8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes for extended host interface
161
Line code 40: Service Selection
Example
40_1<LF>
Line Code 45: Service Action Definition
Example
45_103_c<LF>
Description
Format
Range
Range description
Line code
40
Service request I1
0
Slot configuration
1
Sample tubes without
orders
3
Reserved
4
List of all sample tubes
5
Status of all automatic
service actions
y
Line code 40: Service Selection
Description
Format Range
Range description
Line Code
45
Service Action ID
I3
Counts unit
S1
h
counts are given in
hours
c
counts are given in
test counts
y
Line Code 45: Service Action Definition
ID
Description
0
Deproteinize probes
101
Prime fluid system
103
Backup database
301
Electrode service
304
Activate electrodes
310
Prime ISE calibrators
312
Clean ISE tower automatically
314
Condition ISE tubing
315
Initialize ISE module
y
IDs