Setting Detailed Operational Parameters Using the SDM-PC2 Docking Station PC
3 . Click the Bump & Cal 2 Tab.
Figure 27: Bump & Cal 2 Tab
4 . Select the desired setting for the
Power Off After Pass
parameter. This is the
length of time that will pass after a successful bump test, calibration, or alarm
check before the docking station will automatically shut off the GX-6000. Use
the arrows to the right of the current setting to change the value or highlight the
current setting and type in the desired value. It is defined in seconds with a
maximum setting of
99,999
seconds. The factory setting is
15
seconds.
5 . Select the desired setting for the
Auto Power Off
parameter. This is the length
of time that will pass after the last operation is finished before the docking
station will automatically shut off the GX-6000. If a successful bump test,
calibration, or alarm check is performed, the docking station will automatically
shut off the GX-6000 in the time defined in
Power Off After Pass
regardless
of the
Auto Power Off
setting. Use the arrows to the right of the current setting
to change the value or highlight the current setting and type in the desired
value. It is defined in seconds with a maximum setting of
99,999
seconds. The
factory setting is
600
seconds.
6 . To save the changes, click
SDM Update
and click
Yes
in the window that