HP8643 emulation mode
1GP71_0E
124
Rohde & Schwarz
Status
Command / Function / Remark / Limitation
[SOURce:]PM:STATe value
Set / Query the (activation) state of the modulation.
Value:
ON | OFF | 1 | 0
Default:
OFF
[SOURce:]PM:SOURce value
Set / Query the source of the modulation signal.
Value:
INTernal, EXTernal | EXTernal, INTernal
Default:
INTernal
[SOURce:]PM:COUPling value
Select / Query the coupling type, when modulation signal source is set to external.
Value:
AC | DC | GROund
Default:
DC
Remark:
GROund is not supported
[SOURce:]PM:FREQuency value
Alias to LFSource:FREQuency value.
[SOURce:]PM:FREQuency:STEP[:INCRement] value
Alias to LFSource:FREQuency:STEP[:INCRement] value.
Table 117: PM subsystem commands
•
PMETer Subsystem
Status
Command / Function / Remark / Limitation
PMETer[:POWer]?
Table 118: PMETer subsystem commands
•
PULSe Subsystem
Status
Command / Function / Remark / Limitation
PULSe[:STATe] value
Set / Query the (activation) state of the modulation.
Value:
ON | OFF | 1 | 0
Default:
OFF