AN680XX emulation mode
1GP71_0E
50
Rohde & Schwarz
Status
Command / Function / Remark / Limitation
[:SOURce:]SWEep<n>:DWELl value
Set / Query step time for frequency / power level sweep.
Value:
timemin … timemax [S | MS | US | NS] | MIN | MAX
Default:
timemin
Remark:
SWEep[1] is used for frequency sweep, SWEep2 is used for power sweep.
[:SOURce:]SWEep<n>:DWELl:AUTO value
[:SOURce:]SWEep<n>:GENeration value
[:SOURce:]SWEep<n>:POINts value
Set / Query the number of steps in frequency / power level sweep.
Value:
pointsmin … pointsmax | MIN | MAX
Default:
freqspan / freqstep
powerspan / powerstep
Remark:
SWEep[1] is used for frequency sweep, SWEep2 is used for power sweep.
[:SOURce:]SWEep<n>:POWer:STEP value
Set / Query the step size for power level sweep.
Value:
stepmin … stepmax [DB] | MIN | MAX
Default:
0.02 dB
Remark:
SWEep[1] is used for frequency sweep, SWEep2 is used for power sweep.
[:SOURce:]SWEep<n>:SPACing value
Set / Query the type of sweep.
Value:
LINear | LOGarithmic
Default:
LINear for frequency sweep
LOGarithmic for power sweep
Remark: