8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Line codes for extended host interface
169
Line Code 83: Message String
Example
83_Wrong•DB•script.<LF>
q
o
The maximum length of each message is 100. If a
message exceeds this length it will be split in multiple Lines
83. If the message is shorter also the line will be shorter.
o
If a
newline
character is present in the message string
this will be displayed as <NL> in order to tell it apart
from <LF> which is used in the Hif Protocol structure
Line Code 93: System Date Time
Example
93_23/04/2006_22:07:59<LF>
Line Code 94: Serial Number
Example
94_CN123456<LF>
Description
Format
Range
Line Code
83
String
S100
y
Line Code 83: Message String
Description
Format
Range
Line Code
93
Date
D1
dd/mm/yyyy
Time
T1
hh:mm:ss
y
Line Code 93: System Date Time
Description
Format
Range
Line Code
94
Serial Number
S8
SMP code and Serial
Number
y
Line Code 94: Serial Number