7 Block codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Block codes for extended host interface
125
Block 07: Sample Results with Lot Information
Block 07 is transmitted to the host if
all
of the following
conditions are met:
• After receiving a result request with result type
selector 11 or 17
• If an accepted and not yet transmitted sample result is
available
Line code
Description
Iteration
(1)(2)(3)
(1) w = 1 to 141. The value of w can be different from 1 only if the result
was measured using an absorbance test (Line 06). There are 141
possible lines since each line is produced for every cycle (max 141).
(2) p = 0 to x. The value of x is the number of controls used for a single
test result. It is possible that no control was used because QC was
never performed.
(3) y = 1 to z. The value of z is the total number of reruns executed for a
single test result.
53
Order ID
56
Sample Name
0 to 1
15
Access Date Time
16
CALidx
55
Test ID
11
Reagent Lot Information
00
Result data
14
Validation Status
06/08/09
Raw Data
1 to w
17
Test Execution
18
Cassette Information
19
Calibration Data
21
Control Data
0 to p
27
Rerun results
0 to y
y
Block 07