29 • Wall Mounting the SDM-6000, Optional
b. If you are installing the air filter on an SV-3, the green tubing stub does
not need to be used. Orient the air filter with the arrow pointing toward the
SV-3 and insert the air filter into the AIR recess at the front of the SV-3’s
left side and push it in until the SV-3’s latch secures it in place.
NOTE:
If you are testing a GX-6000 with an IR CO
2
sensor in the 0-10,000
ppm or 0-5.00 %vol range, be sure to attach the CO
2
scrubber to the
air filter to ensure that the CO
2
present in fresh air is scrubbed out. The
arrow on the CO
2
scrubber should be pointing toward the air filter.
5 . Install the 10 foot long 5/16 inch ID flexible tube that is included with the
SDM-6000 on the exhaust fitting. Route the tube to an area where the exhaust
can be safely dispersed, such as an open window.
CAUTION:
The maximum recommended length for the exhaust tube is 15 feet. Do
not use more than 15 feet of tubing or tubing with an ID of less than
5/16 inch for the exhaust tube or the bump test and calibration
accuracy may be adversely affected. The tube that is shipped with the
SDM-6000 has an ID of 5/16 inch and is 10 feet long.
Wall Mounting the SDM-6000, Optional
Single SDM-6000s or multiple SDM-6000s connected together as described in
“Connecting Multiple SDM-6000s Together” on page 23 can be mounted to the wall
using mounting brackets. The wall mounting brackets are not supplied as standard
with the SDM-6000. If you are wall mounting a bank of SDM-6000s, not every
SDM-6000 will need wall mounting brackets but each SDM-6000 that will use wall
mounting brackets needs 3. See the figure below to determine how many wall
mounting brackets you need to order.