49 • Setting Basic Operational Parameters in the Config Menu
•
GAS 1 in the top left corner shows what gas inlet is currently selected for
editing.
•
The Active button in the bottom left corner selects whether the gas inlet is
active. This feature is useful if you don’t want to use a particular gas inlet
for a test. For example, if you have a 4-gas + low range PID GX-6000 but
you only want to test the PID sensor, you could deselect the Active button
for the 4-gas inlet.
NOTE:
Since the SDM-6000 only has 3 gas ports, GAS 4, GAS 5, and GAS 6
cannot be selected as active if the SV-3 is not connected.
•
Gas list in the middle of the screen shows what gas(es) are assigned to the
selected inlet.
6 . To change the gas(es) and/or concentration(s) for a gas inlet:
a. Press the desired gas. You can only change one gas at a time.