7 Block codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
120
Result blocks
Result blocks
In this section
Block 02: Calibration Result (120)
Block 03: Control Result (121)
Block 04: Sample Results (122)
Block 05: Calibration Result with Lot Information (123)
Block 06: Control Result with Lot Information (124)
Block 07: Sample Results with Lot Information (125)
Block 02: Calibration Result
Block 02 is transmitted to the host if
all
of the following
conditions are met:
• After receiving a result request with result type
selector 01 or 08
• If an accepted and not yet transmitted calibration
result is available
• If the instrument is configured accordingly
Line code
Description
Iteration
(1)(2)(3)
(1) m = 1 to 6. m is the number of standard rates in line 03. The
sequence of m iterations must correspond to the sequence of values
in line 03. In addition, the number of iterations m has to be equal to
the number of parameters (standard rates) in line 03.
(2) y = 0 or 1. The value of y depends on whether the antigen excess
(AGE) check is switched ON or OFF. The instrument always transmits
an AGE rate line regardless of this parameter. If the AGE check was
not required by the test, this line contains the values initialized by the
instrument.
(3) n = 1 to 423. The value of n can be different from 1 only if the result
was measured using an absorbance test (Line 06). There are 423
possible lines since each line is produced for every replicate (max. 3)
and every cycle (max 141).
55
Test Id
01
Result Time
03
Standard Rates
1 to y+1
04
Calibrate Curve
00
Result Data
1 to m
06/08/09
Raw Data
1 to n
y
Block 02