93 • Bump Testing
Calibration Gas, SV-3 Attached” on page 35 for calibration gas cylinder options
and calibration gas connection procedures.
NOTE:
If you are testing a GX-6000 with an IR CO
2
sensor in the 0-10,000
ppm or 0-5.00 %vol range, be sure to connect a CO
2
scrubber either to
the AIR fitting (no SV-3 used) or to the air filter (SV-3 used) to ensure
that the CO
2
present in fresh air is scrubbed out. The arrow on the CO
2
scrubber should be pointing toward the AIR inlet fitting.
7 . If necessary, press the Function Tab in the upper left corner to get the Function
Tab to appear.
8 . Press the BUMP button to initiate a bump test.
If you wish to cancel the bump test, press the BUMP Cancel button in the
Function Tab and press Yes in the window that appears.
9 . If there is no SD card installed, a warning screen will appear. If no SD card is
installed, no test data will be saved. Press Yes or No to continue with the bump
test.
10 . Throughout the bump test, the Unit Tab will display the current gas readings
along with the pass/fail status of a bump test or automatic calibration.