL 1 F DEV
Solution 1 factor deviation
29
29
AG EXCESS
Antigen excess
30
30
< TEST RNG
Lower test range exceeded (please refer to following note)
31
31
> TEST RNG
Upper test range exceeded (please refer to following note)
255
33
TEST BLOCKED
Used with collated results if a test is blocked (see note)
y
List of flag CALC