43 • Setting Basic Operational Parameters in the Config Menu
5 . Adjust the AIR Flush, GAS, AIR Purge, and Tolerance parameters as
described below. The procedure is the same for each parameter’s adjustment.
AIR Flush is shown as an example.
a. Press the current value for the parameter you want to change.
b. Use the arrows to adjust the value.
c. Press OK to save the changes and return to the Bump Parameters
Screen.
d. Press cancel to return to the Bump Parameters Screen without saving the
changes.
e. Repeat step a - step d to set any other parameters.
6 . Select or deselect the Fast, Auto Cal., Alarm Check, and Auto Print
parameters to turn them on or off.
7 . When you are done setting the parameters, press the Return arrow button on
the right side of the screen. The Configuration Menu will be displayed.
8 . To exit the Configuration Menu and return to the Function Tab’s main menu,
press the return arrow on the right side of the screen.