63 • Setting Detailed Operational Parameters Using the SDM-PC2 Docking
5 . If necessary, select the desired setting for the
Data Logger Interval
. The
Data
Logger Interval
is the length of time, in days, that will pass before data is
automatically downloaded from a GX-6000 if
Data Logger Auto Download
is
set to
Interval
. If the GX-6000 is connected to an SDM-6000 before the
interval has passed, a data download will not automatically occur. The
Data
Logger Interval
can be set anywhere from
1
to
60
days. The factory setting is
1
day. Use the arrows to the right of the current setting to change the interval
or highlight the current setting and type in the desired value.
6 . Set Manual Download to the desired setting:
•
When set to
ON
(default value), the Download button on the SDM-6000
touchscreen becomes active and a manual download in Standalone Mode
of the GX-6000’s logger data to the SD card can be performed.
•
When set to
OFF
, the Download button is not active and a manual
download of the GX-6000’s logger data to the SD card cannot be
performed in Standalone Mode.
7 . To save the changes, click
SDM Update
and click
Yes
in the window that
appears.
To return to the main window without saving any changes, click
Cancel
and
click
OK
in the window that appears.
Updating the Power Saving Settings
1 . Launch the PC Controller Program and connect your SDM-6000(s) as
described in “Launching the PC Controller Program and Connecting an SDM-
6000” on page 60.
2 . Access the SDM Configuration Window as described in “Accessing the SDM