8 Line codes for extended host interface
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
170
List of line codes for extended HIF
Line Code 97: System status
Example
97_•5_0_0_0_1_•853_0_•••37.1_•••12.6_•••24.2<LF>
Description
Format Range
Range description
Line Code
97
System status
I2
0
Sleeping
1
Initializing
2
Stopped
3
Service
4
Cover Open
5
Standby
6
Running
7
Fatal Error
8
Power Down
9
Auto Service
10
Power On
Cleaner status
I1
0
Ok
1
Warning
2
Error
Cuvette status
I1
0
Ok
1
Warning
2
Error
Water supply
status
I1
0
Ok
1
Warning
2
Error
Cuvette waste
status
I1
0
Ok
1
Warning
2
Error
Cuvette waste
count
I4
0..1000
Waste drain
status
I1
0
Ok
1
Warning
2
Error
Analyzer
temperature
F7
Temperature in degrees
Cassettes
temperature
F7
Temperature in degrees
Cleaner
temperature
F7
Temperature in degrees
y
Line Code 97: System status